Folysil Silicone Foley Catheters, Device Reference Numbers AA6106, AA6108, AA6110, AA6112, AA6114, AA6116, AA6118, AA6120, AA6122, AA6124, AA6308, AA6310, AA6312, AA6314, AA6316, AA6318, AA6320, AA6322, AA6324, AA6408, AA6410, AA6412, AA6414, AA6416, AA6418, AA6420, AA6422, AA6424, AA6C18, AA6C20, AA6C22, and AA6C24.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Device Reference Number AA6106, GTIN 3600040126851, Lot Numbers 7278611, 7326825, 7469914, 7528770, 7540052, 7616917, 7680676, 7726724, 7748128, 7794789, 7874610, 7883406, 7948568, 7987184, 7987186, 7998885, 8097945, 8097947, 8097949, 8097950, 8149951, 8149954, 8156925, 8326511, 8326512, 8392428, 8452914, 8452915, 8519853, 8559628, 8572973, 8572977, 8728905, 8742746, 8796890, 8829455, 8899165, 8911320, 8996565, 9139298, 9177631, 9177632, 9192348, 9294521, 9294523, 9294524, 9315917, 9374195, 9404781, 9404782, 9419436, 9429227, 9429228, 9445801, 9453606, 9700652, 9711505, 9778601, 9778603, 9778606, 9792251, 9792257, 9817387, 9829471, 9843572, 9843573, 9940969 ; Device Reference Number AA6108, GTIN 3600040126981, Lot Numbers 7278611, 7326825, 7469914, 7528770, 7540052, 7616917, 7680676, 7726724, 7748128, 7794789, 7874610, 7883406, 7948568, 7987184, 7987186, 7998885, 8097945, 8097947, 8097949, 8097950, 8149951, 8149954, 8156925, 8326511, 8326512, 8392428, 8452914, 8452915, 8519853, 8559628, 8572973, 8572977, 8728905, 8742746, 8796890, 8829455, 8899165, 8911320, 8996565, 9139298, 9177631, 9177632, 9192348, 9294521, 9294523, 9294524, 9315917, 9374195, 9404781, 9404782, 9419436, 9429227, 9429228, 9445801, 9453606, 9700652, 9711505, 9778601, 9778603, 9778606, 9792251, 9792257, 9817387, 9829471, 9843572, 9843573, 9940969; Device Reference Number AA6110, GTIN 3600040127117, Lot Numbers 7255806, 7270683, 7270685, 7278618, 7303190, 7427087, 7440872, 7469982, 7487187, 7503933, 7564542, 7596635, 7629271, 7703729, 7713958, 7774417, 7774419, 7794951, 8133792, 8156966, 8156972, 8297224, 8313405, 8350384, 8519861, 8539757, 8679575, 8781871, 8796821, 8839854, 8839855, 8985535, 8996599, 9008443, 9031094, 9067753, 9077697, 9090464, 9102793, 9213533, 9213536, 9240194, 9240195, 9240196, 9240197, 9248068, 9261075, 9282360, 9282362, 9294615, 9294624, 9304466, 9315932, 9315933, 9341162, 9351850, 9351851, 9374230, 9385336, 9396721, 9405092, 9420752, 9420753, 9429249, 9445822, 9445823, 9472404, 9472405, 9553577, 9553578, 9610617, 9634462, 9634464, 9634465, 9656853, 9656854, 9656855, 9656856, 9656857, 9666697, 9666698, 9683046, 9683047, 9690316, 9700707, 9711520, 9792840, 9792841, 9843580, 9843581 ; Device Reference Number AA6112, GTIN 3600040127247, Lot Numbers 7174166, 7174170, 7174172, 7338746, 7338747, 7338751, 7470039, 7576132, 7576144, 7585050, 7585053, 7607531, 7670529, 7680781, 7894977, 7929763, 7929766, 7929768, 7929772, 8030455, 8030458, 8122983, 8122990, 8289326, 8297221, 8313525, 8326570, 8326572, 8361200, 8472134, 8506837, 8559544, 8592891, 8679869, 8728871, 8796859, 8948101, 8985560, 9041086, 9102788, 9141405, 9177670, 9192341, 9240188, 9282330, 9282342, 9294599, 9341164, 9351839, 9351840, 9495897, 9529685, 9622813, 9634315, 9682510, 9682512, 9682513, 9690266, 9792211, 9792214, 9792217, 9792219, 9792220, 9792222, 9792236 ; Device Reference Number AA6114, GTIN 3600040127377, Lot Numbers 7174266, 7238934, 7238937, 7290670, 7303277, 7328602, 7328604, 7362452, 7387284, 7396140, 7396147, 7441024, 7680769, 7735761, 7782360, 7782371, 7794999, 7795001, 7942148, 7942154, 7969037, 7998909, 8040737, 8040751, 8098020, 8289321, 8361123, 8361133, 8408149, 8418184, 8442278, 8592968, 8620070, 8679825, 8679833, 8679834, 8728885, 8853718, 8884537, 8911244, 8923138, 8948148, 8975016, 8985570, 8985571, 8985589, 9031124, 9031133, 9102804, 9122861, 9226662, 9226666, 9248080, 9351817, 9385277, 9419405, 9419408, 9453470, 9460169, 9485166, 9529697, 9558587, 9558600, 9585508, 9585518, 9597173, 9597181, 9610603, 9610606, 9700656, 9700662, 9737166, 9748487, 9748831, 9748832, 9755424, 9755427, 9755429, 9755439, 9780645, 9780646, 9780648, 9780661, 9780679, 9780683, 9792890, 9792891, 9792893 ; Device Reference Number AA6116, GTIN 3600040127506, Lot Numbers 7193434, 7255904, 7314786, 7328608, 7328617, 7328618, 7406746, 7416574, 7447516, 7447523, 7487323, 7487328, 7494334, 7607523, 7680801, 7714102, 7794958, 7810306, 7883541, 7883545, 7942168, 8030469, 8087140, 8087162, 8111962, 8200198, 8268459, 8350405, 8361178, 8361186, 8418190, 8431498, 8442273, 8470200, 8472145, 8519842, 8529333, 8559608, 8572962, 8679849, 8679855, 8691267, 8728891, 8829399, 8829402, 8865298, 8884578, 8923523, 8948167, 8997529, 9008469, 9055014, 9077729, 9112123, 9141408, 9201829, 9240174, 9261058, 9261059, 9282296, 9294589, 9316028, 9341175, 9404784, 9429218, 9445790, 9453593, 9478195, 9495829, 9529688, 9540968, 9553558, 9558616, 9558618, 9585503, 9585504, 9610582, 9610598, 9622842, 9634401, 9634405, 9634423, 9634433, 9656837, 9666685, 9700700, 9700701, 9755483, 9755497 ; Device Reference Number AA6118, GTIN 3600040127612, Lot Numbers 7161607, 7229862, 7278686, 7290713, 7350612, 7607542, 7657482, 7657485, 7735781, 7735785, 7794982, 7794987, 7836019, 7919589, 7919593, 7919604, 7948649, 7998889, 7998893, 8244807, 8244816, 8255400, 8289336, 8341241, 8460199, 8559580, 8572905, 8700341, 8742738, 8742739, 8749952, 8829462, 8948269, 9008685, 9008690, 9041121, 9067845, 9090781, 9102910, 9122821, 9143141, 9261278, 9261283, 9261293, 9304728, 9304733, 9362306, 9362321, 9472350, 9472368, 9559982, 9585591, 9585598 ; Device Reference Number AA6120, GTIN 3600040127742, Lot Numbers 7174152, 7314709, 7314711, 7350584, 7362408, 7597839, 7607515, 7774397, 7883536, 7883538, 7942098, 7948628, 8350320, 8350322, 8442250, 8884440, 8985523, 9041132, 9201860, 9429193, 9429194, 9460760, 9460768, 9597150, 9711497, 9792240, 9792243 ; Device Reference Number AA6122, GTIN 3600040127872, Lot Numbers 7405871, 7416497, 7748381, 7942074, 8350342, 8493250, 8519854, 8592989, 8728906, 8985443, 9159829, 9404773, 9445789, 9453600, 9453601, 9485195, 9829431 ; Device Reference Number AA6124, GTIN 3600040128008, Lot Numbers 7416543, 7516929, 7773792, 7850835, 8442227, 8442228, 8442229, 9008561, 9067848, 9656792 ; Device Reference Number AA6308, GTIN 3600040128268, Lot Numbers 7210663, 7440856, 7447457, 7470016, 7516834, 7528782, 7703710, 7761577, 7782071, 7810060, 7883513, 7919523, 7919524, 7977718, 7977720, 8111946, 8111947, 8268401, 8268407, 8326502, 8341188, 8452349, 8559594, 8572917, 8679560, 8817139, 8829340, 8839643, 9090446, 9122812, 9226586, 9240167, 9351797, 9351798, 9429188, 9485153, 9495821, 9634306, 9656800, 9682505, 9690261, 9940979, 9940980, 9940981, 9940982 ; Device Reference Number AA6310, GTIN 3600040128398, Lot Numbers 7224527, 7270676, 7326836, 7503928, 7564546, 7575984, 7629249, 7669034, 7694121, 7782053, 7825239, 7850714, 7850715, 7874609, 7894935, 7942054, 7977212, 8008229, 8229393, 8268433, 8313522, 8350372, 8392310, 8392311, 8442257, 8484903, 8572971, 8572972, 8742701, 8742702, 8806409, 8865389, 8884593, 8996692, 9213520, 9304403, 9351788, 9404767, 9495819, 9505078, 9622811, 9634390, 9778578, 9940983, 9940985, 9940986, 9940987, 9940988 ; Device Reference Number AA6312, GTIN 3600040128527, Lot Numbers 7174140, 7187848, 7187849, 7210666, 7278643, 7362305, 7396116, 7447574, 7540089, 7540090, 7585014, 7748371, 7748372, 7748373, 7894950, 7894951, 7894952, 8009485, 8009487, 8149941, 8149944, 8326518, 8392430, 8620093, 8700281, 8710124, 8710125, 8765639, 8765640, 8765642, 8839922, 8854150, 8865380, 8865381, 8931454, 8979454, 8979455, 9008570, 9008577, 9031600, 9077808, 9077809, 9112304, 9112305, 9112307, 9226831, 9226834, 9351873, 9351874, 9351875, 9385381, 9405093, 9420757, 9472413, 9472414, 9530032, 9541014, 9634457, 9634458, 9634459 ; Device Reference Number AA6314, GTIN 3600040128657, Lot Numbers 7161409, 7199153, 7210671, 7270715, 7303239, 7328572, 7387266, 7597841, 7657501, 7748382, 7810282, 7810284, 7810285, 7883527, 7883528, 7883530, 7942127, 8021005, 8021008, 8021010, 8133760, 8133762, 8210455, 8268417, 8326563, 8361161, 8392440, 8431470, 8442240, 8506820, 8506824, 8506827, 8506829, 8592983, 8691234, 8781858, 8948130, 8996706, 9067761, 9090475, 9142760, 9192349, 9192350, 9315900, 9351810, 9351811, 9374199, 9396686, 9404795, 9405052, 9405054, 9405055, 9405056, 9419450, 9419451, 9429239, 9429242, 9429244, 9429245, 9445846, 9498282, 9498284, 9498291, 9520729, 9583957, 9583958, 9591036, 9683091, 9683093, 9683095 ; Device Reference Number AA6316, GTIN 3600040128787, Lot Numbers 7270691, 7270693, 7270695, 7278620, 7338694, 7372033, 7372058, 7416502, 7440888, 7469922, 7502564, 7504465, 7516897, 7531041, 7543264, 7657521, 7657840, 7810293, 7810295, 7810298, 7874664, 7919607, 7919615, 7948692, 8009509, 8087237, 8156954, 8156959, 8229360, 8229362, 8279953, 8279964, 8289295, 8385832, 8431464, 8442233, 8442236, 8459903, 8506902, 8506911, 8592930, 8620035, 8710056, 8749908, 8765594, 8796808, 8829392, 8884465, 9018628, 9031089, 9031090, 9159876, 9177676, 9201834, 9248070, 9304424, 9304425, 9341146, 9341147, 9351832, 9351834, 9362253, 9362254, 9396699, 9396700, 9405074, 9405076, 9405078, 9405080, 9405082, 9420743, 9420747, 9429258, 9429259, 9429260, 9429264, 9446036, 9446037, 9498317, 9498319, 9498321, 9498325, 9498328, 9553608, 9583682, 9583685, 9583686, 9583951, 9597332, 9610632, 9666715, 9666718, 9666722, 9690346, 9711527, 9711528, 9780661 ; Device Reference Number AA6318, GTIN 3600040128916, Lot Numbers 7187853, 7187855, 7187856, 7210690, 7314706, 7338701, 7387255, 7441009, 7494299, 7540104, 7624498, 7624499, 7774405, 7774409, 7774410, 7836009, 7836010, 7873737, 7874459, 7883405, 7894833, 7909021, 7919584, 7919587, 8063418, 8149978, 8210437, 8326549, 8452921, 8519805, 8519806, 8592976, 8656364, 8710067, 8710071, 8884562, 8899156, 8899157, 8931469, 8948191, 8948192, 9018829, 9055044, 9112167, 9192342, 9248025, 9261042, 9341126, 9341127, 9351828, 9351829, 9351830, 9374215, 9405067, 9419447, 9429238, 9445816, 9472419, 9472424, 9472428, 9505200, 9505205, 9505210, 9597325, 9683098 ; Device Reference Number AA6320, GTIN 3600040129043, Lot Numbers 7470031, 7564265, 7564548, 7694153, 7748259, 7859432, 7859433, 7968997, 7968998, 8111929, 8111931, 8350380, 8408104, 8529305, 8620018, 8620019, 8765651, 8796853, 9102809, 9159909, 9177740, 9261105, 9461967, 9634310, 9656805 ; Device Reference Number AA6322, GTIN 3600040129173, Lot Numbers 7278629, 7427155, 7669083, 7883400, 8020995, 8341192, 8529360, 8806425, 8865242, 9077741, 9261332, 9529711 ; Device Reference Number AA6324, GTIN 3600040129302, Lot Numbers 7576086, 8781856 ; Device Reference Number AA6408, GTIN 3600040129548, Lot Numbers 7161423, 7255852, 7314698, 7338705, 7487153, 7494196, 7503941, 7564628, 7859393, 7919509, 8122919, 8296919, 8313327, 8473741, 8498208, 8559595, 8710047, 8763704, 8829389, 8865402, 9520259, 9529677, 9622807, 9634305 ; Device Reference Number AA6410, GTIN 3600040129654, Lot Numbers 7238889, 7328578, 7504537, 7727224, 7875897, 7968423, 8268274, 8279805, 8289260, 8313332, 8385000, 8572855, 8829332, 8911104, 9041019, 9090452, 9112040, 9122839, 9304740, 9351786, 9404766, 9485150, 9585499, 9597191, 9610626, 9622874, 9645254 ; Device Reference Number AA6412, GTIN 3600040129760, Lot Numbers 7469902, 7487192, 7575986, 7616927, 7657453, 7831467, 7850833, 7894943, 7969018, 8063393, 8484904, 8763741, 8865370, 8996597, 9540949, 9597147 ; Device Reference Number AA6414, GTIN 3600040129876, Lot Numbers 7188119, 7199212, 7440901, 7528790, 7576075, 7714004, 7883386, 7987202, 8157026, 8418146, 8506830, 8620089, 8829335, 8884416, 8931371, 9041046, 9404761, 9634304 ; Device Reference Number AA6416, GTIN 3600040129982, Lot Numbers 7350530, 7447481, 7564676, 7942073, 8030404, 8279949, 8289288, 8341195, 8392320, 8442223, 8506815, 8620005, 8691165, 8765660, 8806426, 8884587, 8911347, 8996693, 9112189, 9404768, 9656818, 9690317 ; Device Reference Number AA6418, GTIN 3600040130094, Lot Numbers 7504609, 7774381, 8728852, 8806464, 8839747, 8865340, 8931610, 9018807, 9315892, 9341121, 9405065, 9711524 ; Device Reference Number AA6420, GTIN 3600040130209, Lot Numbers 7405944, 7528861, 7748339, 8097883, 8174966, 8200073, 8221042 ; Device Reference Number AA6422, GTIN 3600040130315, Lot Numbers 7387180, 7494256, 7504533, 7835619, 7998860, 8072753, 8122937, 8289258, 8361032, 8572867, 8724746, 8853703, 8923157, 8931625, 9018923, 9253805, 9315714, 9420936, 9530112, 9683051 ; Device Reference Number AA6424, GTIN 3600040130421, Lot Numbers 7908254, 8097924, 8111874, 8539720, 8700221, 8865324, 9520261, 9711496 ; Device Reference Number AA6C18, GTIN 3600040132005, Lot Numbers 7576074, 7735670, 7835647, 8529292, 8763748, 8865410, 9090707, 9177753 ; Device Reference Number AA6C20, GTIN 3600040132135, Lot Numbers 7362291, 7405869, 7714003, 8679900 ; Device Reference Number AA6C22, GTIN 3600040132265, Lot Numbers 8055955, 8114931, 8114941, 8114965, 9690288 ; Device Reference Number AA6C24, GTIN 3600040132395, Lot Numbers 8114942, 8498233, 8603437, 8865331, 9132275, 9558659 ;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Coloplast Manufacturing US, LLC
- Reason for Recall:
- A possible sterility issue was detected in Coloplast's facility on some Coloplast products.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Folysil Silicone Foley Catheters, Device Reference Numbers AA6106, AA6108, AA6110, AA6112, AA6114, AA6116, AA6118, AA6120, AA6122, AA6124, AA6308, AA6310, AA6312, AA6314, AA6316, AA6318, AA6320, AA6322, AA6324, AA6408, AA6410, AA6412, AA6414, AA6416, AA6418, AA6420, AA6422, AA6424, AA6C18, AA6C20, AA6C22, and AA6C24.
Product Codes/Lot Numbers:
Device Reference Number AA6106, GTIN 3600040126851, Lot Numbers 7278611, 7326825, 7469914, 7528770, 7540052, 7616917, 7680676, 7726724, 7748128, 7794789, 7874610, 7883406, 7948568, 7987184, 7987186, 7998885, 8097945, 8097947, 8097949, 8097950, 8149951, 8149954, 8156925, 8326511, 8326512, 8392428, 8452914, 8452915, 8519853, 8559628, 8572973, 8572977, 8728905, 8742746, 8796890, 8829455, 8899165, 8911320, 8996565, 9139298, 9177631, 9177632, 9192348, 9294521, 9294523, 9294524, 9315917, 9374195, 9404781, 9404782, 9419436, 9429227, 9429228, 9445801, 9453606, 9700652, 9711505, 9778601, 9778603, 9778606, 9792251, 9792257, 9817387, 9829471, 9843572, 9843573, 9940969 ; Device Reference Number AA6108, GTIN 3600040126981, Lot Numbers 7278611, 7326825, 7469914, 7528770, 7540052, 7616917, 7680676, 7726724, 7748128, 7794789, 7874610, 7883406, 7948568, 7987184, 7987186, 7998885, 8097945, 8097947, 8097949, 8097950, 8149951, 8149954, 8156925, 8326511, 8326512, 8392428, 8452914, 8452915, 8519853, 8559628, 8572973, 8572977, 8728905, 8742746, 8796890, 8829455, 8899165, 8911320, 8996565, 9139298, 9177631, 9177632, 9192348, 9294521, 9294523, 9294524, 9315917, 9374195, 9404781, 9404782, 9419436, 9429227, 9429228, 9445801, 9453606, 9700652, 9711505, 9778601, 9778603, 9778606, 9792251, 9792257, 9817387, 9829471, 9843572, 9843573, 9940969; Device Reference Number AA6110, GTIN 3600040127117, Lot Numbers 7255806, 7270683, 7270685, 7278618, 7303190, 7427087, 7440872, 7469982, 7487187, 7503933, 7564542, 7596635, 7629271, 7703729, 7713958, 7774417, 7774419, 7794951, 8133792, 8156966, 8156972, 8297224, 8313405, 8350384, 8519861, 8539757, 8679575, 8781871, 8796821, 8839854, 8839855, 8985535, 8996599, 9008443, 9031094, 9067753, 9077697, 9090464, 9102793, 9213533, 9213536, 9240194, 9240195, 9240196, 9240197, 9248068, 9261075, 9282360, 9282362, 9294615, 9294624, 9304466, 9315932, 9315933, 9341162, 9351850, 9351851, 9374230, 9385336, 9396721, 9405092, 9420752, 9420753, 9429249, 9445822, 9445823, 9472404, 9472405, 9553577, 9553578, 9610617, 9634462, 9634464, 9634465, 9656853, 9656854, 9656855, 9656856, 9656857, 9666697, 9666698, 9683046, 9683047, 9690316, 9700707, 9711520, 9792840, 9792841, 9843580, 9843581 ; Device Reference Number AA6112, GTIN 3600040127247, Lot Numbers 7174166, 7174170, 7174172, 7338746, 7338747, 7338751, 7470039, 7576132, 7576144, 7585050, 7585053, 7607531, 7670529, 7680781, 7894977, 7929763, 7929766, 7929768, 7929772, 8030455, 8030458, 8122983, 8122990, 8289326, 8297221, 8313525, 8326570, 8326572, 8361200, 8472134, 8506837, 8559544, 8592891, 8679869, 8728871, 8796859, 8948101, 8985560, 9041086, 9102788, 9141405, 9177670, 9192341, 9240188, 9282330, 9282342, 9294599, 9341164, 9351839, 9351840, 9495897, 9529685, 9622813, 9634315, 9682510, 9682512, 9682513, 9690266, 9792211, 9792214, 9792217, 9792219, 9792220, 9792222, 9792236 ; Device Reference Number AA6114, GTIN 3600040127377, Lot Numbers 7174266, 7238934, 7238937, 7290670, 7303277, 7328602, 7328604, 7362452, 7387284, 7396140, 7396147, 7441024, 7680769, 7735761, 7782360, 7782371, 7794999, 7795001, 7942148, 7942154, 7969037, 7998909, 8040737, 8040751, 8098020, 8289321, 8361123, 8361133, 8408149, 8418184, 8442278, 8592968, 8620070, 8679825, 8679833, 8679834, 8728885, 8853718, 8884537, 8911244, 8923138, 8948148, 8975016, 8985570, 8985571, 8985589, 9031124, 9031133, 9102804, 9122861, 9226662, 9226666, 9248080, 9351817, 9385277, 9419405, 9419408, 9453470, 9460169, 9485166, 9529697, 9558587, 9558600, 9585508, 9585518, 9597173, 9597181, 9610603, 9610606, 9700656, 9700662, 9737166, 9748487, 9748831, 9748832, 9755424, 9755427, 9755429, 9755439, 9780645, 9780646, 9780648, 9780661, 9780679, 9780683, 9792890, 9792891, 9792893 ; Device Reference Number AA6116, GTIN 3600040127506, Lot Numbers 7193434, 7255904, 7314786, 7328608, 7328617, 7328618, 7406746, 7416574, 7447516, 7447523, 7487323, 7487328, 7494334, 7607523, 7680801, 7714102, 7794958, 7810306, 7883541, 7883545, 7942168, 8030469, 8087140, 8087162, 8111962, 8200198, 8268459, 8350405, 8361178, 8361186, 8418190, 8431498, 8442273, 8470200, 8472145, 8519842, 8529333, 8559608, 8572962, 8679849, 8679855, 8691267, 8728891, 8829399, 8829402, 8865298, 8884578, 8923523, 8948167, 8997529, 9008469, 9055014, 9077729, 9112123, 9141408, 9201829, 9240174, 9261058, 9261059, 9282296, 9294589, 9316028, 9341175, 9404784, 9429218, 9445790, 9453593, 9478195, 9495829, 9529688, 9540968, 9553558, 9558616, 9558618, 9585503, 9585504, 9610582, 9610598, 9622842, 9634401, 9634405, 9634423, 9634433, 9656837, 9666685, 9700700, 9700701, 9755483, 9755497 ; Device Reference Number AA6118, GTIN 3600040127612, Lot Numbers 7161607, 7229862, 7278686, 7290713, 7350612, 7607542, 7657482, 7657485, 7735781, 7735785, 7794982, 7794987, 7836019, 7919589, 7919593, 7919604, 7948649, 7998889, 7998893, 8244807, 8244816, 8255400, 8289336, 8341241, 8460199, 8559580, 8572905, 8700341, 8742738, 8742739, 8749952, 8829462, 8948269, 9008685, 9008690, 9041121, 9067845, 9090781, 9102910, 9122821, 9143141, 9261278, 9261283, 9261293, 9304728, 9304733, 9362306, 9362321, 9472350, 9472368, 9559982, 9585591, 9585598 ; Device Reference Number AA6120, GTIN 3600040127742, Lot Numbers 7174152, 7314709, 7314711, 7350584, 7362408, 7597839, 7607515, 7774397, 7883536, 7883538, 7942098, 7948628, 8350320, 8350322, 8442250, 8884440, 8985523, 9041132, 9201860, 9429193, 9429194, 9460760, 9460768, 9597150, 9711497, 9792240, 9792243 ; Device Reference Number AA6122, GTIN 3600040127872, Lot Numbers 7405871, 7416497, 7748381, 7942074, 8350342, 8493250, 8519854, 8592989, 8728906, 8985443, 9159829, 9404773, 9445789, 9453600, 9453601, 9485195, 9829431 ; Device Reference Number AA6124, GTIN 3600040128008, Lot Numbers 7416543, 7516929, 7773792, 7850835, 8442227, 8442228, 8442229, 9008561, 9067848, 9656792 ; Device Reference Number AA6308, GTIN 3600040128268, Lot Numbers 7210663, 7440856, 7447457, 7470016, 7516834, 7528782, 7703710, 7761577, 7782071, 7810060, 7883513, 7919523, 7919524, 7977718, 7977720, 8111946, 8111947, 8268401, 8268407, 8326502, 8341188, 8452349, 8559594, 8572917, 8679560, 8817139, 8829340, 8839643, 9090446, 9122812, 9226586, 9240167, 9351797, 9351798, 9429188, 9485153, 9495821, 9634306, 9656800, 9682505, 9690261, 9940979, 9940980, 9940981, 9940982 ; Device Reference Number AA6310, GTIN 3600040128398, Lot Numbers 7224527, 7270676, 7326836, 7503928, 7564546, 7575984, 7629249, 7669034, 7694121, 7782053, 7825239, 7850714, 7850715, 7874609, 7894935, 7942054, 7977212, 8008229, 8229393, 8268433, 8313522, 8350372, 8392310, 8392311, 8442257, 8484903, 8572971, 8572972, 8742701, 8742702, 8806409, 8865389, 8884593, 8996692, 9213520, 9304403, 9351788, 9404767, 9495819, 9505078, 9622811, 9634390, 9778578, 9940983, 9940985, 9940986, 9940987, 9940988 ; Device Reference Number AA6312, GTIN 3600040128527, Lot Numbers 7174140, 7187848, 7187849, 7210666, 7278643, 7362305, 7396116, 7447574, 7540089, 7540090, 7585014, 7748371, 7748372, 7748373, 7894950, 7894951, 7894952, 8009485, 8009487, 8149941, 8149944, 8326518, 8392430, 8620093, 8700281, 8710124, 8710125, 8765639, 8765640, 8765642, 8839922, 8854150, 8865380, 8865381, 8931454, 8979454, 8979455, 9008570, 9008577, 9031600, 9077808, 9077809, 9112304, 9112305, 9112307, 9226831, 9226834, 9351873, 9351874, 9351875, 9385381, 9405093, 9420757, 9472413, 9472414, 9530032, 9541014, 9634457, 9634458, 9634459 ; Device Reference Number AA6314, GTIN 3600040128657, Lot Numbers 7161409, 7199153, 7210671, 7270715, 7303239, 7328572, 7387266, 7597841, 7657501, 7748382, 7810282, 7810284, 7810285, 7883527, 7883528, 7883530, 7942127, 8021005, 8021008, 8021010, 8133760, 8133762, 8210455, 8268417, 8326563, 8361161, 8392440, 8431470, 8442240, 8506820, 8506824, 8506827, 8506829, 8592983, 8691234, 8781858, 8948130, 8996706, 9067761, 9090475, 9142760, 9192349, 9192350, 9315900, 9351810, 9351811, 9374199, 9396686, 9404795, 9405052, 9405054, 9405055, 9405056, 9419450, 9419451, 9429239, 9429242, 9429244, 9429245, 9445846, 9498282, 9498284, 9498291, 9520729, 9583957, 9583958, 9591036, 9683091, 9683093, 9683095 ; Device Reference Number AA6316, GTIN 3600040128787, Lot Numbers 7270691, 7270693, 7270695, 7278620, 7338694, 7372033, 7372058, 7416502, 7440888, 7469922, 7502564, 7504465, 7516897, 7531041, 7543264, 7657521, 7657840, 7810293, 7810295, 7810298, 7874664, 7919607, 7919615, 7948692, 8009509, 8087237, 8156954, 8156959, 8229360, 8229362, 8279953, 8279964, 8289295, 8385832, 8431464, 8442233, 8442236, 8459903, 8506902, 8506911, 8592930, 8620035, 8710056, 8749908, 8765594, 8796808, 8829392, 8884465, 9018628, 9031089, 9031090, 9159876, 9177676, 9201834, 9248070, 9304424, 9304425, 9341146, 9341147, 9351832, 9351834, 9362253, 9362254, 9396699, 9396700, 9405074, 9405076, 9405078, 9405080, 9405082, 9420743, 9420747, 9429258, 9429259, 9429260, 9429264, 9446036, 9446037, 9498317, 9498319, 9498321, 9498325, 9498328, 9553608, 9583682, 9583685, 9583686, 9583951, 9597332, 9610632, 9666715, 9666718, 9666722, 9690346, 9711527, 9711528, 9780661 ; Device Reference Number AA6318, GTIN 3600040128916, Lot Numbers 7187853, 7187855, 7187856, 7210690, 7314706, 7338701, 7387255, 7441009, 7494299, 7540104, 7624498, 7624499, 7774405, 7774409, 7774410, 7836009, 7836010, 7873737, 7874459, 7883405, 7894833, 7909021, 7919584, 7919587, 8063418, 8149978, 8210437, 8326549, 8452921, 8519805, 8519806, 8592976, 8656364, 8710067, 8710071, 8884562, 8899156, 8899157, 8931469, 8948191, 8948192, 9018829, 9055044, 9112167, 9192342, 9248025, 9261042, 9341126, 9341127, 9351828, 9351829, 9351830, 9374215, 9405067, 9419447, 9429238, 9445816, 9472419, 9472424, 9472428, 9505200, 9505205, 9505210, 9597325, 9683098 ; Device Reference Number AA6320, GTIN 3600040129043, Lot Numbers 7470031, 7564265, 7564548, 7694153, 7748259, 7859432, 7859433, 7968997, 7968998, 8111929, 8111931, 8350380, 8408104, 8529305, 8620018, 8620019, 8765651, 8796853, 9102809, 9159909, 9177740, 9261105, 9461967, 9634310, 9656805 ; Device Reference Number AA6322, GTIN 3600040129173, Lot Numbers 7278629, 7427155, 7669083, 7883400, 8020995, 8341192, 8529360, 8806425, 8865242, 9077741, 9261332, 9529711 ; Device Reference Number AA6324, GTIN 3600040129302, Lot Numbers 7576086, 8781856 ; Device Reference Number AA6408, GTIN 3600040129548, Lot Numbers 7161423, 7255852, 7314698, 7338705, 7487153, 7494196, 7503941, 7564628, 7859393, 7919509, 8122919, 8296919, 8313327, 8473741, 8498208, 8559595, 8710047, 8763704, 8829389, 8865402, 9520259, 9529677, 9622807, 9634305 ; Device Reference Number AA6410, GTIN 3600040129654, Lot Numbers 7238889, 7328578, 7504537, 7727224, 7875897, 7968423, 8268274, 8279805, 8289260, 8313332, 8385000, 8572855, 8829332, 8911104, 9041019, 9090452, 9112040, 9122839, 9304740, 9351786, 9404766, 9485150, 9585499, 9597191, 9610626, 9622874, 9645254 ; Device Reference Number AA6412, GTIN 3600040129760, Lot Numbers 7469902, 7487192, 7575986, 7616927, 7657453, 7831467, 7850833, 7894943, 7969018, 8063393, 8484904, 8763741, 8865370, 8996597, 9540949, 9597147 ; Device Reference Number AA6414, GTIN 3600040129876, Lot Numbers 7188119, 7199212, 7440901, 7528790, 7576075, 7714004, 7883386, 7987202, 8157026, 8418146, 8506830, 8620089, 8829335, 8884416, 8931371, 9041046, 9404761, 9634304 ; Device Reference Number AA6416, GTIN 3600040129982, Lot Numbers 7350530, 7447481, 7564676, 7942073, 8030404, 8279949, 8289288, 8341195, 8392320, 8442223, 8506815, 8620005, 8691165, 8765660, 8806426, 8884587, 8911347, 8996693, 9112189, 9404768, 9656818, 9690317 ; Device Reference Number AA6418, GTIN 3600040130094, Lot Numbers 7504609, 7774381, 8728852, 8806464, 8839747, 8865340, 8931610, 9018807, 9315892, 9341121, 9405065, 9711524 ; Device Reference Number AA6420, GTIN 3600040130209, Lot Numbers 7405944, 7528861, 7748339, 8097883, 8174966, 8200073, 8221042 ; Device Reference Number AA6422, GTIN 3600040130315, Lot Numbers 7387180, 7494256, 7504533, 7835619, 7998860, 8072753, 8122937, 8289258, 8361032, 8572867, 8724746, 8853703, 8923157, 8931625, 9018923, 9253805, 9315714, 9420936, 9530112, 9683051 ; Device Reference Number AA6424, GTIN 3600040130421, Lot Numbers 7908254, 8097924, 8111874, 8539720, 8700221, 8865324, 9520261, 9711496 ; Device Reference Number AA6C18, GTIN 3600040132005, Lot Numbers 7576074, 7735670, 7835647, 8529292, 8763748, 8865410, 9090707, 9177753 ; Device Reference Number AA6C20, GTIN 3600040132135, Lot Numbers 7362291, 7405869, 7714003, 8679900 ; Device Reference Number AA6C22, GTIN 3600040132265, Lot Numbers 8055955, 8114931, 8114941, 8114965, 9690288 ; Device Reference Number AA6C24, GTIN 3600040132395, Lot Numbers 8114942, 8498233, 8603437, 8865331, 9132275, 9558659 ;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0844-2025
Related Recalls
A possible sterility issue was detected in Coloplast's facility on some Coloplast products.
Ureteral dilator, REF RBD014
Coloplast Manufacturing US
A possible sterility issue (e.g. a breach in sterile packaging) was detected in Coloplast's facility on some ureteral dilator products. This issue on the ureteral dilator packaging has been identified during testing in our facility.
Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.