1) The FEMTO LDV Z2 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery or other treatment requiring lamellar resection of the cornea at constant depth relative to the corneal surface. 2) The FEMTO LDV Z4 Femtosecond Surgical Laser is an opthalmic surgical laser indicated for use in the creation of the corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface that does not enclose a volume of the cornea. 3) The FEMTO LDV Z6 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants, lamellar keratoplasty or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface that may enclose a volume of the cornea.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: FLM4135, FLM3281, FLM2963, FLM3907, FLM3817
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ziemer Usa Inc
- Reason for Recall:
- A defect in the controller board can lead to the condenser to the video monitor becoming too warm. This could potentially cause the video monitor to shut off.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
1) The FEMTO LDV Z2 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery or other treatment requiring lamellar resection of the cornea at constant depth relative to the corneal surface. 2) The FEMTO LDV Z4 Femtosecond Surgical Laser is an opthalmic surgical laser indicated for use in the creation of the corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface that does not enclose a volume of the cornea. 3) The FEMTO LDV Z6 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants, lamellar keratoplasty or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface that may enclose a volume of the cornea.
Product Codes/Lot Numbers:
Serial Numbers: FLM4135, FLM3281, FLM2963, FLM3907, FLM3817
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0811-2014
Related Recalls
Femto LDV Z-Generation Femtosecond Surgical Lasers are used to perform ophthalmic surgeries.
Ziemer Usa