Hillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with Warming Blankets and Mattresses.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI 00850011479582, Serial Numbers: 23010166, 23010167, 23010168, 23010169, 23010170, 23010171, 23010172, 23010173, 23010174, 23010175, 23010176, 23010177, 23010178, 23010179, 23010180, 23010181, 23010182, 23010183, 23010184, 23010185, 24020001, 24020002, 24020003, 24020004, 24020005, 24020006, 24030001, 24030002, 24030003, 24030004, 24030005, 24030021, 24030022, 24030023, 24030024, 24030025, 24030026, 24030027, 24030028, 24030029, 24030030, 24030016, 24030017, 24030018, 24030019, 24030020, 22120033, 24030011, 24030012, 24030013, 24030014, 24030015, 24030006, 24030007, 24030008, 24030009, 24030010, 24060001, 24060002, 24060003, 24060004, 24060005, 24060006, 24060007, 24060008, 24060009, 24060010, 24060024, 24060025, 24060026, 24060027, 24060028, 24060011, 24060012, 24060013, 24060014, 24060015.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Augustine Temperature Management, LLC
Reason for Recall:
HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
completed

Product Information

Full Description:

Hillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with Warming Blankets and Mattresses.

Product Codes/Lot Numbers:

UDI-DI 00850011479582, Serial Numbers: 23010166, 23010167, 23010168, 23010169, 23010170, 23010171, 23010172, 23010173, 23010174, 23010175, 23010176, 23010177, 23010178, 23010179, 23010180, 23010181, 23010182, 23010183, 23010184, 23010185, 24020001, 24020002, 24020003, 24020004, 24020005, 24020006, 24030001, 24030002, 24030003, 24030004, 24030005, 24030021, 24030022, 24030023, 24030024, 24030025, 24030026, 24030027, 24030028, 24030029, 24030030, 24030016, 24030017, 24030018, 24030019, 24030020, 22120033, 24030011, 24030012, 24030013, 24030014, 24030015, 24030006, 24030007, 24030008, 24030009, 24030010, 24060001, 24060002, 24060003, 24060004, 24060005, 24060006, 24060007, 24060008, 24060009, 24060010, 24060024, 24060025, 24060026, 24060027, 24060028, 24060011, 24060012, 24060013, 24060014, 24060015.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0841-2025

Related Recalls

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

Dec 16, 2024 Other Medical Devices Nationwide View Details →

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

Dec 16, 2024 Other Medical Devices Nationwide View Details →