Maquet ELS Cannula 15 FR. Used for extracorporeal membrane oxygenation (ECMO) for the treatment of respiratory insufficiency in infants.
Class I - DangerousWhat Should You Do?
- Check if you have this product: US Distribution Part number 701017471 Lot numbers 92046825 92065723 92076940 OUS Distribution Part Number Lot Numbers 701002797 92050616 701002797 92052823 701002797 92052913 701002797 92058153 701002797 92063518 701002797 92068260 701002797 92069267 701002797 92072177 701002797 92075413 701002797 92077862 701005092 92081862 701005092 92049129 701005092 92070554 701005093 92049130 701005093 92056899 701005093 92065588 701005093 92069429 701005093 92070553 701005093 92071852 701005093 92073598 701005093 92076875 701017471 92076940 701046644 92063587 701050978 92047882 701050978 92053930 701050978 92054156 701050978 92056171 701050978 92056804 701050978 92065094 701050978 92067382 701050978 92068224 701050978 92069286 701050978 92071806
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Maquet Cardiovascular Us Sales, Llc
- Reason for Recall:
- ELS 15 Fr Cannula could potentially contain an occluded arterial lumen.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Maquet ELS Cannula 15 FR. Used for extracorporeal membrane oxygenation (ECMO) for the treatment of respiratory insufficiency in infants.
Product Codes/Lot Numbers:
US Distribution Part number 701017471 Lot numbers 92046825 92065723 92076940 OUS Distribution Part Number Lot Numbers 701002797 92050616 701002797 92052823 701002797 92052913 701002797 92058153 701002797 92063518 701002797 92068260 701002797 92069267 701002797 92072177 701002797 92075413 701002797 92077862 701005092 92081862 701005092 92049129 701005092 92070554 701005093 92049130 701005093 92056899 701005093 92065588 701005093 92069429 701005093 92070553 701005093 92071852 701005093 92073598 701005093 92076875 701017471 92076940 701046644 92063587 701050978 92047882 701050978 92053930 701050978 92054156 701050978 92056171 701050978 92056804 701050978 92065094 701050978 92067382 701050978 92068224 701050978 92069286 701050978 92071806
Distribution:
Distributed in: AZ, CO, FL, TN, TX, UK
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0839-2013
Related Recalls
HLS Set Advanced Model Name: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code/Part Number: 70105.2797 - Product Usage: These devices are used with the CARDIOHELP system in order to provide extracorporeal circulation for cardiac support and/or pulmonary support.
Maquet Cardiovascular Us Sales
The HLS Set Advanced (disposable for Cardiohelp) is being recalled due to breaches in the product's sterility barrier.
HLS Set Advanced, Model: BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Product Code/Part Number: 70105.2794 - Product Usage: These devices are used with the CARDIOHELP system in order to provide extracorporeal circulation for cardiac support and/or pulmonary support.
Maquet Cardiovascular Us Sales
The HLS Set Advanced (disposable for Cardiohelp) is being recalled due to breaches in the product's sterility barrier.
The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.