RayStation 4-11B, and RayPlan 1, 2, 7-11B including some service packs: RayStation 4.0 (4.0.0.14) to RayStation 5 Service Pack 2 (5.0.2.35), RayStation 5 Service Pack 3 (5.0.3.17), RayStation 6/RayPlan 2 (6.0.0.24), RayStation 6/RayPlan 2 Service Pack 1 (6.1.1.2), RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7), RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6), RayStation/RayPlan 7 (7.0.0.19), RayStation/RayPlan 8A (8.0.0.61), RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) Software versions: 4.0.0.14, 4.0.3.4, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 4.7.6.7, 4.9.0.42, 5.0.1.11, 5.0.2.35, 5.0.3.17, 6.0.0.24, 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52, 10.1.0.613, 10.1.1.54, 11.0.0.951, 11.0.1.29, 11.0.3.116 or 12.0.0.932

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI # 07350002010020, 07350002010013, 07350002010082, 07350002010075, 07350002010242, 07350002010068, 07350002010112, 07350002010136, 07350002010129, 07350002010204, 07350002010235, 07350002010174, 07350002010266, 07350002010297, 07350002010303, 07350002010365, 07350002010310, 07350002010471, 07350002010389, 07350002010433, 07350002010440, 07350002010426
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
RaySearch America Inc
Reason for Recall:
Incorrect source to skin or surface distance (SSD) calculation may result in dose error when using RayStation system.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

RayStation 4-11B, and RayPlan 1, 2, 7-11B including some service packs: RayStation 4.0 (4.0.0.14) to RayStation 5 Service Pack 2 (5.0.2.35), RayStation 5 Service Pack 3 (5.0.3.17), RayStation 6/RayPlan 2 (6.0.0.24), RayStation 6/RayPlan 2 Service Pack 1 (6.1.1.2), RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7), RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6), RayStation/RayPlan 7 (7.0.0.19), RayStation/RayPlan 8A (8.0.0.61), RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) Software versions: 4.0.0.14, 4.0.3.4, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 4.7.6.7, 4.9.0.42, 5.0.1.11, 5.0.2.35, 5.0.3.17, 6.0.0.24, 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52, 10.1.0.613, 10.1.1.54, 11.0.0.951, 11.0.1.29, 11.0.3.116 or 12.0.0.932

Product Codes/Lot Numbers:

UDI # 07350002010020, 07350002010013, 07350002010082, 07350002010075, 07350002010242, 07350002010068, 07350002010112, 07350002010136, 07350002010129, 07350002010204, 07350002010235, 07350002010174, 07350002010266, 07350002010297, 07350002010303, 07350002010365, 07350002010310, 07350002010471, 07350002010389, 07350002010433, 07350002010440, 07350002010426

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0850-2022

Related Recalls

A software bug affecting results when using deep learning (DL) segmentation followed by geometrical functions: Create wall function, Expand Contract function or ROI algebra function, the results are incorrect. If dose statistics for an incorrectly created ROI are used to determine if a plan is acceptable for treatment, this could potentially lead to an inappropriate plan being approved. The error could lead to a more conservative or more liberal plan than intended.

Jan 22, 2021 Other Medical Devices View Details →