Naturalyte Liquid Acid concentrate bottles, Product Code 08-2301-3 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers: 20NXAC013 20NXAC035 Additional Affected lots as of 3/17/21: 20LXAC056 20LXAC041 20LXAC055 20LXAC057 20NXAC013 20NXAC035
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Fresenius Medical Care Holdings, Inc.
- Reason for Recall:
- The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Naturalyte Liquid Acid concentrate bottles, Product Code 08-2301-3 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Product Codes/Lot Numbers:
Lot Numbers: 20NXAC013 20NXAC035 Additional Affected lots as of 3/17/21: 20LXAC056 20LXAC041 20LXAC055 20LXAC057 20NXAC013 20NXAC035
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0835-2021
Related Recalls
Bicarby Dialysate; Model number: RFP-400-G;
Fresenius Medical Care Holdings
There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.
Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;
Fresenius Medical Care Holdings
Given the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.
DELFLEX, Dextrose Peritoneal Dialysis Solution with attached stay-safe Exchange Set for Intraperitoneal Dialysis Only, 2.5% DEX. LM/LC, 2L 5PK, Part Number 054-20222, Fresenius Medical Care North America, 920 Winter Street, Waltham, MA 02451.
Fresenius Medical Care Holdings
Lack of Assurance of Sterility: Potential leaks from perforations in bags.