Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software Versions: 1.4.0 and 1.4.1. UDI-DI: 10841522107177, 10841522100246
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Spacelabs Healthcare, Inc.
- Reason for Recall:
- Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.
Product Codes/Lot Numbers:
Software Versions: 1.4.0 and 1.4.1. UDI-DI: 10841522107177, 10841522100246
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0810-2025
Related Recalls
Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.