Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Catalog #s 8416400 and 8417400
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Draeger Medical, Inc.
Reason for Recall:
The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Product Codes/Lot Numbers:

Catalog #s 8416400 and 8417400

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0827-2016