LIFEPAK 15 MONITOR/DEFIBRILLATOR that was either manufactured with or received an upgrade kit that contained an affected keypad.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial numbers: 37964613 37964617 37964621 37964628 37964630 37968424 37968429 37964596 37964601 37964607 37964610 37968715 37968716 37968717 37968718 37968721 37968722 37968724 38046540 38046549 38046552 38046565 38046571 38046572 38046573 38051949 38051951 38051954 38051955 38051958 38051963 38051969 38051972 38051973 38051974 38051980 38053216 38053218 38053220 38053221 38053225 38053231 38053233 38053236 38053239 38059568 38059570 38059574 38059591 38120769 38120772 38120775 38120780 38120781 38120783 38120784 38120787 38120788 38120790 38124016 38124030 38124035 38124042 38124044 38124047 38125969 38125971 38125972 38125980 38125982 38125987 38125990 38126255 38126258 38126260 38126265 38126267 38126269 38126272 38126273 38126277 38127130 38127131 38127132 38127134 38127136 38127138 38127139 38127140 38127141 38127143 38127144 38130272 38130273 38130276 38130277 38130279 38130282 38130283 38130285 38130286 38130287 38131588 38131589 38131592 38131597 38131600 38131602 38131603 38131605 38131607 38131608 38131756 38131757 38131760 38131761 38131762 38131764 38131765 38131767 38131773 38131775 38131776 38131777 38131842 38131845 38132370 38132951 38134512 38134524 38135205 38135212 38135215 38136507 38136508 38136513 38136516 38137137 38137143 38137149 38137155 38137168 38137174 38138844 38138846 38141005 38142743 38150694 38151893 38151898 38153123 38153124 38153136 38155919 38158080 38159137 38159181 38159183 38159187 38159197 38161892 38161893 38161898 38161899 38161902 38161907 38161909 38161912 38162531 38162561 38162597 38164391 38164403 38164420 38164448 38166449 38166461 38166462 38166475 38166479 38166485 38166486 38166497 38166507 38166513 38166516 38166541 38166542 38166543 38166544 38166549 38166550 38166555 38166558 38166569 38166571 38166581 38166582 38166586 38166594 38166610 38166614 38166628 38166632 38166634 38166642 38166647 38166674 38166675 38166676 38166677 38166678 38177270 38177272 38177275 38191219 38191235 38200435 38200436 38200441 38200443 38200444 38200446 38200447 38200448 38200450 38200451 38200452 38200453 38200454 38200457 38200458 38200461 38200462 38200464 38200465 38200466 38200467 38200469 38200470 38200471 38200472 38200474 38200476 38202790 38202791 38202802 38202803 38202809 38202821 38205398 38205404 38205407 38205410 38205413 38205781 38205783 38205784 38205794 38206256 38211038 38211041 38211042 38211043 38211044 38211051 38211059 38214239 38214242 38214247 38214259 38214266 38214271 38214279 38217387 38217390 38217400 38221456 38221457 38224395 38224601 38227461 38253472 38253478 38253768 38253769 38253770 38253786 38253790 38253798 38253805 38263304 38263329 38272217 38273859 38274178 38274396 38274508 38279197 38279198 38279199 38279202 38279206 38281774 38282979 38282983 38282987 38287524 38287525 38287530 38287845 38287846 38290137 38290140 38290141 38290535 38290536 38290538 38290539 38307505 38315311 38315330 38322516 38323219 38323233 38323263 38323311 38324196 38324198 38324199 38324201 38324210 38324213 38327183 38332345 38332349 38332355 38332358 38332363 38332748 38332749 38332753 38332754 38332761 38332940 38338635 38340566 38341984 38343138 38343162 38343235 38343241 38345048 38346560 38353024 38353025 38353033 38354810 38355627 38355629 38355631 38355637 38355638 38355639 38355645 38359027 38359559 38362860 38364966 38365244 38365249 38365937 38365940 38368303 38368304 38368307 38368327 38368329 38368616 38369586 38370274 38370275 38370277 38370280 38370282 38370293 38370425 38370431 38370433 38370434 38371200 38371201 38371218 38371701 38371702 38371708 38371709 38371711 38371714 38372527 38373618 38374119 38374686 38374688 38374696 38374701 38432103 38441912 38441916 38441918 38441920 38441931 38441934 38441935 38441936 38445594 38445597 38445604 38445626 38445640 38445649 38449384 38449391 38449400 38449410 38458698 38472562 38517179 38517227 38520270 38520276 38521672 38523568 38523584 38523596 38523602 38523604 38523606 38523607 38523610 38528452 38528458 38528471 38534441 38536323 38536351 38536916 38536922 38536929 38537159 38537164 38537168 38537704 38537706 38537758 38537759 38537768 38537774 38537776 38537779 38538715 38538719 38538721 38538879 38538881 38538926 38538931 38538932 38538935 38538940 38538953 38538956 38538957 38538969 38538983 38540917 38540923 38540935 38540937 38540940 38540944 38540948 38540955 38540956 38540959 38540960 38540961 38540964 38540967 38540971 38540972 38540974 38543095 38543100 38543102 38543108 38543111 38543112 38543113 38544618 38544622 38544623 38544625 38544627 38544635 38544637 38544638 38544646 38544649 38544658 38544672 38544673 38544675 38544763 38544764 38544771 38548002 38548004 38548012 38548022 38548037 38548038 38548040 38553528 38553529 38553537 38553545 38559431 38559446 38559447 38560013 38560014 38560018 38560019 38560020 38560021 38560030 38560031 38560032 38560033 38560034 38560035 38560036 38560037 38560038 38560039 38560040 38560041 38561014 38561021 38561022 38561023 38561024 38561027 38564049 38564051 38564679 38567823 38567825 38568460 38568463 38568468 38568473 38568874 38568875 38568876 38568877 38568878 38568879 38568882 38568890 38568891 38569281 38574882 38578213 38611410 38611411 38611413 38611414 38611509 38611513 38611516 38611517 38611518 38611520 38611521 38611522 38611523 38611524 38611526 38611529 38611530 38611531 38611533 38611536 38611537 38611553 38611554 38611556 38611557 38611559 38611561 38613942 38613943 38613945 38613949 38613950 38613952 38613954 38613967 38619552 38620457 38620459 38620466 38620477 38620479 38620480 38620481 38620495 38624129 38624139 38624141 38624145 38624149 38624150 38624151 38624153 38624159 38626953 38626956 38626958 38626963 38626967 38626991 38626995 38631721 38631723 38631725 38631730 38631746 38648317 38648319 38720330 38723976 38724387 38725753 38725771 38726757 38726761 38726794 38726796 38733842 38733852 38733858 38735077 38738213 38738217 38738219 38738222 38743122 38756198 38763408 38763413 38763423 38763427 38774131 38778622 38784168 38784951 38784969 38785815 38785821 38790361 38790364 38790367 38790373 38790381 38790383 38790389 38790398 38790399 38790403 38790407 38793515 38793516 38793529 38793539 38793540 38793546 38793552 38793555 38793559 38793560 38793562 38793563 38793565 38793566 38793569 38793575 38793580 38795890 38795898 38795906 38798710 38798718 38798795 38798801 38800559 38800568 38800581 38800586 38800587 38800588 38800594 38800595 38802775 38802783 38802785 38802787 38802789 38802791 38802797 38802823 38802828 38802846 38805559 38805572 38805575 38805589 38805592 38805596 38805602 38805605 38805609 38805611 38805612 38805616 38805619 38807481 38807487 38807488 38807491 38807492 38807493 38807494 38807496 38807498 38807501 38807502 38807508 38807509 38809007 38809009 38809010 38809015 38809017 38809019 38809020 38809022 38809025 38809035 38811712 38811713 38811714 38811717 38811720 38811731 38811732 38817539 38817544 38817554 38817555 38817557 38817558 38817559 38817560 38817562 38817563 38817564 38817772 38817776 38817777 38819691 38819700 38819712 38821747 38821771 38821775 38821787 38821792 38821799 38821800 38821801 38821803 38821804 38821805 38821807 38824055 38824080 38839958 38839964 38839966 38839972 38840725 38852928 38854322 38854328 38854329 38856558 38856571 38856576 38856580 38856589 38870845 38870852 38871061 38871062 38877850 38878667 38878682 38878686 38878688 38878692 38881890 38881891 38897292 38897298 38897493 38897495 38897499 38898618 38898619 38898622 38898623 38898625 38898629 38898631 38898634 38898636 38898640 38898647 38898651 38934844 38934845 38934848 38934850 38934868 38948322 38988236 38988242 38988247 38988251 38988256 38988257 38988258 38988259 38989256 38989260 38989261 38989264 38989265 38989267 38989269 38989275 38989336 38992044 38992045 38992061 38996375 38996376 38996377 38996378 38996379 38996380 38996381 38996382 38996383 38996386 38996389 38996390 38996392 38996393 38996395 38996397 38996398 38996408 38996792 38996793 38996794 38996799 38996807 38999803 38999805 38999809 38999810 38999812 39000250 39004380 39004384 39023652 39025293 39025303 39029063 39029067 39029068 39029073 39029143 39029147 39029161 39029168 39029176 39029200 39029205 39029206 39029215 39029218 39039179 39039192 39040605 39048208 39048209 39048488 39048504 39048506 39050458 39050475 39050492 39050508 39053292 39053297 39053316 39054574 39054577 39054578 39054582 39054583 39054584 39054586 39054589 39054593 39054598 39054602 39054603 39054605 39054606 39054607 39054622 39054634 39054636 39054642 39054646 39054648 39054651 39054654 39054666 39054673 39054702 39054719 39054729 39054736 39054741 39054749 39054751 39054759 39054766 39054773 39054787 39054788 39054809 39054830 39054833 39054834 39054835 39054836 39054842 39054850 39054852 39054854 39054857 39054861 39059003 39059006 39059474 39059477 39059479 39059482 39059485 39059486 39060195 39060196 39084702 39090471 39108559 39108565 39108570 39108571 39108572 39108585 39108586 39108589 39109072 39109078 39109080 39109081 39119994 39119995 39119996 39119997 39119998 39119999 39120000 39120001 39120002 39120003 39126759 39139844 39139853 39150420 39150421 39150423 39150430 39150865 39152709 39152710 39152711 39152713 39153383 39155791 39178106 39178108 39178109 39178110 39178111 39178112 39178113 39179615 39179708 39179732 39179733 39179734 39179735 39179736 39179737 39179738 39179739 39179740 39179742 39179743 39179744 39179745 39179746 39179747 39179748 39179749 37923203 37923205 37923208 37923213 37924152 37928794 37928964 37930942 37930944 37930964 37930966 37930977 37930990 37930993 37931003 37933241 37933244 37933249 37933250 37933262 37937842 37937843 37937848 37937849 37937851 37937854 37937855 37937856 37937862 37937863 37937864 37937866 37937868 37937870 37937871 37937872 37937873 37937874 37937885 37937886 37937887 37937889 37937891 37937892 37963295 37963300 37963302 37963323 37963324 37963325 37963328 37963926 37963928 37963935 37964552 37964554 37964557 37964560 37964564 37964567 37964575 37964585 37964587 37964588 39179750 39179751 39179752 39179754 39179755 39180764 39180766 39180767 39180768 39180770 39180771 39180773 39180774 39180775 39180776 39180779 39180780 39180783 39180784 39180785 39180787 39180788 39180789 39180790 39182063 39182066 39182070 39182076 39182077 39182078 39182080 39182081 39183493 39183494 39183495 39183496 39183497 39183498 39183499 39183500 39183501 39183502 39183640 39183641 39183642 39183643 39183644 39183645 39183646 39183648 39183649 39183650 39183651 39183652 39183655 39183657 39184048 39184050 39184200 39186322 39186323 39186325 39186326 39186327 39186328 39186329 39186332 39186333 39186335 39186336 39186337 39186340 39186342 39186344 39186348 39186352 39186354 39186355 39186359 39186362 39186363 39186366 39186367 39186369 39186371 39186373 39186374 39186376 39186379 39186400 39186402 39186404 39186405 39186408 39186409 39186410 39186411 39186413 39186415 39186416 39186417 39186418 39186419 39187293 39187295 39187297 39208847 39208852 39208853 39208854 39208855 39208856 39208857 39208858 39208859 39208860 39208861 39208862 39208863 39208864 39208865 39208866 39208867 39208868 39208869 39208870 39208871 39208872 39208873 39208875 39208876 39208877 39208878 39208879 39208880 39208881 39208884 39210858 39210864 39210873 39210874 39210900 39210916 39210919 39210925 39214577 39217541 39217788 39217790 39217801 39219298 39219313 39220661 39220666 39220667 39220668 39220669 39220672 39220673 39220674 39220675 39220676 39220677 39220678 39220679 39220681 39220682 39220683 39220685 39220687 39221056 39221057 39221058 39221059 39221060 39221061 39221062 39221063 39221064 39221065 39221066 39221067 39221069 39221075 39221089 39221090 39221091 39221092 39221093 39221095 39221096 39221097 39221098 39221099 39221100 39221101 39221102 39221103 39221107 39221108 39221109 39221110 39221112 39221113 39221114 39221115 39221116 39221117 39221118 39221119 39227222 39227223 39227225 39227226 39227377 39229414 39229416 39229417 39229418 39229420 39229421 39229423 39230325 39230326 39230327 39230328 39230329 39230332 39230333 39230334 39230335 39230336 39230337 39230338 39230339 39230342 39230343 39230344 39230345 39230346 39230347 39230348 39230350 39230351 39230352 39230355 39230356 39230357 39230359 39231492 39231634 39231947 39231949 39231950 39231951 39231952 39231953 39231954 39231955 39231959 39231961 39233294 39233295 39233299 39233453 39233455 39233913 39235157 39235921 39235930 39235932 39235940 39235941 39235942 39235948 39235949 39237814 39237861 39237863 39239025 39239048 39239052 39239053 39239054 39239055 39242835 39242836 39242838 39242840 39242841 39242842 39242843 39242844 39242845 39242846 39242847 39242848 39246165 39246167 39246169 39246170 39246171 39246174 39246175 39246176 39246177 39246179 39246182 39246183 39246184 39246186 39246187 39246189 39246190 39246192 39246904 39246905 39246906 39246910 39246914 39246916 39246917 39246918 39246919 39246920 39246922 39246923 39246924 39246925 39246926 39246927 39246929 39246931 39247647 39247648 39247649 39247661 39247662 39247663 39247706 39247712 39247713 39247714 39247715 39247716 39248611 39248612 39248615 39248619 39248623 39248628 39248630 39248668 39248669 39248670 39248671 39248672 39248673 39248674 39251180 39251181 39251182 39251183 39251184 39251185 39251186 39251187 39251188 39258079 39258080 39258081 39258082 39258083 39258084 39258085 39258086 39258087 39258088 39258089 39258090 39258091 39258092 39258328 39258329 39258769 39258770 39258771 39258772 39258773 39258774 39258777 39258778 39258779 39258780 39258781 39258782 39258783 39258784 39258785 39258786 39258787 39258790 39258791 39258792 39258793 39258794 39258796 39263611 39263612 39263613 39263614 39263615 39263616 39263617 39263618 39263620 39263621 39263622 39263623 39263625 39263626 39263627 39263628 39263629 39263630 39263631 39263632 39263633 39263634 39263635 39263636 39263637 39263639 39263640 39263641 39264431 39264433 39264434 39264435 39264436 39264437 39264438 39264440 39264441 39264442 39264443 39264444 39264445 39264447 39264448 39264449 39264450 39264451 39264453 39264454 39264458 39265345 39265350 39265355 39265877 39265878 39265879 39265880 39277869 39277871 39277893 39284362 39284366 39284367 39284370 39284371 39284372 39284373 39284374 39284375 39284377 39284381 39284382 39284383 39284384 39284385 39284386 39284387 39284388 39284566 39284568 39284570 39287018 39287023 39287024 39287027 39287028 39287029 39287690 39287691 39287692 39287693 39287694 39291376 39291377 39291378 39291379 39291380 39291381 39291382 39291383 39291384 39291385 39291387 39291388 39291389 39293335 39293345 39293347 39293349 39293351 39293353 39293354 39293355 39293356 39293357 39293620 39293626 39293627 39293635 39295335 39295376 39295391 39295397 39298627 39301190 39301191 39301192 39301193 39301194 39301195 39317626 39317627 39317630 39317632 39317634 39317635 39317636 39317638 39317639 39317641 39317737 39317738 39318161 39318162 39318163 39318165 39318166 39318168 39318169 39318171 39318899 39320708 39328295 39328296 39328299 39328301 39328304 39328307 39328310 39328311 39328312 39328313 39328314 39328318 39328588 39328589 39328594 39328595 39328635 39328636 39328637 39328638 39328639 39328640 39328641 39328642 39328644 39328645 39328647 39328648 39328649 39328650 39328651 39328653 39328655 39328656 39328658 39328659 39328661 39328663 39328664 39328665 39330756 39330757 39330758 39330759 39330760 39330764 39330765 39332142 39332146 39332148 39332149 39332151 39332152 39332153 39332154 39332157 39332158 39332161 39332162 39332163 39339698 39339699 39339700 39339701 39339702 39339704 39341441 39341442 39341447 39341450 39341452 39341454 39342707 39342708 39342709 39342712 39342713 39342714 39342715 39342716 39342718 39342719 39342720 39342721 39342725 39342727 39342729 39342730 39342731 39346226 39346228 39346229 39346231 39346232 39346233 39346234 39346236 39346238 39346239 39346972 39346977 39346981 39346983 39346990 39347727 39347733 39347735 39347736 39347742 39347756 39347764 39347770 39347771 39347772 39347774 39347775 39348719 39348720 39348721 39348722 39357772 39366532 39368498 39368689 39368691 39368697 39370581 39371149 39371236 39371237 39371241 39371243 39371244 39371245 39371246 39371247 39371248 39371894 39371895 39371896 39371897 39371898 39371900 39371901 39371902 39371903 39371904 39372107 39372108 39372109 39372110 39372111 39372112 39372115 39372117 39372120 39372123 39373364 39373365 39373366 39373367 39373368 39373369 39373370 39373371 39373372 39373373 39373374 39373375 39373376 39373377 39373378 39373379 39373380 39373381 39373383 39373384 39373853 39373854 39373855 39373856 39373857 39373858 39373859 39373860 39373861 39373862 39373863 39373864 39373865 39373866 39373867 39373868 39373869 39373870 39373871 39373872 39374676 39374677 39374678 39374679 39374680 39374681 39374682 39374683 39374684 39374685 39374686 39374687 39374688 39374689 39374690 39374691 39374692 39374693 39374694 39374695 39375603 39375604 39375605 39375606 39375607 39375608 39375609 39375610 39375611 39375612 39375949 39375951 39375954 39375955 39375957 39375958 39381176 39381243 39381603 39381604 39381606 39381608 39381609 39381610 39381611 39381612 39381613 39381614 39381615 39382544 39382546 39382551 39382553 39382554 39382555 39382556 39382557 39382558 39382560 39382562 39382720 39382721 39382722 39382724 39382726 39382727 39382728 39382729 39382730 39382731 39382732 39382733 39382735 39382737 39382738 39382739 39383356 39383358 39383360 39383362 39383364 39383365 39383367 39383369 39383370 39383373 39383374 39383690 39383693 39383696 39383697 39383698 39383699 39383700 39383701 39383702 39383703 39383704 39383705 39383706 39383707 39383708 39383709 39383710 39383711 39383712 39383713 39383714 39383715 39383716 39383717 39383718 39383719 39389110 39389112 39389113 39391679 39391680 39391682 39391683 39391686 39391687 39392115 39392120 39392124 39392125 39392126 39392127 39392128 39392841 39392842 39392844 39392848 39392850 39392851 39392852 39392853 39392854 39392855 39392856 39392857 39394294 39394295 39394296 39394298 39394299 39394300 39394302 39394304 39396783 39396784 39396785 39396786 39396788 39396792 39396795 39396796 39396798 39396799 39396802 39396804 39396807 39399975 39399977 39399978 39399979 39399980 39399981 39399983 39399985 39400046 39400048 39400167 39410379 39410380 39410381 39410382 39410385 39410386 39410389 39410391 39410393 39410394 39410395 39410396 39410397 39410398 39410400 39410402 39410403 39410404 39410405 39410406 39410407 39410408 39410410 39410411 39410412 39410413 39410415 39410416 39410417 39410418 39410419 39410420 39410421 39410422 39410423 39410425 39410426 39410427 39410428 39410430 39410431 39410432 39410433 39410434 39410435 39410436 39412578 39415108 39415109 39415110 39415111 39415119 39415122 39415123 39415127 39415128 39415134 39415137 39415143 39415144 39415145 39415148 39415149 39415150 39415151 39415152 39415153 39415154 39415155 39419836 39419837 39419838 39419839 39419840 39419853 39419854 39419856 39419857 39419858 39419859 39419860 39420144 39420145 39420146 39420147 39420148 39420149 39420150 39420151 39420152 39420153 39420155 39420156 39420157 39420158 39420161 39420163 39420165 39420166 39420167 39420168 39420171 39420172 39420177 39420178 39423274 39423275 39423276 39423277 39423636 39423637 39423639 39423644 39423647 39423653 39423656 39423657 39423658 39423659 39423660 39423662 39423663 39423667 39423669 39423670 39423671 39423674 39423680 39423681 39423683 39423685 39423686 39423687 39423688 39423691 39423692 39423693 39423694 39423695 39423696 39423697 39423698 39426098 39426099 39426100 39426101 39426103 39426104 39426105 39426106 39426107 39426108 39426109 39426110 39426111 39426112 39426113 39426114 39426115 39426116 39426117 39426118 39426119 39426120 39426121 39426122 39427826 39427829 39427830 39427831 39427832 39427836 39427837 39427840 39427842 39427843 39427844 39427845 39427846 39427848 39427849 39430831 39430832 39430833 39430836 39430837 39430838 39430839 39430840 39430842 39430843 39430844 39430845 39430846 39430847 39430848 39430849 39430850 39430851 39430852 39430853 39430854 39430855 39430857 39430858 39430860 39430861 39430862 39430863 39430864 39430865 39430867 39430868 39430869 39430870 39430872 39430877 39430878 39430879 39430880 39430881 39430882 39430883 39430884 39430885 39430886 39430887 39430888 39430889 39430890 39430891 39430892 39430893 39430894 39430895 39430896 39430898 39430899 39430900 39430901 39430903 39430904 39430905 39430906 39430907 39430908 39430909 39430910 39430911 39430912 39430914 39430915 39430916 39430917 39430918 39430919 39430920 39430921 39430922 39430923 39431817 39431819 39436350 39436351 39436352 39437060 39438722 39438723 39439857 39439860 39439861 39439862 39439863 39442605 39442606 39442607 39442608 39452801 39452804 39461900 39461902 39461903 39461905 39461906 39463873 39463876 39463877 39463878 39463879 39463880 39463881 39463882 39463885 39463886 39463887 39463889 39463890 39463893 39463898 39463899 39463902 39463903 39463904 39463905 39463906 39463907 39463908 39463910 39463911 39463913 39463914 39463917 39463918 39463919 39465797 39465798 39465800 39465801 39465804 39465805 39465806 39465809 39465814 39465815 39465817 39468698 39468700 39468702 39468704 39468705 39468706 39468707 39468708 39468709 39469259 39475013 39475014 39475015 39475016 39475117 39476003 39476092 39476097 39476098 39476100 39476101 39476103 39476105 39476674 39476675 39476676 39476678 39476690 39476695 39476696 39476699 39476700 39476701 39477013 39477014 39477015 39477017 39477019 39477020 39477332 39477335 39477336 39477337 39492354 39519587 39519737 39519738 39519739 39519740 39519915 39519916 39519917 39519918 39519919 39519920 39519921 39519922 39519923 39519924 39519964 39519965 39519966 39519967 39519968 39519969 39519970 39519971 39519972 39519973 39520124 39520125 39520126 39520127 39520128 39520129 39520130 39520131 39520132 39520133 39520155 39520156 39520157 39520158 39520159 39520160 39520370 39520373 39520376 39520525 39520526 39520531 39524167 39524639 39524642 39524643 39524645 39524646 39524648 39524649 39524650 39524651 39524652 39524653 39524654 39524655 39524656 39524657 39525819 39526027 39526030 39526032 39526033 39526035 39526037 39526038 39526039 39526040 39526041 39526042 39526043 39526044 39526045 39526046 39526047 39527938 39527939 39527940 39527942 39527943 39527945 39527946 39527948 39527950 39527951 39527952 39527953 39527954 39527955 39527956 39527957 39528208 39528210 39528211 39528213 39528214 39528215 39528227 39528234 39528236 39528238 39528240 39528241 39528456 39528457 39528458 39528682 39529105 39529106 39529107 39529109 39529110 39529111 39529112 39529114 39529116 39529509 39529513 39529519 39529522 39529523 39529524 39529525 39529526 39529527 39529528 39529529 39529530 39529531 39529532 39529533 39529535 39530545 39530546 39530547 39530548 39530549 39530550 39530551 39530552 39530553 39530555 39530556 39530557 39530558 39530559 39530560 39530561 39530562 39530563 39530564 39530565 39530566 39530567 39530568 39530570 39530571 39530572 39530573 39530574 39530575 39531608 39531611 39531612 39531613 39531614 39531615 39531616 39531617 39531618 39531619 39531620 39531621 39531622 39531623 39531710 39531711 39531712 39531713 39531714 39531715 39531716 39531717 39531718 39531719 39531720 39531721 39531722 39531723 39531724 39536076 39536077 39536078 39536079 39536080 39536081 39536082 39536083 39536084 39536085 39536086 39536087 39536088 39536089 39536090 39536091 39536093 39536094 39536095 39536137 39536138 39536139 39536140 39536141 39536142 39536143 39536145 39536147 39536151 39536427 39536428 39536429 39536432 39536433 39536435 39536437 39536438 39536439 39536443 39536444 39536445 39536446 39536789 39536790 39536791 39536792 39536793 39536794 39536796 39537506 39537507 39537508 39537509 39537510 39537511 39537512 39537513 39537514 39537515 39537516 39537517 39537518 39537520 39537521 39537522 39537523 39537524 39537525 39538380 39538381 39538382 39538383 39538384 39538385 39538386 39538387 39538388 39538389 39538390 39538393 39538394 39538396 39538397 39538399 39538401 39538404 39538405 39541352 39541654 39541655 39541656 39541657 39541663 39541664 39541666 39541667 39541668 39541669 39541670 39541671 39541673 39541674 39541675 39541676 39541677 39541678 39541679 39541680 39541681 39542100 39543706 39563968 39563973 39564332 39573887 39573888 39573889 39573890 39573892 39573893 39573930 39573932 39573933 39573934 39578064 39578065 39578083 39578084 39578085 39578086 39578087 39578088 39578089 39578090 39578091 39578092 39579344 39579345 39579347 39583907 39583908 39583909 39583910 39583912 39583924 39583927 39583929 39583931 39583932 39583935 39583939 39584159 39584160 39584162 39584163 39584165 39584166 39584167 39584168 39584169 39584170 39584172 39584173 39584174 39584175 39584176 39584177 39584178 39586286 39586287 39586748 39586749 39586750 39586801 39586802 39586803 39586804 39586805 39586806 39586807 39586808 39586809 39586810 39586812 39587200 39587201 39587202 39587203 39587204 39587205 39587210 39587214 39587218 39587219 39587522 39587525 39587526 39587528 39587533 39587535 39587537 39587539 39587540 39587541 39590273 39590274 39590278 39590279 39590280 39590281 39590282 39591021 39591022 39591024 39591025 39591026 39591027 39591028 39591029 39591032 39591033 39591034 39591035 39591037 39591038 39591039 39591041 39591043 39591044 39593713 39593714 39593715 39593716 39593717 39593718 39593719 39593720 39593721 39593722 39593725 39593726 39593727 39593728 39593729 39593730 39593731 39594854 39594855 39594856 39594857 39594859 39594860 39594861 39594862 39594863 39594864 39595080 39595081 39595085 39595086 39595088 39595090 39595108 39595117 39595151 39595153 39595156 39595159 39595166 39595173 39596668 39596671 39596674 39596675 39596678 39596682 39596685 39596686 39596750 39596751 39596752 39596753 39596754 39596755 39596756 39596757 39596758 39596759 39597908 39597910 39597911 39597912 39597913 39597915 39598371 39598372 39598374 39598375 39598376 39598378 39598380 39605732 39605734 39605735 39605736 39605737 39605738 39605740 39605741 39605742 39605744 39605768 39605769 39605770 39605771 39605772 39605773 39605774 39605775 39605776 39605777 39605778 39605779 39605780 39605781 39605782 39607430 39607431 39607432 39607433 39607434 39607435 39607436 39607437 39607438 39607439 39607440 39607443 39607445 39607446 39607447 39607448 39607450 39607451 39607454 39607455 39607458 39607460 39607466 39608254 39608261 39608333 39608334 39608337 39608343 39608345 39608346 39608347 39608349 39608351 39608359 39608363 39608370 39608373 39608377 39608378 39608383 39608384 39608386 39608391 39608394 39608396 39608397 39610108 39610111 39610112 39610114 39610115 39610116 39610117 39610118 39610119 39610120 39610121 39610122 39610123 39610124 39610125 39610126 39610127 39611628 39611629 39611630 39611631 39611633 39611634 39611635 39611636 39611637 39611638 39611639 39611640 39611641 39611642 39611643 39611644 39611645 39611646 39611647 39612259 39612261 39612262 39612263 39612264 39612265 39612266 39612267 39612268 39612275 39612279 39612283 39612285 39612289 39612299 39612302 39612303 39612307 39612311 39612314 39612317 39612322 39612326 39613528 39613530 39613531 39613532 39613533 39613534 39613535 39613536 39613537 39613538 39613539 39613540 39613541 39613542 39613543 39613544 39613545 39613546 39613547 39615271 39615272 39615273 39615274 39615275 39615276 39615277 39615278 39615279 39615281 39615282 39615283 39615284 39615285 39615286 39615287 39615288 39615289 39615290 39615291 39615292 39615293 39615294 39615295 39615296 39615297 39615298 39615299 39615300 39615771 39615772 39615799 39615800 39615801 39615802 39615806 39615818 39615834 39615835 39615836 39615837 39615838 39615839 39615840 39615841 39615842 39615843 39615844 39615845 39615846 39615847 39615848 39615849 39615850 39615851 39615852 39615853 39616901 39616903 39616904 39616905 39616906 39616907 39616908 39616909 39616910 39616911 39616912 39616913 39616915 39616917 39616924 39616926 39618493 39618496 39618500 39618501 39618502 39618503 39618505 39618506 39618507 39618509 39618511 39618512 39618514 39618519 39619243 39621252 39621253 39621255 39621263 39621265 39621266 39621268 39621269 39621270 39621271 39622460 39623189 39623191 39623192 39625596 39625608 39625609 39625611 39625965 39625968 39625970 39625971 39625972 39625973 39625974 39625979 39625984 39625985 39625987 39625988 39625991 39625992 39626148 39626911 39627031 39628851 39628853 39628854 39628855 39628856 39628858 39632122 39632159 39635113 39635430 39635431 39643271 39643275 39643467 39643468 39643901 39643903 39643904 39644287 39647487 39647488 39647489 39647490 39647492 39647494 39647496 39647497 39647499 39647505 39647510 39647511 39647513 39647514 39647590 39647591 39648463 39648464 39648465 39648466 39648467 39648468 39648469 39648470 39648471 39648472 39648473 39648474 39648475 39648476 39648477 39648903 39648904 39648905 39648906 39648907 39648908 39648909 39648910 39648911 39648912 39648913 39648914 39648915 39648916 39648917 39648918 39648919 39648920 39648921 39648922 39649245 39649248 39649249 39649250 396
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Physio-Control, Inc.
Reason for Recall:
Monitor/Defibrillator may not deliver a shock after the "Shock" button on the keypad is pressed as a result of oxidation that has formed over time within the button.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

LIFEPAK 15 MONITOR/DEFIBRILLATOR that was either manufactured with or received an upgrade kit that contained an affected keypad.

Product Codes/Lot Numbers:

Serial numbers: 37964613 37964617 37964621 37964628 37964630 37968424 37968429 37964596 37964601 37964607 37964610 37968715 37968716 37968717 37968718 37968721 37968722 37968724 38046540 38046549 38046552 38046565 38046571 38046572 38046573 38051949 38051951 38051954 38051955 38051958 38051963 38051969 38051972 38051973 38051974 38051980 38053216 38053218 38053220 38053221 38053225 38053231 38053233 38053236 38053239 38059568 38059570 38059574 38059591 38120769 38120772 38120775 38120780 38120781 38120783 38120784 38120787 38120788 38120790 38124016 38124030 38124035 38124042 38124044 38124047 38125969 38125971 38125972 38125980 38125982 38125987 38125990 38126255 38126258 38126260 38126265 38126267 38126269 38126272 38126273 38126277 38127130 38127131 38127132 38127134 38127136 38127138 38127139 38127140 38127141 38127143 38127144 38130272 38130273 38130276 38130277 38130279 38130282 38130283 38130285 38130286 38130287 38131588 38131589 38131592 38131597 38131600 38131602 38131603 38131605 38131607 38131608 38131756 38131757 38131760 38131761 38131762 38131764 38131765 38131767 38131773 38131775 38131776 38131777 38131842 38131845 38132370 38132951 38134512 38134524 38135205 38135212 38135215 38136507 38136508 38136513 38136516 38137137 38137143 38137149 38137155 38137168 38137174 38138844 38138846 38141005 38142743 38150694 38151893 38151898 38153123 38153124 38153136 38155919 38158080 38159137 38159181 38159183 38159187 38159197 38161892 38161893 38161898 38161899 38161902 38161907 38161909 38161912 38162531 38162561 38162597 38164391 38164403 38164420 38164448 38166449 38166461 38166462 38166475 38166479 38166485 38166486 38166497 38166507 38166513 38166516 38166541 38166542 38166543 38166544 38166549 38166550 38166555 38166558 38166569 38166571 38166581 38166582 38166586 38166594 38166610 38166614 38166628 38166632 38166634 38166642 38166647 38166674 38166675 38166676 38166677 38166678 38177270 38177272 38177275 38191219 38191235 38200435 38200436 38200441 38200443 38200444 38200446 38200447 38200448 38200450 38200451 38200452 38200453 38200454 38200457 38200458 38200461 38200462 38200464 38200465 38200466 38200467 38200469 38200470 38200471 38200472 38200474 38200476 38202790 38202791 38202802 38202803 38202809 38202821 38205398 38205404 38205407 38205410 38205413 38205781 38205783 38205784 38205794 38206256 38211038 38211041 38211042 38211043 38211044 38211051 38211059 38214239 38214242 38214247 38214259 38214266 38214271 38214279 38217387 38217390 38217400 38221456 38221457 38224395 38224601 38227461 38253472 38253478 38253768 38253769 38253770 38253786 38253790 38253798 38253805 38263304 38263329 38272217 38273859 38274178 38274396 38274508 38279197 38279198 38279199 38279202 38279206 38281774 38282979 38282983 38282987 38287524 38287525 38287530 38287845 38287846 38290137 38290140 38290141 38290535 38290536 38290538 38290539 38307505 38315311 38315330 38322516 38323219 38323233 38323263 38323311 38324196 38324198 38324199 38324201 38324210 38324213 38327183 38332345 38332349 38332355 38332358 38332363 38332748 38332749 38332753 38332754 38332761 38332940 38338635 38340566 38341984 38343138 38343162 38343235 38343241 38345048 38346560 38353024 38353025 38353033 38354810 38355627 38355629 38355631 38355637 38355638 38355639 38355645 38359027 38359559 38362860 38364966 38365244 38365249 38365937 38365940 38368303 38368304 38368307 38368327 38368329 38368616 38369586 38370274 38370275 38370277 38370280 38370282 38370293 38370425 38370431 38370433 38370434 38371200 38371201 38371218 38371701 38371702 38371708 38371709 38371711 38371714 38372527 38373618 38374119 38374686 38374688 38374696 38374701 38432103 38441912 38441916 38441918 38441920 38441931 38441934 38441935 38441936 38445594 38445597 38445604 38445626 38445640 38445649 38449384 38449391 38449400 38449410 38458698 38472562 38517179 38517227 38520270 38520276 38521672 38523568 38523584 38523596 38523602 38523604 38523606 38523607 38523610 38528452 38528458 38528471 38534441 38536323 38536351 38536916 38536922 38536929 38537159 38537164 38537168 38537704 38537706 38537758 38537759 38537768 38537774 38537776 38537779 38538715 38538719 38538721 38538879 38538881 38538926 38538931 38538932 38538935 38538940 38538953 38538956 38538957 38538969 38538983 38540917 38540923 38540935 38540937 38540940 38540944 38540948 38540955 38540956 38540959 38540960 38540961 38540964 38540967 38540971 38540972 38540974 38543095 38543100 38543102 38543108 38543111 38543112 38543113 38544618 38544622 38544623 38544625 38544627 38544635 38544637 38544638 38544646 38544649 38544658 38544672 38544673 38544675 38544763 38544764 38544771 38548002 38548004 38548012 38548022 38548037 38548038 38548040 38553528 38553529 38553537 38553545 38559431 38559446 38559447 38560013 38560014 38560018 38560019 38560020 38560021 38560030 38560031 38560032 38560033 38560034 38560035 38560036 38560037 38560038 38560039 38560040 38560041 38561014 38561021 38561022 38561023 38561024 38561027 38564049 38564051 38564679 38567823 38567825 38568460 38568463 38568468 38568473 38568874 38568875 38568876 38568877 38568878 38568879 38568882 38568890 38568891 38569281 38574882 38578213 38611410 38611411 38611413 38611414 38611509 38611513 38611516 38611517 38611518 38611520 38611521 38611522 38611523 38611524 38611526 38611529 38611530 38611531 38611533 38611536 38611537 38611553 38611554 38611556 38611557 38611559 38611561 38613942 38613943 38613945 38613949 38613950 38613952 38613954 38613967 38619552 38620457 38620459 38620466 38620477 38620479 38620480 38620481 38620495 38624129 38624139 38624141 38624145 38624149 38624150 38624151 38624153 38624159 38626953 38626956 38626958 38626963 38626967 38626991 38626995 38631721 38631723 38631725 38631730 38631746 38648317 38648319 38720330 38723976 38724387 38725753 38725771 38726757 38726761 38726794 38726796 38733842 38733852 38733858 38735077 38738213 38738217 38738219 38738222 38743122 38756198 38763408 38763413 38763423 38763427 38774131 38778622 38784168 38784951 38784969 38785815 38785821 38790361 38790364 38790367 38790373 38790381 38790383 38790389 38790398 38790399 38790403 38790407 38793515 38793516 38793529 38793539 38793540 38793546 38793552 38793555 38793559 38793560 38793562 38793563 38793565 38793566 38793569 38793575 38793580 38795890 38795898 38795906 38798710 38798718 38798795 38798801 38800559 38800568 38800581 38800586 38800587 38800588 38800594 38800595 38802775 38802783 38802785 38802787 38802789 38802791 38802797 38802823 38802828 38802846 38805559 38805572 38805575 38805589 38805592 38805596 38805602 38805605 38805609 38805611 38805612 38805616 38805619 38807481 38807487 38807488 38807491 38807492 38807493 38807494 38807496 38807498 38807501 38807502 38807508 38807509 38809007 38809009 38809010 38809015 38809017 38809019 38809020 38809022 38809025 38809035 38811712 38811713 38811714 38811717 38811720 38811731 38811732 38817539 38817544 38817554 38817555 38817557 38817558 38817559 38817560 38817562 38817563 38817564 38817772 38817776 38817777 38819691 38819700 38819712 38821747 38821771 38821775 38821787 38821792 38821799 38821800 38821801 38821803 38821804 38821805 38821807 38824055 38824080 38839958 38839964 38839966 38839972 38840725 38852928 38854322 38854328 38854329 38856558 38856571 38856576 38856580 38856589 38870845 38870852 38871061 38871062 38877850 38878667 38878682 38878686 38878688 38878692 38881890 38881891 38897292 38897298 38897493 38897495 38897499 38898618 38898619 38898622 38898623 38898625 38898629 38898631 38898634 38898636 38898640 38898647 38898651 38934844 38934845 38934848 38934850 38934868 38948322 38988236 38988242 38988247 38988251 38988256 38988257 38988258 38988259 38989256 38989260 38989261 38989264 38989265 38989267 38989269 38989275 38989336 38992044 38992045 38992061 38996375 38996376 38996377 38996378 38996379 38996380 38996381 38996382 38996383 38996386 38996389 38996390 38996392 38996393 38996395 38996397 38996398 38996408 38996792 38996793 38996794 38996799 38996807 38999803 38999805 38999809 38999810 38999812 39000250 39004380 39004384 39023652 39025293 39025303 39029063 39029067 39029068 39029073 39029143 39029147 39029161 39029168 39029176 39029200 39029205 39029206 39029215 39029218 39039179 39039192 39040605 39048208 39048209 39048488 39048504 39048506 39050458 39050475 39050492 39050508 39053292 39053297 39053316 39054574 39054577 39054578 39054582 39054583 39054584 39054586 39054589 39054593 39054598 39054602 39054603 39054605 39054606 39054607 39054622 39054634 39054636 39054642 39054646 39054648 39054651 39054654 39054666 39054673 39054702 39054719 39054729 39054736 39054741 39054749 39054751 39054759 39054766 39054773 39054787 39054788 39054809 39054830 39054833 39054834 39054835 39054836 39054842 39054850 39054852 39054854 39054857 39054861 39059003 39059006 39059474 39059477 39059479 39059482 39059485 39059486 39060195 39060196 39084702 39090471 39108559 39108565 39108570 39108571 39108572 39108585 39108586 39108589 39109072 39109078 39109080 39109081 39119994 39119995 39119996 39119997 39119998 39119999 39120000 39120001 39120002 39120003 39126759 39139844 39139853 39150420 39150421 39150423 39150430 39150865 39152709 39152710 39152711 39152713 39153383 39155791 39178106 39178108 39178109 39178110 39178111 39178112 39178113 39179615 39179708 39179732 39179733 39179734 39179735 39179736 39179737 39179738 39179739 39179740 39179742 39179743 39179744 39179745 39179746 39179747 39179748 39179749 37923203 37923205 37923208 37923213 37924152 37928794 37928964 37930942 37930944 37930964 37930966 37930977 37930990 37930993 37931003 37933241 37933244 37933249 37933250 37933262 37937842 37937843 37937848 37937849 37937851 37937854 37937855 37937856 37937862 37937863 37937864 37937866 37937868 37937870 37937871 37937872 37937873 37937874 37937885 37937886 37937887 37937889 37937891 37937892 37963295 37963300 37963302 37963323 37963324 37963325 37963328 37963926 37963928 37963935 37964552 37964554 37964557 37964560 37964564 37964567 37964575 37964585 37964587 37964588 39179750 39179751 39179752 39179754 39179755 39180764 39180766 39180767 39180768 39180770 39180771 39180773 39180774 39180775 39180776 39180779 39180780 39180783 39180784 39180785 39180787 39180788 39180789 39180790 39182063 39182066 39182070 39182076 39182077 39182078 39182080 39182081 39183493 39183494 39183495 39183496 39183497 39183498 39183499 39183500 39183501 39183502 39183640 39183641 39183642 39183643 39183644 39183645 39183646 39183648 39183649 39183650 39183651 39183652 39183655 39183657 39184048 39184050 39184200 39186322 39186323 39186325 39186326 39186327 39186328 39186329 39186332 39186333 39186335 39186336 39186337 39186340 39186342 39186344 39186348 39186352 39186354 39186355 39186359 39186362 39186363 39186366 39186367 39186369 39186371 39186373 39186374 39186376 39186379 39186400 39186402 39186404 39186405 39186408 39186409 39186410 39186411 39186413 39186415 39186416 39186417 39186418 39186419 39187293 39187295 39187297 39208847 39208852 39208853 39208854 39208855 39208856 39208857 39208858 39208859 39208860 39208861 39208862 39208863 39208864 39208865 39208866 39208867 39208868 39208869 39208870 39208871 39208872 39208873 39208875 39208876 39208877 39208878 39208879 39208880 39208881 39208884 39210858 39210864 39210873 39210874 39210900 39210916 39210919 39210925 39214577 39217541 39217788 39217790 39217801 39219298 39219313 39220661 39220666 39220667 39220668 39220669 39220672 39220673 39220674 39220675 39220676 39220677 39220678 39220679 39220681 39220682 39220683 39220685 39220687 39221056 39221057 39221058 39221059 39221060 39221061 39221062 39221063 39221064 39221065 39221066 39221067 39221069 39221075 39221089 39221090 39221091 39221092 39221093 39221095 39221096 39221097 39221098 39221099 39221100 39221101 39221102 39221103 39221107 39221108 39221109 39221110 39221112 39221113 39221114 39221115 39221116 39221117 39221118 39221119 39227222 39227223 39227225 39227226 39227377 39229414 39229416 39229417 39229418 39229420 39229421 39229423 39230325 39230326 39230327 39230328 39230329 39230332 39230333 39230334 39230335 39230336 39230337 39230338 39230339 39230342 39230343 39230344 39230345 39230346 39230347 39230348 39230350 39230351 39230352 39230355 39230356 39230357 39230359 39231492 39231634 39231947 39231949 39231950 39231951 39231952 39231953 39231954 39231955 39231959 39231961 39233294 39233295 39233299 39233453 39233455 39233913 39235157 39235921 39235930 39235932 39235940 39235941 39235942 39235948 39235949 39237814 39237861 39237863 39239025 39239048 39239052 39239053 39239054 39239055 39242835 39242836 39242838 39242840 39242841 39242842 39242843 39242844 39242845 39242846 39242847 39242848 39246165 39246167 39246169 39246170 39246171 39246174 39246175 39246176 39246177 39246179 39246182 39246183 39246184 39246186 39246187 39246189 39246190 39246192 39246904 39246905 39246906 39246910 39246914 39246916 39246917 39246918 39246919 39246920 39246922 39246923 39246924 39246925 39246926 39246927 39246929 39246931 39247647 39247648 39247649 39247661 39247662 39247663 39247706 39247712 39247713 39247714 39247715 39247716 39248611 39248612 39248615 39248619 39248623 39248628 39248630 39248668 39248669 39248670 39248671 39248672 39248673 39248674 39251180 39251181 39251182 39251183 39251184 39251185 39251186 39251187 39251188 39258079 39258080 39258081 39258082 39258083 39258084 39258085 39258086 39258087 39258088 39258089 39258090 39258091 39258092 39258328 39258329 39258769 39258770 39258771 39258772 39258773 39258774 39258777 39258778 39258779 39258780 39258781 39258782 39258783 39258784 39258785 39258786 39258787 39258790 39258791 39258792 39258793 39258794 39258796 39263611 39263612 39263613 39263614 39263615 39263616 39263617 39263618 39263620 39263621 39263622 39263623 39263625 39263626 39263627 39263628 39263629 39263630 39263631 39263632 39263633 39263634 39263635 39263636 39263637 39263639 39263640 39263641 39264431 39264433 39264434 39264435 39264436 39264437 39264438 39264440 39264441 39264442 39264443 39264444 39264445 39264447 39264448 39264449 39264450 39264451 39264453 39264454 39264458 39265345 39265350 39265355 39265877 39265878 39265879 39265880 39277869 39277871 39277893 39284362 39284366 39284367 39284370 39284371 39284372 39284373 39284374 39284375 39284377 39284381 39284382 39284383 39284384 39284385 39284386 39284387 39284388 39284566 39284568 39284570 39287018 39287023 39287024 39287027 39287028 39287029 39287690 39287691 39287692 39287693 39287694 39291376 39291377 39291378 39291379 39291380 39291381 39291382 39291383 39291384 39291385 39291387 39291388 39291389 39293335 39293345 39293347 39293349 39293351 39293353 39293354 39293355 39293356 39293357 39293620 39293626 39293627 39293635 39295335 39295376 39295391 39295397 39298627 39301190 39301191 39301192 39301193 39301194 39301195 39317626 39317627 39317630 39317632 39317634 39317635 39317636 39317638 39317639 39317641 39317737 39317738 39318161 39318162 39318163 39318165 39318166 39318168 39318169 39318171 39318899 39320708 39328295 39328296 39328299 39328301 39328304 39328307 39328310 39328311 39328312 39328313 39328314 39328318 39328588 39328589 39328594 39328595 39328635 39328636 39328637 39328638 39328639 39328640 39328641 39328642 39328644 39328645 39328647 39328648 39328649 39328650 39328651 39328653 39328655 39328656 39328658 39328659 39328661 39328663 39328664 39328665 39330756 39330757 39330758 39330759 39330760 39330764 39330765 39332142 39332146 39332148 39332149 39332151 39332152 39332153 39332154 39332157 39332158 39332161 39332162 39332163 39339698 39339699 39339700 39339701 39339702 39339704 39341441 39341442 39341447 39341450 39341452 39341454 39342707 39342708 39342709 39342712 39342713 39342714 39342715 39342716 39342718 39342719 39342720 39342721 39342725 39342727 39342729 39342730 39342731 39346226 39346228 39346229 39346231 39346232 39346233 39346234 39346236 39346238 39346239 39346972 39346977 39346981 39346983 39346990 39347727 39347733 39347735 39347736 39347742 39347756 39347764 39347770 39347771 39347772 39347774 39347775 39348719 39348720 39348721 39348722 39357772 39366532 39368498 39368689 39368691 39368697 39370581 39371149 39371236 39371237 39371241 39371243 39371244 39371245 39371246 39371247 39371248 39371894 39371895 39371896 39371897 39371898 39371900 39371901 39371902 39371903 39371904 39372107 39372108 39372109 39372110 39372111 39372112 39372115 39372117 39372120 39372123 39373364 39373365 39373366 39373367 39373368 39373369 39373370 39373371 39373372 39373373 39373374 39373375 39373376 39373377 39373378 39373379 39373380 39373381 39373383 39373384 39373853 39373854 39373855 39373856 39373857 39373858 39373859 39373860 39373861 39373862 39373863 39373864 39373865 39373866 39373867 39373868 39373869 39373870 39373871 39373872 39374676 39374677 39374678 39374679 39374680 39374681 39374682 39374683 39374684 39374685 39374686 39374687 39374688 39374689 39374690 39374691 39374692 39374693 39374694 39374695 39375603 39375604 39375605 39375606 39375607 39375608 39375609 39375610 39375611 39375612 39375949 39375951 39375954 39375955 39375957 39375958 39381176 39381243 39381603 39381604 39381606 39381608 39381609 39381610 39381611 39381612 39381613 39381614 39381615 39382544 39382546 39382551 39382553 39382554 39382555 39382556 39382557 39382558 39382560 39382562 39382720 39382721 39382722 39382724 39382726 39382727 39382728 39382729 39382730 39382731 39382732 39382733 39382735 39382737 39382738 39382739 39383356 39383358 39383360 39383362 39383364 39383365 39383367 39383369 39383370 39383373 39383374 39383690 39383693 39383696 39383697 39383698 39383699 39383700 39383701 39383702 39383703 39383704 39383705 39383706 39383707 39383708 39383709 39383710 39383711 39383712 39383713 39383714 39383715 39383716 39383717 39383718 39383719 39389110 39389112 39389113 39391679 39391680 39391682 39391683 39391686 39391687 39392115 39392120 39392124 39392125 39392126 39392127 39392128 39392841 39392842 39392844 39392848 39392850 39392851 39392852 39392853 39392854 39392855 39392856 39392857 39394294 39394295 39394296 39394298 39394299 39394300 39394302 39394304 39396783 39396784 39396785 39396786 39396788 39396792 39396795 39396796 39396798 39396799 39396802 39396804 39396807 39399975 39399977 39399978 39399979 39399980 39399981 39399983 39399985 39400046 39400048 39400167 39410379 39410380 39410381 39410382 39410385 39410386 39410389 39410391 39410393 39410394 39410395 39410396 39410397 39410398 39410400 39410402 39410403 39410404 39410405 39410406 39410407 39410408 39410410 39410411 39410412 39410413 39410415 39410416 39410417 39410418 39410419 39410420 39410421 39410422 39410423 39410425 39410426 39410427 39410428 39410430 39410431 39410432 39410433 39410434 39410435 39410436 39412578 39415108 39415109 39415110 39415111 39415119 39415122 39415123 39415127 39415128 39415134 39415137 39415143 39415144 39415145 39415148 39415149 39415150 39415151 39415152 39415153 39415154 39415155 39419836 39419837 39419838 39419839 39419840 39419853 39419854 39419856 39419857 39419858 39419859 39419860 39420144 39420145 39420146 39420147 39420148 39420149 39420150 39420151 39420152 39420153 39420155 39420156 39420157 39420158 39420161 39420163 39420165 39420166 39420167 39420168 39420171 39420172 39420177 39420178 39423274 39423275 39423276 39423277 39423636 39423637 39423639 39423644 39423647 39423653 39423656 39423657 39423658 39423659 39423660 39423662 39423663 39423667 39423669 39423670 39423671 39423674 39423680 39423681 39423683 39423685 39423686 39423687 39423688 39423691 39423692 39423693 39423694 39423695 39423696 39423697 39423698 39426098 39426099 39426100 39426101 39426103 39426104 39426105 39426106 39426107 39426108 39426109 39426110 39426111 39426112 39426113 39426114 39426115 39426116 39426117 39426118 39426119 39426120 39426121 39426122 39427826 39427829 39427830 39427831 39427832 39427836 39427837 39427840 39427842 39427843 39427844 39427845 39427846 39427848 39427849 39430831 39430832 39430833 39430836 39430837 39430838 39430839 39430840 39430842 39430843 39430844 39430845 39430846 39430847 39430848 39430849 39430850 39430851 39430852 39430853 39430854 39430855 39430857 39430858 39430860 39430861 39430862 39430863 39430864 39430865 39430867 39430868 39430869 39430870 39430872 39430877 39430878 39430879 39430880 39430881 39430882 39430883 39430884 39430885 39430886 39430887 39430888 39430889 39430890 39430891 39430892 39430893 39430894 39430895 39430896 39430898 39430899 39430900 39430901 39430903 39430904 39430905 39430906 39430907 39430908 39430909 39430910 39430911 39430912 39430914 39430915 39430916 39430917 39430918 39430919 39430920 39430921 39430922 39430923 39431817 39431819 39436350 39436351 39436352 39437060 39438722 39438723 39439857 39439860 39439861 39439862 39439863 39442605 39442606 39442607 39442608 39452801 39452804 39461900 39461902 39461903 39461905 39461906 39463873 39463876 39463877 39463878 39463879 39463880 39463881 39463882 39463885 39463886 39463887 39463889 39463890 39463893 39463898 39463899 39463902 39463903 39463904 39463905 39463906 39463907 39463908 39463910 39463911 39463913 39463914 39463917 39463918 39463919 39465797 39465798 39465800 39465801 39465804 39465805 39465806 39465809 39465814 39465815 39465817 39468698 39468700 39468702 39468704 39468705 39468706 39468707 39468708 39468709 39469259 39475013 39475014 39475015 39475016 39475117 39476003 39476092 39476097 39476098 39476100 39476101 39476103 39476105 39476674 39476675 39476676 39476678 39476690 39476695 39476696 39476699 39476700 39476701 39477013 39477014 39477015 39477017 39477019 39477020 39477332 39477335 39477336 39477337 39492354 39519587 39519737 39519738 39519739 39519740 39519915 39519916 39519917 39519918 39519919 39519920 39519921 39519922 39519923 39519924 39519964 39519965 39519966 39519967 39519968 39519969 39519970 39519971 39519972 39519973 39520124 39520125 39520126 39520127 39520128 39520129 39520130 39520131 39520132 39520133 39520155 39520156 39520157 39520158 39520159 39520160 39520370 39520373 39520376 39520525 39520526 39520531 39524167 39524639 39524642 39524643 39524645 39524646 39524648 39524649 39524650 39524651 39524652 39524653 39524654 39524655 39524656 39524657 39525819 39526027 39526030 39526032 39526033 39526035 39526037 39526038 39526039 39526040 39526041 39526042 39526043 39526044 39526045 39526046 39526047 39527938 39527939 39527940 39527942 39527943 39527945 39527946 39527948 39527950 39527951 39527952 39527953 39527954 39527955 39527956 39527957 39528208 39528210 39528211 39528213 39528214 39528215 39528227 39528234 39528236 39528238 39528240 39528241 39528456 39528457 39528458 39528682 39529105 39529106 39529107 39529109 39529110 39529111 39529112 39529114 39529116 39529509 39529513 39529519 39529522 39529523 39529524 39529525 39529526 39529527 39529528 39529529 39529530 39529531 39529532 39529533 39529535 39530545 39530546 39530547 39530548 39530549 39530550 39530551 39530552 39530553 39530555 39530556 39530557 39530558 39530559 39530560 39530561 39530562 39530563 39530564 39530565 39530566 39530567 39530568 39530570 39530571 39530572 39530573 39530574 39530575 39531608 39531611 39531612 39531613 39531614 39531615 39531616 39531617 39531618 39531619 39531620 39531621 39531622 39531623 39531710 39531711 39531712 39531713 39531714 39531715 39531716 39531717 39531718 39531719 39531720 39531721 39531722 39531723 39531724 39536076 39536077 39536078 39536079 39536080 39536081 39536082 39536083 39536084 39536085 39536086 39536087 39536088 39536089 39536090 39536091 39536093 39536094 39536095 39536137 39536138 39536139 39536140 39536141 39536142 39536143 39536145 39536147 39536151 39536427 39536428 39536429 39536432 39536433 39536435 39536437 39536438 39536439 39536443 39536444 39536445 39536446 39536789 39536790 39536791 39536792 39536793 39536794 39536796 39537506 39537507 39537508 39537509 39537510 39537511 39537512 39537513 39537514 39537515 39537516 39537517 39537518 39537520 39537521 39537522 39537523 39537524 39537525 39538380 39538381 39538382 39538383 39538384 39538385 39538386 39538387 39538388 39538389 39538390 39538393 39538394 39538396 39538397 39538399 39538401 39538404 39538405 39541352 39541654 39541655 39541656 39541657 39541663 39541664 39541666 39541667 39541668 39541669 39541670 39541671 39541673 39541674 39541675 39541676 39541677 39541678 39541679 39541680 39541681 39542100 39543706 39563968 39563973 39564332 39573887 39573888 39573889 39573890 39573892 39573893 39573930 39573932 39573933 39573934 39578064 39578065 39578083 39578084 39578085 39578086 39578087 39578088 39578089 39578090 39578091 39578092 39579344 39579345 39579347 39583907 39583908 39583909 39583910 39583912 39583924 39583927 39583929 39583931 39583932 39583935 39583939 39584159 39584160 39584162 39584163 39584165 39584166 39584167 39584168 39584169 39584170 39584172 39584173 39584174 39584175 39584176 39584177 39584178 39586286 39586287 39586748 39586749 39586750 39586801 39586802 39586803 39586804 39586805 39586806 39586807 39586808 39586809 39586810 39586812 39587200 39587201 39587202 39587203 39587204 39587205 39587210 39587214 39587218 39587219 39587522 39587525 39587526 39587528 39587533 39587535 39587537 39587539 39587540 39587541 39590273 39590274 39590278 39590279 39590280 39590281 39590282 39591021 39591022 39591024 39591025 39591026 39591027 39591028 39591029 39591032 39591033 39591034 39591035 39591037 39591038 39591039 39591041 39591043 39591044 39593713 39593714 39593715 39593716 39593717 39593718 39593719 39593720 39593721 39593722 39593725 39593726 39593727 39593728 39593729 39593730 39593731 39594854 39594855 39594856 39594857 39594859 39594860 39594861 39594862 39594863 39594864 39595080 39595081 39595085 39595086 39595088 39595090 39595108 39595117 39595151 39595153 39595156 39595159 39595166 39595173 39596668 39596671 39596674 39596675 39596678 39596682 39596685 39596686 39596750 39596751 39596752 39596753 39596754 39596755 39596756 39596757 39596758 39596759 39597908 39597910 39597911 39597912 39597913 39597915 39598371 39598372 39598374 39598375 39598376 39598378 39598380 39605732 39605734 39605735 39605736 39605737 39605738 39605740 39605741 39605742 39605744 39605768 39605769 39605770 39605771 39605772 39605773 39605774 39605775 39605776 39605777 39605778 39605779 39605780 39605781 39605782 39607430 39607431 39607432 39607433 39607434 39607435 39607436 39607437 39607438 39607439 39607440 39607443 39607445 39607446 39607447 39607448 39607450 39607451 39607454 39607455 39607458 39607460 39607466 39608254 39608261 39608333 39608334 39608337 39608343 39608345 39608346 39608347 39608349 39608351 39608359 39608363 39608370 39608373 39608377 39608378 39608383 39608384 39608386 39608391 39608394 39608396 39608397 39610108 39610111 39610112 39610114 39610115 39610116 39610117 39610118 39610119 39610120 39610121 39610122 39610123 39610124 39610125 39610126 39610127 39611628 39611629 39611630 39611631 39611633 39611634 39611635 39611636 39611637 39611638 39611639 39611640 39611641 39611642 39611643 39611644 39611645 39611646 39611647 39612259 39612261 39612262 39612263 39612264 39612265 39612266 39612267 39612268 39612275 39612279 39612283 39612285 39612289 39612299 39612302 39612303 39612307 39612311 39612314 39612317 39612322 39612326 39613528 39613530 39613531 39613532 39613533 39613534 39613535 39613536 39613537 39613538 39613539 39613540 39613541 39613542 39613543 39613544 39613545 39613546 39613547 39615271 39615272 39615273 39615274 39615275 39615276 39615277 39615278 39615279 39615281 39615282 39615283 39615284 39615285 39615286 39615287 39615288 39615289 39615290 39615291 39615292 39615293 39615294 39615295 39615296 39615297 39615298 39615299 39615300 39615771 39615772 39615799 39615800 39615801 39615802 39615806 39615818 39615834 39615835 39615836 39615837 39615838 39615839 39615840 39615841 39615842 39615843 39615844 39615845 39615846 39615847 39615848 39615849 39615850 39615851 39615852 39615853 39616901 39616903 39616904 39616905 39616906 39616907 39616908 39616909 39616910 39616911 39616912 39616913 39616915 39616917 39616924 39616926 39618493 39618496 39618500 39618501 39618502 39618503 39618505 39618506 39618507 39618509 39618511 39618512 39618514 39618519 39619243 39621252 39621253 39621255 39621263 39621265 39621266 39621268 39621269 39621270 39621271 39622460 39623189 39623191 39623192 39625596 39625608 39625609 39625611 39625965 39625968 39625970 39625971 39625972 39625973 39625974 39625979 39625984 39625985 39625987 39625988 39625991 39625992 39626148 39626911 39627031 39628851 39628853 39628854 39628855 39628856 39628858 39632122 39632159 39635113 39635430 39635431 39643271 39643275 39643467 39643468 39643901 39643903 39643904 39644287 39647487 39647488 39647489 39647490 39647492 39647494 39647496 39647497 39647499 39647505 39647510 39647511 39647513 39647514 39647590 39647591 39648463 39648464 39648465 39648466 39648467 39648468 39648469 39648470 39648471 39648472 39648473 39648474 39648475 39648476 39648477 39648903 39648904 39648905 39648906 39648907 39648908 39648909 39648910 39648911 39648912 39648913 39648914 39648915 39648916 39648917 39648918 39648919 39648920 39648921 39648922 39649245 39649248 39649249 39649250 396

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0816-2020

Related Recalls

Due to required inspections not being performed on products/units that have gone through servicing.

Jul 22, 2025 Infusion Pumps Nationwide View Details →

Due to required inspections not being performed on products/units that have gone through servicing.

Jul 22, 2025 Diagnostic Equipment Nationwide View Details →

Due to required inspections not being performed on products/units that have gone through servicing.

Jul 22, 2025 Diagnostic Equipment Nationwide View Details →