Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) C-SECTION PACK, kit number FDCS01AA; b) C-SECTION PACK, kit number FDCS01AB; c) C-SECTION STERILE PACK, kit number HGCS68J; d) C-SECTION PACK, kit number HICS50H; e) C-SECTION PPS, kit number LLCI66; f) C-Section drape Pack, kit number RBCS12R; g) C-Section, kit number THCS49B; h) C-SECTION PACK (PS 020632), kit number UICS50AG;
Class I - DangerousWhat Should You Do?
- Check if you have this product: a) FDCS01AA, UDI/DI 00191072179890, Lot Numbers: 708221, exp. 10/23/2023; 646221, exp. 12/20/2023; b) FDCS01AB, UDI/DI 00191072185273, Lot Numbers: 886231, exp. 4/24/2024; c) HGCS68J, UDI/DI none, Lot Numbers: 702221, exp. 10/22/2023; 654221, exp. 12/12/2023; 960231, exp. 2/9/2024; 903231, exp. 4/7/2024; d) HICS50H, UDI/DI none, Lot Numbers: 909221, exp. 4/1/2023; 835221, exp. 6/14/2023; 755221, exp. 9/2/2023; e) LLCI66, UDI/DI 00191072119025, Lot Numbers: 750221, exp. 9/7/2023; 727221, exp. 9/30/2023; 678221, exp. 11/9/2023; 994231, exp. 11/9/2023; 948231, exp. 11/9/2023; 926231, exp. 11/9/2023; f) RBCS12R, UDI/DI 00191072131812, Lot Numbers: 770221, exp. 8/18/2023; 695221, exp. 11/1/2023; g) THCS49B, UDI/DI 00191072153357, Lot Numbers: 770221, exp. 8/18/2023; h) UICS50AG, UDI/DI none, Lot Numbers: 679221, exp. 11/17/2023; 946231, exp. 2/23/2024; 876231, exp. 5/4/2024;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- American Contract Systems, Inc.
- Reason for Recall:
- During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) C-SECTION PACK, kit number FDCS01AA; b) C-SECTION PACK, kit number FDCS01AB; c) C-SECTION STERILE PACK, kit number HGCS68J; d) C-SECTION PACK, kit number HICS50H; e) C-SECTION PPS, kit number LLCI66; f) C-Section drape Pack, kit number RBCS12R; g) C-Section, kit number THCS49B; h) C-SECTION PACK (PS 020632), kit number UICS50AG;
Product Codes/Lot Numbers:
a) FDCS01AA, UDI/DI 00191072179890, Lot Numbers: 708221, exp. 10/23/2023; 646221, exp. 12/20/2023; b) FDCS01AB, UDI/DI 00191072185273, Lot Numbers: 886231, exp. 4/24/2024; c) HGCS68J, UDI/DI none, Lot Numbers: 702221, exp. 10/22/2023; 654221, exp. 12/12/2023; 960231, exp. 2/9/2024; 903231, exp. 4/7/2024; d) HICS50H, UDI/DI none, Lot Numbers: 909221, exp. 4/1/2023; 835221, exp. 6/14/2023; 755221, exp. 9/2/2023; e) LLCI66, UDI/DI 00191072119025, Lot Numbers: 750221, exp. 9/7/2023; 727221, exp. 9/30/2023; 678221, exp. 11/9/2023; 994231, exp. 11/9/2023; 948231, exp. 11/9/2023; 926231, exp. 11/9/2023; f) RBCS12R, UDI/DI 00191072131812, Lot Numbers: 770221, exp. 8/18/2023; 695221, exp. 11/1/2023; g) THCS49B, UDI/DI 00191072153357, Lot Numbers: 770221, exp. 8/18/2023; h) UICS50AG, UDI/DI none, Lot Numbers: 679221, exp. 11/17/2023; 946231, exp. 2/23/2024; 876231, exp. 5/4/2024;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0830-2024
Related Recalls
Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ
American Contract Systems
Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.
Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.