Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Service Manuals P/N 070-2409-07 and Prior. UDI-DI: 10841522107177, 10841522100246
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Spacelabs Healthcare, Inc.
Reason for Recall:
Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.

Product Codes/Lot Numbers:

Service Manuals P/N 070-2409-07 and Prior. UDI-DI: 10841522107177, 10841522100246

Distribution:

Nationwide distribution

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0809-2025

Related Recalls

Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).

Mar 26, 2026 Diagnostic Equipment Nationwide View Details →

Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record

Sep 16, 2025 Other Medical Devices Nationwide View Details →

Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

Sep 5, 2025 Other Medical Devices Nationwide View Details →