IVEA 500A Intended for Med-Surg patient care.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Product Part Number 16100. Serial Numbers 100200-101068, except for serial numbers within this range not released into final inventory for other reasons.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Firefly Medical, Inc.
Reason for Recall:
The dowel pin could potentially slip out of its press fit, resulting in the product potentially slumping to one side which could cause the patient to stumble.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

IVEA 500A Intended for Med-Surg patient care.

Product Codes/Lot Numbers:

Product Part Number 16100. Serial Numbers 100200-101068, except for serial numbers within this range not released into final inventory for other reasons.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0767-2017

Related Recalls

Class I - Dangerous

The dowel pins (P/N 16082) that fit in the left and right leg pivot castings (P/Ns 16049/16056) can rotate and gradually maneuver out of the leg casting after several hundred deployment and folding cycles (designed for easy storage) of the IVEA unit.

Aug 24, 2017 Infusion Pumps View Details →