Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
Class I - DangerousWhat Should You Do?
- Check if you have this product: *************Serial Numbers of Units distributed in the US ************** 004329, 004330, 004349, 004360, 004361, 004362, 004363, 004364, 004365, 004366, 004367, 004369, 004370, 004371, 004417, 004421, 004426, 004436, 004437, 004443, 004448, 004449, 004454, 004456, 004473, 004481, 004488, 004490, 004494, 004495, 004502, 004507, 004517, 004523, 004525, 004526, 004527, 004528, 004530, 004533, 004539, 004541, 004544, 004545, 004546, 004548, 004552, 004553, 004554, 004555, 004556, 004557, 004558, 004567, 004568, 004573, 004578, 004581, 004582, 004584, 004588, 004591, 004602, 004608, 004630, 004632, 004637, 004640, 004647, 004659, 004667, 004668, 004670, 004672, 004673, 004677, 004678, 004679, 004681, 004683, 004684, 004685, 004686, 004693, 004694, 004696, 004700, 004706, 004707, 004708, 004709, 004710, 004711, 004712, 004713, 004714, 004715, 004716, 004717, 004718, 004719, 004720, 004721, 004722, 004723, 004724, 004725, 004726, 004727, 004728, 004729, 004730, 004731, 004732, 004733, 004734, 004735, 004736, 004737, 004738, 004739, 004740, 004741, 004742, 004743, 004745, 004746, 004747, 004749, 004750, 004753, 004754, 004756, 004757, 004760, 004761, 004762, 004763, 004764, 004766, 004767, 004768, 004771, 004772, 004773, 004774, 004775, 004776, 004777, 004778, 004779, 004780, 004781, 004782, 004783, 004784, 004785, 004787, 004788, 004789, 004790, 004791, 004793, 004794, 004795, 004796, 004797, 004798, 004800, 004801, 004802, 004803, 004804, 004805, 004806, 004807, 004808, 004810, 004811, 004812, 004814, 004815, 004816, 004818, 004819, 004820, 004822, 004823, 004824, 004825, 004826, 004827, 004828, 004829, 004830, 004831, 004832, 004833, 004834, 004835, 004836, 004837, 004838, 004839, 004840, 004841, 004842, 004843, 004844, 004845, 004846, 004847, 004848, 004849, 004850, 004851, 004852, 004853, 004854, 004855, 004856, 004857, 004858, 004859, 004860, 004861, 004862, 004863, 004864, 004865, 004866, 004867, 004868, 004869, 004870, 004871, 004872, 004873, 004874, 004876, 004877, 004878, 004879, 004880, 004881, 004882, 004884, 004885, 004886, 004887, 004888, 004890, 004891, 004893, 004894, 004895, 004896, 004897, 004899, 004900, 004901, 004902, 004903, 004904, 004905, 004906, 004907, 004909, 004910, 004997, 004998, 004999, 005007, 005008, 005009, 005010, 005011, 005012, 005013, 005014, 005015, 005016, 005017, 005018, 005019, 005020, 005021, 005022, 005023, 005024, 005025, 005026, 005027, 005028, 005029, 005030, 005031, 005032, 005033, 005034, 005035, 005036, 005037, 005039, 005040, 005041, 005042, 005043, 005044, 005045, 005046, 005047, 005048, 005049, 005050, 005051, 005052, 005053, 005054, 005055, 005056, 005057, 005058, 005059, 005060, 005061, 005062, 005063, 005064, 005065, 005066, 005067, 005068, 005069, 005070, 005071, 005072, 005073, 005074, 005075, 005076, 005077, 005078, 005079, 005080, 005081, 005082, 005083, 005084, 005085, 005086, 005087, 005088, 005089, 005090, 005091, 005092, 005093, 005094, 005095, 005096, 005097, 005098, 005099, 005100, 005101, 005103, 005104, 005105, 005106, 005107, 005108, 005112, 005113, 005115, 005117, 005118, 005119, 005120, 005121, 005122, 005123, 005124, 005125, 005128, 005136, 005137, 005138, 005139, 005140, 005141, 005142, 005143, 005144, 005145, 005146, 005147, 005148, 005149, 005150, 005151, 005152, and 005153. *************Serial Numbers of Units distributed in the US ************** 004283, 004284, 004285, 004286, 004287, 004288, 004290, 004291, 004292, 004293, 004294, 004295, 004296, 004298, 004299, 004302, 004303, 004305, 004306, 004307, 004308, 004309, 004310, 004312, 004313, 004314, 004315, 004316, 004319, 004320, 004321, 004322, 004325, 004326, 004327, 004328, 004332, 004333, 004334, 004335, 004337, 004338, 004339, 004340, 004341, 004343, 004344, 004345, 004347, 004348, 004350, 004351, 004353, 004354, 004355, 004356, 004357, 004359, 004372, 004373, 004374, 004375, 004379, 004380, 004381, 004382, 004383, 004384, 004385, 004386, 004388, 004389, 004390, 004391, 004392, 004393, 004394, 004395, 004396, 004397, 004399, 004400, 004401, 004402, 004403, 004404, 004405, 004406, 004408, 004411, 004412, 004413, 004414, 004415, 004416, 004418, 004419, 004420, 004422, 004423, 004424, 004425, 004427, 004428, 004430, 004431, 004432, 004433, 004439, 004440, 004442, 004445, 004446, 004447, 004450, 004451, 004452, 004453, 004455, 004457, 004458, 004459, 004460, 004462, 004463, 004464, 004465, 004466, 004467, 004468, 004469, 004471, 004472, 004474, 004475, 004476, 004477, 004478, 004479, 004480, 004484, 004485, 004486, 004489, 004491, 004492, 004496, 004497, 004498, 004499, 004500, 004501, 004503, 004504, 004505, 004508, 004510, 004511, 004512, 004513, 004514, 004515, 004516, 004518, 004519, 004520, 004521, 004522, 004524, 004531, 004532, 004536, 004542, 004543, 004549, 004551, 004565, 004569, 004570, 004571, 004572, 004574, 004575, 004576, 004577, 004585, 004587, 004589, 004590, 004592, 004593, 004594, 004596, 004597, 004598, 004599, 004600, 004601, 004604, 004605, 004606, 004609, 004610, 004611, 004612, 004613, 004614, 004615, 004616, 004617, 004618, 004619, 004620, 004621, 004622, 004623, 004624, 004625, 004626, 004628, 004629, 004631, 004633, 004634, 004635, 004638, 004639, 004641, 004642, 004643, 004644, 004645, 004646, 004648, 004649, 004650, 004651, 004652, 004653, 004655, 004656, 004657, 004658, 004660, 004662, 004663, 004664, 004665, 004666, 004669, 004671, 004674, 004675, 004680, 004682, 004688, 004689, 004690, 004692, 004695, 004697, 004698, 004699, 004701, 004702, 004703, 004704, 004705, 004744, 004748, 004751, 004752, 004755, 004758, 004765, 004769, 004770, 004792, 004799, 004809, 004813, 004817, 004821, 004875, 004889, 004898, 004908, 005102, 005111, 005114, 005116, 005126, 005127, 005129, 005130, 005131, 005132, 005133, 005134, 005135, 005154, 005155, 005239, 005240, 005241, 4317, 4323, 4324, 4368, and 4377.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Natus Medical Incorporated
- Reason for Recall:
- The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
Product Codes/Lot Numbers:
*************Serial Numbers of Units distributed in the US ************** 004329, 004330, 004349, 004360, 004361, 004362, 004363, 004364, 004365, 004366, 004367, 004369, 004370, 004371, 004417, 004421, 004426, 004436, 004437, 004443, 004448, 004449, 004454, 004456, 004473, 004481, 004488, 004490, 004494, 004495, 004502, 004507, 004517, 004523, 004525, 004526, 004527, 004528, 004530, 004533, 004539, 004541, 004544, 004545, 004546, 004548, 004552, 004553, 004554, 004555, 004556, 004557, 004558, 004567, 004568, 004573, 004578, 004581, 004582, 004584, 004588, 004591, 004602, 004608, 004630, 004632, 004637, 004640, 004647, 004659, 004667, 004668, 004670, 004672, 004673, 004677, 004678, 004679, 004681, 004683, 004684, 004685, 004686, 004693, 004694, 004696, 004700, 004706, 004707, 004708, 004709, 004710, 004711, 004712, 004713, 004714, 004715, 004716, 004717, 004718, 004719, 004720, 004721, 004722, 004723, 004724, 004725, 004726, 004727, 004728, 004729, 004730, 004731, 004732, 004733, 004734, 004735, 004736, 004737, 004738, 004739, 004740, 004741, 004742, 004743, 004745, 004746, 004747, 004749, 004750, 004753, 004754, 004756, 004757, 004760, 004761, 004762, 004763, 004764, 004766, 004767, 004768, 004771, 004772, 004773, 004774, 004775, 004776, 004777, 004778, 004779, 004780, 004781, 004782, 004783, 004784, 004785, 004787, 004788, 004789, 004790, 004791, 004793, 004794, 004795, 004796, 004797, 004798, 004800, 004801, 004802, 004803, 004804, 004805, 004806, 004807, 004808, 004810, 004811, 004812, 004814, 004815, 004816, 004818, 004819, 004820, 004822, 004823, 004824, 004825, 004826, 004827, 004828, 004829, 004830, 004831, 004832, 004833, 004834, 004835, 004836, 004837, 004838, 004839, 004840, 004841, 004842, 004843, 004844, 004845, 004846, 004847, 004848, 004849, 004850, 004851, 004852, 004853, 004854, 004855, 004856, 004857, 004858, 004859, 004860, 004861, 004862, 004863, 004864, 004865, 004866, 004867, 004868, 004869, 004870, 004871, 004872, 004873, 004874, 004876, 004877, 004878, 004879, 004880, 004881, 004882, 004884, 004885, 004886, 004887, 004888, 004890, 004891, 004893, 004894, 004895, 004896, 004897, 004899, 004900, 004901, 004902, 004903, 004904, 004905, 004906, 004907, 004909, 004910, 004997, 004998, 004999, 005007, 005008, 005009, 005010, 005011, 005012, 005013, 005014, 005015, 005016, 005017, 005018, 005019, 005020, 005021, 005022, 005023, 005024, 005025, 005026, 005027, 005028, 005029, 005030, 005031, 005032, 005033, 005034, 005035, 005036, 005037, 005039, 005040, 005041, 005042, 005043, 005044, 005045, 005046, 005047, 005048, 005049, 005050, 005051, 005052, 005053, 005054, 005055, 005056, 005057, 005058, 005059, 005060, 005061, 005062, 005063, 005064, 005065, 005066, 005067, 005068, 005069, 005070, 005071, 005072, 005073, 005074, 005075, 005076, 005077, 005078, 005079, 005080, 005081, 005082, 005083, 005084, 005085, 005086, 005087, 005088, 005089, 005090, 005091, 005092, 005093, 005094, 005095, 005096, 005097, 005098, 005099, 005100, 005101, 005103, 005104, 005105, 005106, 005107, 005108, 005112, 005113, 005115, 005117, 005118, 005119, 005120, 005121, 005122, 005123, 005124, 005125, 005128, 005136, 005137, 005138, 005139, 005140, 005141, 005142, 005143, 005144, 005145, 005146, 005147, 005148, 005149, 005150, 005151, 005152, and 005153. *************Serial Numbers of Units distributed in the US ************** 004283, 004284, 004285, 004286, 004287, 004288, 004290, 004291, 004292, 004293, 004294, 004295, 004296, 004298, 004299, 004302, 004303, 004305, 004306, 004307, 004308, 004309, 004310, 004312, 004313, 004314, 004315, 004316, 004319, 004320, 004321, 004322, 004325, 004326, 004327, 004328, 004332, 004333, 004334, 004335, 004337, 004338, 004339, 004340, 004341, 004343, 004344, 004345, 004347, 004348, 004350, 004351, 004353, 004354, 004355, 004356, 004357, 004359, 004372, 004373, 004374, 004375, 004379, 004380, 004381, 004382, 004383, 004384, 004385, 004386, 004388, 004389, 004390, 004391, 004392, 004393, 004394, 004395, 004396, 004397, 004399, 004400, 004401, 004402, 004403, 004404, 004405, 004406, 004408, 004411, 004412, 004413, 004414, 004415, 004416, 004418, 004419, 004420, 004422, 004423, 004424, 004425, 004427, 004428, 004430, 004431, 004432, 004433, 004439, 004440, 004442, 004445, 004446, 004447, 004450, 004451, 004452, 004453, 004455, 004457, 004458, 004459, 004460, 004462, 004463, 004464, 004465, 004466, 004467, 004468, 004469, 004471, 004472, 004474, 004475, 004476, 004477, 004478, 004479, 004480, 004484, 004485, 004486, 004489, 004491, 004492, 004496, 004497, 004498, 004499, 004500, 004501, 004503, 004504, 004505, 004508, 004510, 004511, 004512, 004513, 004514, 004515, 004516, 004518, 004519, 004520, 004521, 004522, 004524, 004531, 004532, 004536, 004542, 004543, 004549, 004551, 004565, 004569, 004570, 004571, 004572, 004574, 004575, 004576, 004577, 004585, 004587, 004589, 004590, 004592, 004593, 004594, 004596, 004597, 004598, 004599, 004600, 004601, 004604, 004605, 004606, 004609, 004610, 004611, 004612, 004613, 004614, 004615, 004616, 004617, 004618, 004619, 004620, 004621, 004622, 004623, 004624, 004625, 004626, 004628, 004629, 004631, 004633, 004634, 004635, 004638, 004639, 004641, 004642, 004643, 004644, 004645, 004646, 004648, 004649, 004650, 004651, 004652, 004653, 004655, 004656, 004657, 004658, 004660, 004662, 004663, 004664, 004665, 004666, 004669, 004671, 004674, 004675, 004680, 004682, 004688, 004689, 004690, 004692, 004695, 004697, 004698, 004699, 004701, 004702, 004703, 004704, 004705, 004744, 004748, 004751, 004752, 004755, 004758, 004765, 004769, 004770, 004792, 004799, 004809, 004813, 004817, 004821, 004875, 004889, 004898, 004908, 005102, 005111, 005114, 005116, 005126, 005127, 005129, 005130, 005131, 005132, 005133, 005134, 005135, 005154, 005155, 005239, 005240, 005241, 4317, 4323, 4324, 4368, and 4377.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0769-2017
Related Recalls
EDS 3 CSF External Drainage System, REF: NT821731C, STERILEEO, Rx only
Natus Medical Incorporated
Sterility assurance cannot be guaranteed for external drainage systems due to incomplete bioburden testing.
External Drainage System Collection Bag, REF: NT821732C, STERILEEO, Rx only
Natus Medical Incorporated
Sterility assurance cannot be guaranteed for external drainage systems due to incomplete bioburden testing.
Due to Cranial Access Kits containing recalled sterile surgical drapes. The sterile surgical drapes are being recalled due to the liner on the adhesive component is difficult to remove without damaging the product and may render the product unusable.