SIGMA HP MBT Non-Keel Punch Knee Instrument. Designed to be used as an option in stabilizing the tibial trial during trial reduction.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Cat. No. GTIN Lot Description 950502016 10603295227366 ABB92402 SIGMA HP MBT Cemented Punch size 1-1.5 950502017 10603295227373 ABB84491 SIGMA HP MBT Cemented Punch size SZ 2-3 950502018 10603295227380 ABB92358 SIGMA HP MBT Cemented Punch size SZ 4-7 950502020 10603295227403 ABC33125 SIGMA HP MBT Noncemented Punch size 2-3 950502020 10603295227403 ABB81733 SIGMA HP MBT Noncemented Punch size SZ 2-3 950502020 10603295227403 ABB81732 SIGMA HP MBT Noncemented Punch size SZ 2-3 950502021 10603295227410 ABB39952 SIGMA HP MBT Noncemented Punch size SZ 4-7 950502021 10603295227410 ABC33123 SIGMA HP MBT Noncemented Punch size SZ 4-7
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- DePuy Orthopaedics, Inc.
- Reason for Recall:
- Increased punch height, causing a 0.063 in (1.6 mm) gap between the trial and the instrument. This may cause the surgeon to select an insert that is too thin. Surgical delay may result if the punch dislodges from the inserter and needs to be removed from the joint space. Insert Dislocation or Spin Out, or Poor Joint Mechanics may result, if not noted during surgery. May require revision surgery.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
SIGMA HP MBT Non-Keel Punch Knee Instrument. Designed to be used as an option in stabilizing the tibial trial during trial reduction.
Product Codes/Lot Numbers:
Cat. No. GTIN Lot Description 950502016 10603295227366 ABB92402 SIGMA HP MBT Cemented Punch size 1-1.5 950502017 10603295227373 ABB84491 SIGMA HP MBT Cemented Punch size SZ 2-3 950502018 10603295227380 ABB92358 SIGMA HP MBT Cemented Punch size SZ 4-7 950502020 10603295227403 ABC33125 SIGMA HP MBT Noncemented Punch size 2-3 950502020 10603295227403 ABB81733 SIGMA HP MBT Noncemented Punch size SZ 2-3 950502020 10603295227403 ABB81732 SIGMA HP MBT Noncemented Punch size SZ 2-3 950502021 10603295227410 ABB39952 SIGMA HP MBT Noncemented Punch size SZ 4-7 950502021 10603295227410 ABC33123 SIGMA HP MBT Noncemented Punch size SZ 4-7
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0787-2016
Related Recalls
ATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total Knee Replacement System.
DePuy Orthopaedics
Product incorrectly labelled.
Attune Revision Limb Preservation System (LPS) Tibial Insert XXSM 18MM, Part Number 151760118
DePuy Orthopaedics
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.
Attune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 151760312
DePuy Orthopaedics
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.