Custom procedural convenience kits and trays, general hospital use, labeled as: a) Baby PICC Kit, kit number UIBP90J; b) Baby PICC Kit, kit number UIBP90K; c) PORT PACK (PS 058330), kit number UIPO82AB; d) Pediatric PICC Kit, kit number UIPP76K;
Class I - DangerousWhat Should You Do?
- Check if you have this product: a) UIBP90J, UDI/DI 00191072117038, Lot Numbers: 996231, exp. 1/4/2024; b) UIBP90K, UDI/DI 00191072189356, Lot Numbers: 911231, exp. 7/26/2023; c) UIPO82AB, UDI/DI 00191072172006, Lot Numbers: 727221, exp. 9/30/2023; 680221, exp. 11/16/2023; d) UIPP76K, UDI/DI 00191072142849, Lot Numbers: 637221, exp. 12/29/2023; 976231, exp. 1/24/2024; 898231, exp. 4/12/2024
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- American Contract Systems, Inc.
- Reason for Recall:
- During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Custom procedural convenience kits and trays, general hospital use, labeled as: a) Baby PICC Kit, kit number UIBP90J; b) Baby PICC Kit, kit number UIBP90K; c) PORT PACK (PS 058330), kit number UIPO82AB; d) Pediatric PICC Kit, kit number UIPP76K;
Product Codes/Lot Numbers:
a) UIBP90J, UDI/DI 00191072117038, Lot Numbers: 996231, exp. 1/4/2024; b) UIBP90K, UDI/DI 00191072189356, Lot Numbers: 911231, exp. 7/26/2023; c) UIPO82AB, UDI/DI 00191072172006, Lot Numbers: 727221, exp. 9/30/2023; 680221, exp. 11/16/2023; d) UIPP76K, UDI/DI 00191072142849, Lot Numbers: 637221, exp. 12/29/2023; 976231, exp. 1/24/2024; 898231, exp. 4/12/2024
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0826-2024
Related Recalls
Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO PACK MHCY35AE CYSTO PACK- 212487 PLBY52E CYSTO PACK UDCY31Z CYSTO PACK - 206047 UICY28L MOR CYSTO PACK
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.