Mevion S250 and S250i Proton Therapy Systems hand pendant (P/N 10263 or P/N 37808)

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All theryapy systems, all hand pendant serial numbers. Therapy system UDI 00864366000100 and UDI 00864366000124.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Mevion Medical Systems, Inc.
Reason for Recall:
Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other objects in the treatment room.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing