T:slim Insulin Delivery System Insulin Infusion Pump
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers: M000857, M001414, M001454, M001963, M002028, M000869, M001415, M001455, M001964, M002029, M001344, M001416, M001456, M001973, M002030, M001345, M001417, M001457, M001974, M002082, M001346, M001420, M001458, M001979, M002083, M001347, M001421, M001459, M001980, M002096, M001389, M001422, M001460, M001987, M002097, M001390, M001423, M001528, M001988, M002099, M001391, M001451, M001529, M001990, M002100, M001392, M001452, M001530, M001991, M002119, M001393, M001453, M001532, M002027, M002120.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Tandem Diabetes Care Inc
- Reason for Recall:
- Specific lots of insulin cartridges that are used with the T:slim Insulin Pump may be at risk for leaking. A cartridge leak could result in the device delivering too much or too little insulin, which can lead to a serious adverse event.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
T:slim Insulin Delivery System Insulin Infusion Pump
Product Codes/Lot Numbers:
Lot Numbers: M000857, M001414, M001454, M001963, M002028, M000869, M001415, M001455, M001964, M002029, M001344, M001416, M001456, M001973, M002030, M001345, M001417, M001457, M001974, M002082, M001346, M001420, M001458, M001979, M002083, M001347, M001421, M001459, M001980, M002096, M001389, M001422, M001460, M001987, M002097, M001390, M001423, M001528, M001988, M002099, M001391, M001451, M001529, M001990, M002100, M001392, M001452, M001530, M001991, M002119, M001393, M001453, M001532, M002027, M002120.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0827-2014
Related Recalls
Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.
t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)
Tandem Diabetes Care
an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.