MCA7-15SB - 1.5MM Fluted Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lots are included in recall
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- The Anspach Effort, Inc.
- Reason for Recall:
- The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
MCA7-15SB - 1.5MM Fluted Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.
Product Codes/Lot Numbers:
All lots are included in recall
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0764-2014
Related Recalls
Pediatric Craniotome Ref:CRANI-P-G1
The Anspach Effort
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
Adult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium
The Anspach Effort
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury