ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T NON-ECHO S PG, Item Number H787065970655, Catalog No.REF 06597065, 2) 4 F Stiffened Micro-Introducer Kit, STF 4F M.I. KIT 45CM NT/T NON-ECHO S PG, Item Number H787065970695, Catalog No.REF 06597069 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Batch/Lot: 1) 5066410, 2) 5029429, 5043441, 5074176
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Angiodynamics, Inc.
Reason for Recall:
Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T NON-ECHO S PG, Item Number H787065970655, Catalog No.REF 06597065, 2) 4 F Stiffened Micro-Introducer Kit, STF 4F M.I. KIT 45CM NT/T NON-ECHO S PG, Item Number H787065970695, Catalog No.REF 06597069 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick

Product Codes/Lot Numbers:

Batch/Lot: 1) 5066410, 2) 5029429, 5043441, 5074176

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0776-2017

Related Recalls

The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.

Mar 3, 2026 Surgical Instruments Nationwide View Details β†’