The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Programmer Model Number: 3650 UDI-DI code: 05414734509725 Software Version Number / Serial Numbers: 28.0.1 REV 2 12995 28.0.1 REV 2 16103 28.0.1 REV 2 30831 28.0.1 REV 2 12040778 28.0.1 REV 2 12041825 28.0.1 REV 2 12048943 28.0.1 REV 2 12050154 28.0.1 REV 2 12051352 28.0.1 REV 2 12051615 28.0.1 REV 2 12056133 28.0.1 REV 2 12713 28.0.1 REV 2 14655 28.0.1 REV 2 23324 28.0.1 REV 2 24050 28.0.1 REV 2 12042174 28.0.1 REV 2 12042239 28.0.1 REV 2 12048730 28.0.1 REV 2 11959 28.0.1 REV 2 16116 28.0.1 REV 2 16571 28.0.1 REV 2 27499 28.0.1 REV 2 12056125 28.0.1 REV 2 15378 28.0.1 REV 2 18041 28.0.1 REV 2 27164 28.0.1 REV 2 12045425 28.0.1 REV 2 12045483 28.0.1 REV 2 12870 28.0.1 REV 2 14910 28.0.1 REV 2 18143 28.0.1 REV 2 21904 28.0.1 REV 2 23317 28.0.1 REV 2 12043799 28.0.1 REV 2 12044806 28.0.1 REV 2 12055973 28.0.1 REV 2 24842 28.0.1 REV 2 12040739 28.0.1 REV 2 12042686 28.0.1 REV 2 12044212 28.0.1 REV 2 12045527 28.0.1 REV 2 12052141 28.0.1 REV 2 12055384 28.0.1 REV 2 12052537 28.0.1 REV 2 19660 28.0.1 REV 2 21099 28.0.1 REV 2 24610 28.0.1 REV 2 30022 28.0.1 REV 2 12044540 28.0.1 REV 2 12047076 28.0.1 REV 2 12053304 28.0.1 REV 2 12706 28.0.1 REV 2 20207 28.0.1 REV 2 20933 28.0.1 REV 2 22212 28.0.1 REV 2 25022 28.0.1 REV 2 27371 28.0.1 REV 2 12044759 28.0.1 REV 2 12045223 28.0.1 REV 2 12052322 28.0.1 REV 2 12056132 28.0.1 REV 2 12060627 28.0.1 REV 2 12061640 28.0.1 REV 2 11628 28.0.1 REV 2 11891 28.0.1 REV 2 20861 28.0.1 REV 2 22163 28.0.1 REV 2 25164 28.0.1 REV 2 25942 28.0.1 REV 2 26189 28.0.1 REV 2 12040940 28.0.1 REV 2 12041991 28.0.1 REV 2 12045260 28.0.1 REV 2 12046359 28.0.1 REV 2 12051613 28.0.1 REV 2 12052588 28.0.1 REV 2 12055272 28.0.1 REV 2 12061523 28.0.1 REV 2 12073 28.0.1 REV 2 12386 28.0.1 REV 2 21011 28.0.1 REV 2 22744 28.0.1 REV 2 23629 28.0.1 REV 2 23881 28.0.1 REV 2 12048206 28.0.1 REV 2 12048470 28.0.1 REV 2 12138 28.0.1 REV 2 12330 28.0.1 REV 2 16714 28.0.1 REV 2 20209 28.0.1 REV 2 21538 28.0.1 REV 2 23327 28.0.1 REV 2 24046 28.0.1 REV 2 25837 28.0.1 REV 2 12050165 28.0.1 REV 2 12051284 28.0.1 REV 2 12055859 28.0.1 REV 2 12055880 28.0.1 REV 2 16005 28.0.1 REV 2 12044210 28.0.1 REV 2 12044518 28.0.1 REV 2 12045375 28.0.1 REV 2 15043 28.0.1 REV 2 16127 28.0.1 REV 2 20025 28.0.1 REV 2 20956 28.0.1 REV 2 23113 28.0.1 REV 2 12042772 28.0.1 REV 2 12043605 28.0.1 REV 2 12048109 28.0.1 REV 2 11809 28.0.1 REV 2 16010 28.0.1 REV 2 25882 28.0.1 REV 2 26257 28.0.1 REV 2 27159 28.0.1 REV 2 12042806 28.0.1 REV 2 12061445 28.0.1 REV 2 11991 28.0.1 REV 2 12554 28.0.1 REV 2 16954 28.0.1 REV 2 17842 28.0.1 REV 2 20400 28.0.1 REV 2 26405 28.0.1 REV 2 12041928 28.0.1 REV 2 12044597 28.0.1 REV 2 12048812 28.0.1 REV 2 12060584 28.0.1 REV 2 24146 28.0.1 REV 2 12046557 28.0.1 REV 2 12048311 28.0.1 REV 2 12060407 28.0.1 REV 2 15175 28.0.1 REV 2 19180 28.0.1 REV 2 12045434 28.0.1 REV 2 12052162 28.0.1 REV 2 12055217 28.0.1 REV 2 11577 28.0.1 REV 2 16703 28.0.1 REV 2 16840 28.0.1 REV 2 19150 28.0.1 REV 2 26198 28.0.1 REV 2 26493 28.0.1 REV 2 28038 28.0.1 REV 2 12044032 28.0.1 REV 2 12050131 28.0.1 REV 2 12052590 28.0.1 REV 2 12055295 28.0.1 REV 2 12060629 28.0.1 REV 2 15257 28.0.1 REV 2 17658 28.0.1 REV 2 30735 28.0.1 REV 2 12045895 28.0.1 REV 2 12046696 28.0.1 REV 2 12047207 28.0.1 REV 2 12050175 28.0.1 REV 2 12051144 28.0.1 REV 2 22419 28.0.1 REV 2 22943 28.0.1 REV 2 30704 28.0.1 REV 2 12040805 28.0.1 REV 2 12044618 28.0.1 REV 2 12046363 28.0.1 REV 2 25194 28.0.1 REV 2 12045242 28.0.1 REV 2 12045276 28.0.1 REV 2 12051266 28.0.1 REV 2 11373 28.0.1 REV 2 18152 28.0.1 REV 2 12040924 28.0.1 REV 2 12041295 28.0.1 REV 2 22299 28.0.1 REV 2 24386 28.0.1 REV 2 24492 28.0.1 REV 2 25569 28.0.1 REV 2 12048312 28.0.1 REV 2 12048422 28.0.1 REV 2 12051199 28.0.1 REV 2 12052525 28.0.1 REV 2 11635 28.0.1 REV 2 17295 28.0.1 REV 2 20479 28.0.1 REV 2 23627 28.0.1 REV 2 12044434 28.0.1 REV 2 12045558 28.0.1 REV 2 12055864 28.0.1 REV 2 16105 28.0.1 REV 2 24762 28.0.1 REV 2 25147 28.0.1 REV 2 26011 28.0.1 REV 2 12041951 28.0.1 REV 2 12051079 28.0.1 REV 2 12051168 28.0.1 REV 2 12060736 28.0.1 REV 2 15926 28.0.1 REV 2 21177 28.0.1 REV 2 30016 28.0.1 REV 2 12060805 28.0.1 REV 2 16966 28.0.1 REV 2 20250 28.0.1 REV 2 20886 28.0.1 REV 2 21876 28.0.1 REV 2 24760 28.0.1 REV 2 26628 28.0.1 REV 2 27295 28.0.1 REV 2 27700 28.0.1 REV 2 27933 28.0.1 REV 2 12047177 28.0.1 REV 2 12048506 28.0.1 REV 2 12049432 28.0.1 REV 2 12060785 28.0.1 REV 2 24999 28.0.1 REV 2 12048294 28.0.1 REV 2 16828 28.0.1 REV 2 16907 28.0.1 REV 2 22241 28.0.1 REV 2 24884 28.0.1 REV 2 25075 28.0.1 REV 2 23091 28.0.1 REV 2 24696 28.0.1 REV 2 25999 28.0.1 REV 2 30823 28.0.1 REV 2 12045679 28.0.1 REV 2 12052118 28.0.1 REV 2 12052305 28.0.1 REV 2 18554 28.0.1 REV 2 20197 28.0.1 REV 2 21939 28.0.1 REV 2 23161 28.0.1 REV 2 25940 28.0.1 REV 2 12042222 28.0.1 REV 2 12046932 28.0.1 REV 2 12051416 28.0.1 REV 2 12055874 28.0.1 REV 2 19968 28.0.1 REV 2 21014 28.0.1 REV 2 25308 28.0.1 REV 2 25876 28.0.1 REV 2 12049933 28.0.1 REV 2 12061366 28.0.1 REV 2 21140 28.0.1 REV 2 24552 28.0.1 REV 2 12044482 28.0.1 REV 2 23466 28.0.1 REV 2 23422 28.0.1 REV 2 11688 28.0.1 REV 2 12052631 28.0.1 REV 2 25422 28.0.1 REV 2 12045682 28.0.1 REV 2 24946 28.0.1 REV 2 20876 28.0.1 REV 2 12049422 28.0.1 REV 2 25962 28.0.1 REV 2 12043694 28.0.1 REV 2 19241 28.0.1 REV 2 20042 28.0.1 REV 2 23762 28.0.1 REV 2 12048510 28.0.1 REV 2 12051108 28.0.1 REV 2 12061482 28.0.1 REV 2 12048450 28.0.1 REV 2 19699 28.0.1 REV 2 12052609 28.0.1 REV 2 14964 28.0.1 REV 2 21909 28.0.1 REV 3 15024 28.0.1 REV 3 15071 28.0.1 REV 3 15394 28.0.1 REV 3 17875 28.0.1 REV 3 12041047 28.0.1 REV 3 12042155 28.0.1 REV 3 14785 28.0.1 REV 3 14882 28.0.1 REV 3 15272 28.0.1 REV 3 15645 28.0.1 REV 3 19453 28.0.1 REV 3 21665 28.0.1 REV 3 23850 28.0.1 REV 3 25024 28.0.1 REV 3 12044251 28.0.1 REV 3 12052328 28.0.1 REV 3 11971 28.0.1 REV 3 16152 28.0.1 REV 3 20827 28.0.1 REV 3 27705 28.0.1 REV 3 12044657 28.0.1 REV 3 12049919 28.0.1 REV 3 12056081 28.0.1 REV 3 12061517 28.0.1 REV 3 12061540 28.0.1 REV 3 26178 28.0.1 REV 3 12041349 28.0.1 REV 3 14075 28.0.1 REV 3 15373 28.0.1 REV 3 19105 28.0.1 REV 3 19228 28.0.1 REV 3 21087 28.0.1 REV 3 24744 28.0.1 REV 3 12044142 28.0.1 REV 3 14007 28.0.1 REV 3 14782 28.0.1 REV 3 15928 28.0.1 REV 3 22448 28.0.1 REV 3 24561 28.0.1 REV 3 12041404 28.0.1 REV 3 12042086 28.0.1 REV 3 12043580 28.0.1 REV 3 12045905 28.0.1 REV 3 12048844 28.0.1 REV 3 12454 28.0.1 REV 3 14740 28.0.1 REV 3 15261 28.0.1 REV 3 20202 28.0.1 REV 3 26332 28.0.1 REV 3 27934 28.0.1 REV 3 12041063 28.0.1 REV 3 12048271 28.0.1 REV 3 12045488 28.0.1 REV 3 15423 28.0.1 REV 3 27848 28.0.1 REV 3 12045216 28.0.1 REV 3 12045453 28.0.1 REV 3 12048228 28.0.1 REV 3 11646 28.0.1 REV 3 23911 28.0.1 REV 3 26264 28.0.1 REV 3 12041335 28.0.1 REV 3 12051341 28.0.1 REV 3 12051667 28.0.1 REV 3 12060801 28.0.1 REV 3 11364 28.0.1 REV 3 12135 28.0.1 REV 3 14455 28.0.1 REV 3 12040818 28.0.1 REV 3 12040953 28.0.1 REV 3 12043618 28.0.1 REV 3 12046946 28.0.1 REV 3 19381 28.0.1 REV 3 16386 28.0.1 REV 3 16901 28.0.1 REV 3 12040798 28.0.1 REV 3 12051705 28.0.1 REV 3 11682 28.0.1 REV 3 20450 28.0.1 REV 3 24039 28.0.1 REV 3 12048318 28.0.1 REV 3 12054689 28.0.1 REV 3 12055869 28.0.1 REV 3 11371 28.0.1 REV 3 11708 28.0.1 REV 3 17280 28.0.1 REV 3 18084 28.0.1 REV 3 19840 28.0.1 REV 3 20380 28.0.1 REV 3 20459 28.0.1 REV 3 24679 28.0.1 REV 3 12042574 28.0.1 REV 3 12045237 28.0.1 REV 3 12050182 28.0.1 REV 3 12055311 28.0.1 REV 3 15305 28.0.1 REV 3 16595 28.0.1 REV 3 20483 28.0.1 REV 3 24217 28.0.1 REV 3 25972 28.0.1 REV 3 27666 28.0.1 REV 3 12056061 28.0.1 REV 3 11975 28.0.1 REV 3 16928 28.0.1 REV 3 20496 28.0.1 REV 3 20867 28.0.1 REV 3 25426 28.0.1 REV 3 26454 28.0.1 REV 3 12045804 28.0.1 REV 3 12051437 28.0.1 REV 3 12060033 28.0.1 REV 3 13668 28.0.1 REV 3 14794 28.0.1 REV 3 14983 28.0.1 REV 3 20065 28.0.1 REV 3 24357 28.0.1 REV 3 12045468 28.0.1 REV 3 17058 28.0.1 REV 3 19680 28.0.1 REV 3 27643 28.0.1 REV 3 12045570 28.0.1 REV 3 12050171 28.0.1 REV 3 12050187 28.0.1 REV 3 20188 28.0.1 REV 3 20442 28.0.1 REV 3 24027 28.0.1 REV 3 26432 28.0.1 REV 3 27592 28.0.1 REV 3 27743 28.0.1 REV 3 12049304 28.0.1 REV 3 15994 28.0.1 REV 3 28107 28.0.1 REV 3 12048078 28.0.1 REV 3 12060766 28.0.1 REV 3 18229 28.0.1 REV 3 22233 28.0.1 REV 3 23497 28.0.1 REV 3 25921 28.0.1 REV 3 30275 28.0.1 REV 3 12043589 28.0.1 REV 3 12044455 28.0.1 REV 3 18130 28.0.1 REV 3 23380 28.0.1 REV 3 23707 28.0.1 REV 3 12051098 28.0.1 REV 3 12055293 28.0.1 REV 3 20896 28.0.1 REV 3 23664 28.0.1 REV 3 27681 28.0.1 REV 3 12041300 28.0.1 REV 3 12044249 28.0.1 REV 3 12044699 28.0.1 REV 3 12051081 28.0.1 REV 3 12054754 28.0.1 REV 3 12055190 28.0.1 REV 3 12055550 28.0.1 REV 3 12398 28.0.1 REV 3 14676 28.0.1 REV 3 15073 28.0.1 REV 3 18068 28.0.1 REV 3 25127 28.0.1 REV 3 12042011 28.0.1 REV 3 12045140 28.0.1 REV 3 12045872 28.0.1 REV 3 12046244 28.0.1 REV 3 12052159 28.0.1 REV 3 12061378 28.0.1 REV 3 11790 28.0.1 REV 3 22079 28.0.1 REV 3 22797 28.0.1 REV 3 24471 28.0.1 REV 3 25938 28.0.1 REV 3 25968 28.0.1 REV 3 12041308 28.0.1 REV 3 12042662 28.0.1 REV 3 12045013 28.0.1 REV 3 12045721 28.0.1 REV 3 12048202 28.0.1 REV 3 12052134 28.0.1 REV 3 12054714 28.0.1 REV 3 12060802 28.0.1 REV 3 12469 28.0.1 REV 3 23798 28.0.1 REV 3 24089 28.0.1 REV 3 30709 28.0.1 REV 3 12044994 28.0.1 REV 3 12049444 28.0.1 REV 3 12051334 28.0.1 REV 3 11854 28.0.1 REV 3 17005 28.0.1 REV 3 19258 28.0.1 REV 3 21172 28.0.1 REV 3 23439 28.0.1 REV 3 24825 28.0.1 REV 3 27910 28.0.1 REV 3 12041331 28.0.1 REV 3 12043679 28.0.1 REV 3 12044986 28.0.1 REV 3 12048439 28.0.1 REV 3 12052263 28.0.1 REV 3 16702 28.0.1 REV 3 12044642 28.0.1 REV 3 12048466 28.0.1 REV 3 12060668 28.0.1 REV 3 15187 28.0.1 REV 3 23268 28.0.1 REV 3 24507 28.0.1 REV 3 12044610 28.0.1 REV 3 12052087 28.0.1 REV 3 12060663 28.0.1 REV 3 11863 28.0.1 REV 3 13959 28.0.1 REV 3 17266 28.0.1 REV 3 17741 28.0.1 REV 3 20356 28.0.1 REV 3 23582 28.0.1 REV 3 24413 28.0.1 REV 3 24972 28.0.1 REV 3 25773 28.0.1 REV 3 26112 28.0.1 REV 3 28029 28.0.1 REV 3 30012 28.0.1 REV 3 12050063 28.0.1 REV 3 15955 28.0.1 REV 3 25761 28.0.1 REV 3 26739 28.0.1 REV 3 30793 28.0.1 REV 3 12049901 28.0.1 REV 3 12051664 28.0.1 REV 3 12055810 28.0.1 REV 3 25265 28.0.1 REV 3 12055902 28.0.1 REV 3 22214 28.0.1 REV 3 22314 28.0.1 REV 3 12045485 28.0.1 REV 3 20005 28.0.1 REV 3 12046761 28.0.1 REV 3 12041886 28.0.1 REV 3 15004 28.0.1 REV 3 14988 28.0.1 REV 3 15513 28.0.1 REV 3 12381 28.0.1 REV 3 27344 28.0.1 REV 3 17640 28.0.1 REV 3 20222 28.0.1 REV 3 12429 28.0.1 REV 3 12044445 28.0.1 REV 3 19395 28.0.1 REV 3 26455 28.0.1 REV 3 19154 28.0.1 REV 3 12055956 28.0.1 REV 3 12041978 28.0.1 REV 3 12056082 28.0.1 REV 3 12042687 28.0.1 REV 3 27966 28.0.1 REV 3 14911 28.0.1 REV 3 17551 28.0.1 REV 3 12047002 28.0.1 REV 3 17462 28.0.1 REV 3 12055923 28.1.1 REV 1 18003 28.1.1 REV 1 19608 28.1.1 REV 1 23031 28.1.1 REV 1 25892 28.1.1 REV 1 12047062 28.1.1 REV 1 12053048 28.1.1 REV 1 12060701 28.1.1 REV 1 12310 28.1.1 REV 1 13817 28.1.1 REV 1 18019 28.1.1 REV 1 12043395 28.1.1 REV 1 12044648 28.1.1 REV 1 12045608 28.1.1 REV 1 12052959 28.1.1 REV 1 12053251 28.1.1 REV 1 12054669 28.1.1 REV 1 11819 28.1.1 REV 1 12152 28.1.1 REV 1 12206 28.1.1 REV 1 15031 28.1.1 REV 1 15476 28.1.1 REV 1 17871 28.1.1 REV 1 17912 28.1.1 REV 1 22167 28.1.1 REV 1 23984 28.1.1 REV 1 25969 28.1.1 REV 1 26712 28.1.1 REV 1 12040705 28.1.1 REV 1 12040809 28.1.1 REV 1 12042039 28.1.1 REV 1 12044422 28.1.1 REV 1 12048164 28.1.1 REV 1 12051084 28.1.1 REV 1 12051382 28.1.1 REV 1 12052034 28.1.1 REV 1 12052050 28.1.1 REV 1 12053203 28.1.1 REV 1 12005 28.1.1 REV 1 12564 28.1.1 REV 1 25820 28.1.1 REV 1 26419 28.1.1 REV 1 12042211 28.1.1 REV 1 12045204 28.1.1 REV 1 12045683 28.1.1 REV 1 12052633 28.1.1 REV 1 12256 28.1.1 REV 1 13627 28.1.1 REV 1 15312 28.1.1 REV 1 17877 28.1.1 REV 1 18020 28.1.1 REV 1 21117 28.1.1 REV 1 27039 28.1.1 REV 1 27673 28.1.1 REV 1 27913 28.1.1 REV 1 12041851 28.1.1 REV 1 12042084 28.1.1 REV 1 12046961 28.1.1 REV 1 12048566 28.1.1 REV 1 12051078 28.1.1 REV 1 12053189 28.1.1 REV 1 14952 28.1.1 REV 1 18057 28.1.1 REV 1 19093 28.1.1 REV 1 20076 28.1.1 REV 1 23687 28.1.1 REV 1 24797 28.1.1 REV 1 26345 28.1.1 REV 1 27152 28.1.1 REV 1 12042101 28.1.1 REV 1 12044565 28.1.1 REV 1 12048290 28.1.1 REV 1 20848 28.1.1 REV 1 23442 28.1.1 REV 1 25713 28.1.1 REV 1 30013 28.1.1 REV 1 12041867 28.1.1 REV 1 12044646 28.1.1 REV 1 12046545 28.1.1 REV 1 12047088 28.1.1 REV 1 12051391 28.1.1 REV 1 11886 28.1.1 REV 1 15176 28.1.1 REV 1 15323 28.1.1 REV 1 17288 28.1.1 REV 1 17499 28.1.1 REV 1 19537 28.1.1 REV 1 19577 28.1.1 REV 1 20317 28.1.1 REV 1 25692 28.1.1 REV 1 25849 28.1.1 REV 1 25878 28.1.1 REV 1 27500 28.1.1 REV 1 12041823 28.1.1 REV 1 12042668 28.1.1 REV 1 12043876 28.1.1 REV 1 12045277 28.1.1 REV 1 12045765 28.1.1 REV 1 12046291 28.1.1 REV 1 12048315 28.1.1 REV 1 12049278 28.1.1 REV 1 12051628 28.1.1 REV 1 12053335 28.1.1 REV 1 11746 28.1.1 REV 1 11760 28.1.1 REV 1 13842 28.1.1 REV 1 16730 28.1.1 REV 1 18573 28.1.1 REV 1 19363 28.1.1 REV 1 19484 28.1.1 REV 1 19638 28.1.1 REV 1 20012 28.1.1 REV 1 22306 28.1.1 REV 1 23950 28.1.1 REV 1 24820 28.1.1 REV 1 25833 28.1.1 REV 1 25872 28.1.1 REV 1 25924 28.1.1 REV 1 26393 28.1.1 REV 1 26656 28.1.1 REV 1 27018 28.1.1 REV 1 27101 28.1.1 REV 1 27141 28.1.1 REV 1 27322 28.1.1 REV 1 27631 28.1.1 REV 1 27803 28.1.1 REV 1 12040563 28.1.1 REV 1 12040847 28.1.1 REV 1 12043952 28.1.1 REV 1 12043959 28.1.1 REV 1 12045003 28.1.1 REV 1 12045238 28.1.1 REV 1 12047077 28.1.1 REV 1 12050195 28.1.1 REV 1 12051121 28.1.1 REV 1 12052022 28.1.1 REV 1 12311 28.1.1 REV 1 18894 28.1.1 REV 1 19440 28.1.1 REV 1 19980 28.1.1 REV 1 20870 28.1.1 REV 1 23573 28.1.1 REV 1 24194 28.1.1 REV 1 24455 28.1.1 REV 1 25129 28.1.1 REV 1 26453 28.1.1 REV 1 26611 28.1.1 REV 1 27038 28.1.1 REV 1 12042880 28.1.1 REV 1 12043603 28.1.1 REV 1 12044984 28.1.1 REV 1 12045047 28.1.1 REV 1 12048247 28.1.1 REV 1 12051126 28.1.1 REV 1 12052616 28.1.1 REV 1 14692 28.1.1 REV 1 18111 28.1.1 REV 1 20291 28.1.1 REV 1 20809 28.1.1 REV 1 21243 28.1.1 REV 1 22358 28.1.1 REV 1 23366 28.1.1 REV 1 26337 28.1.1 REV 1 12044069 28.1.1 REV 1 12061618 28.1.1 REV 1 11637 28.1.1 REV 1 12121 28.1.1 REV 1 12786 28.1.1 REV 1 13670 28.1.1 REV 1 19366 28.1.1 REV 1 20236 28.1.1 REV 1 21165 28.1.1 REV 1 22287 28.1.1 REV 1 24110 28.1.1 REV 1 12047215 28.1.1 REV 1 12052591 28.1.1 REV 1 13742 28.1.1 REV 1 19175 28.1.1 REV 1 19600 28.1.1 REV 1 19634 28.1.1 REV 1 23139 28.1.1 REV 1 23168 28.1.1 REV 1 24086 28.1.1 REV 1 24181 28.1.1 REV 1 26365 28.1.1 REV 1 27645 28.1.1 REV 1 12045810 28.1.1 REV 1 12047217 28.1.1 REV 1 12050191 28.1.1 REV 1 11513 28.1.1 REV 1 12234 28.1.1 REV 1 16330 28.1.1 REV 1 18096 28.1.1 REV 1 19503 28.1.1 REV 1 23715 28.1.1 REV 1 24690 28.1.1 REV 1 12042468 28.1.1 REV 1 12045677 28.1.1 REV 1 12045862 28.1.1 REV 1 11700 28.1.1 REV 1 13506 28.1.1 REV 1 13653 28.1.1 REV 1 15628 28.1.1 REV 1 17930 28.1.1 REV 1 20758 28.1.1 REV 1 21218 28.1.1 REV 1 22579 28.1.1 REV 1 23738 28.1.1 REV 1 25757 28.1.1 REV 1 25908 28.1.1 REV 1 26765 28.1.1 REV 1 26988 28.1.1 REV 1 12041700 28.1.1 REV 1 12041906 28.1.1 REV 1 12042654 28.1.1 REV 1 12042788 28.1.1 REV 1 12044245 28.1.1 REV 1 12050089 28.1.1 REV 1 12052036 28.1.1 REV 1 11387 28.1.1 REV 1 12149 28.1.1 REV 1 16416 28.1.1 REV 1 17843 28.1.1 REV 1 19470 28.1.1 REV 1 20003 28.1.1 REV 1 20205 28.1.1 REV 1 24233 28.1.1 REV 1 24568 28.1.1 REV 1 24883 28.1.1 REV 1 27372 28.1.1 REV 1 12044421 28.1.1 REV 1 12044602 28.1.1 REV 1 12046643 28.1.1 REV 1 12048902 28.1.1 REV 1 11862 28.1.1 REV 1 13999 28.1.1 REV 1 16198 28.1.1 REV 1 16266 28.1.1 REV 1 17634 28.1.1 REV 1 19912 28.1.1 REV 1 21112 28.1.1 REV 1 22996 28.1.1 REV 1 23109 28.1.1 REV 1 23188 28.1.1 REV 1 23204 28.1.1 REV 1 23944 28.1.1 REV 1 25385 28.1.1 REV 1 27845 28.1.1 REV 1 12043699 28.1.1 REV 1 12044203 28.1.1 REV 1 12046468 28.1.1 REV 1 12047639 28.1.1 REV 1 12048323 28.1.1 REV 1 12052358 28.1.1 REV 1 11653 28.1.1 REV 1 11756 28.1.1 REV 1 16323 28.1.1 REV 1 17751 28.1.1 REV 1 25023 28.1.1 REV 1 12041344 28.1.1 REV 1 12041989 28.1.1 REV 1 12043925 28.1.1 REV 1 12051648 28.1.1 REV 1 12132 28.1.1 REV 1 14748 28.1.1 REV 1 14771 28.1.1 REV 1 16442 28.1.1 REV 1 19263 28.1.1 REV 1 20473 28.1.1 REV 1 21823 28.1.1 REV 1 21912 28.1.1 REV 1 22507 28.1.1 REV 1 23192 28.1.1 REV 1 25342 28.1.1 REV 1 25817 28.1.1 REV 1 26002 28.1.1 REV 1 30678 28.1.1 REV 1 31052 28.1.1 REV 1 12040399 28.1.1 REV 1 12044417 28.1.1 REV 1 12046553 28.1.1 REV 1 12052095 28.1.1 REV 1 12054725 28.1.1 REV 1 12055198 28.1.1 REV 1 12437 28.1.1 REV 1 15060 28.1.1 REV 1 15199 28.1.1 REV 1 15277 28.1.1 REV 1 16244 28.1.1 REV 1 16290 28.1.1 REV 1 17011 28.1.1 REV 1 19223 28.1.1 REV 1 19433 28.1.1 REV 1 20314 28.1.1 REV 1 20379 28.1.1 REV 1 21849 28.1.1 REV 1 25509 28.1.1 REV 1 30799 28.1.1 REV 1 12045775 28.1.1 REV 1 12049483 28.1.1 REV 1 12052116 28.1.1 REV 1 11820 28.1.1 REV 1 12605 28.1.1 REV 1 14298 28.1.1 REV 1 15826 28.1.1 REV 1 17616 28.1.1 REV 1 22708 28.1.1 REV 1 23410 28.1.1 REV 1 23764 28.1.1 REV 1 26880 28.1.1 REV 1 12040678 28.1.1 REV 1 12041743 28.1.1 REV 1 12042022 28.1.1 REV 1 12044945 28.1.1 REV 1 12045827 28.1.1 REV 1 12046237 28.1.1 REV 1 12049456 28.1.1 REV 1 12052064 28.1.1 REV 1 12052163 28.1.1 REV 1 12052550 28.1.1 REV 1 12055900 28.1.1 REV 1 12061598 28.1.1 REV 1 11785 28.1.1 REV 1 12196 28.1.1 REV 1 12683 28.1.1 REV 1 15439 28.1.1 REV 1 20312 28.1.1 REV 1 20847 28.1.1 REV 1 20932 28.1.1 REV 1 22378 28.1.1 REV 1 24186 28.1.1 REV 1 25030 28.1.1 REV 1 25879 28.1.1 REV 1 25956 28.1.1 REV 1 26339 28.1.1 REV 1 26649 28.1.1 REV 1 27215 28.1.1 REV 1 30968 28.1.1 REV 1 12041299 28.1.1 REV 1 12041805 28.1.1 REV 1 12042670 28.1.1 REV 1 12042720 28.1.1 REV 1 12044147 28.1.1 REV 1 12046106 28.1.1 REV 1 12048257 28.1.1 REV 1 14675 28.1.1 REV 1 15374 28.1.1 REV 1 16785 28.1.1 REV 1 19505 28.1.1 REV 1 20386 28.1.1 REV 1 21758 28.1.1 REV 1 24526 28.1.1 REV 1 26424 28.1.1 REV 1 12040876 28.1.1 REV 1 12041031 28.1.1 REV 1 12042122 28.1.1 REV 1 12042220 28.1.1 REV 1 12044611 28.1.1 REV 1 12045669 28.1.1 REV 1 12061546 28.1.1 REV 1 11824 28.1.1 REV 1 11913 28.1.1 REV 1 12194 28.1.1 REV 1 12211 28.1.1 REV 1 13702 28.1.1 REV 1 17959 28.1.1 REV 1 19636 28.1.1 REV 1 20335 28.1.1 REV 1 20802 28.1.1 REV 1 21316 28.1.1 REV 1 21946 28.1.1 REV 1 22201 28.1.1 REV 1 23025 28.1.1 REV 1 23088 28.1.1 REV 1 23677 28.1.1 REV 1 24971 28.1.1 REV 1 26592 28.1.1 REV 1 27419 28.1.1 REV 1 27775 28.1.1 REV 1 12040943 28.1.1 REV 1 12042634 28.1.1 REV 1 12043593 28.1.1 REV 1 12044183 28.1.1 REV 1 12045728 28.1.1 REV 1 12057 28.1.1 REV 1 17549 28.1.1 REV 1 17964 28.1.1 REV 1 25781 28.1.1 REV 1 12043807 28.1.1 REV 1 12045435 28.1.1 REV 1 12050014 28.1.1 REV 1 12065 28.1.1 REV 1 17116 28.1.1 REV 1 17286 28.1.1 REV 1 19686 28.1.1 REV 1 20824 28.1.1 REV 1 23444 28.1.1 REV 1 24756 28.1.1 REV 1 24871 28.1.1 REV 1 24873 28.1.1 REV 1 12042204 28.1.1 REV 1 12043920 28.1.1 REV 1 12043937 28.1.1 REV 1 12045700 28.1.1 REV 1 12049429 28.1.1 REV 1 12051285 28.1.1 REV 1 12052131 28.1.1 REV 1 12053213 28.1.1 REV 1 12348 28.1.1 REV 1 12475 28.1.1 REV 1 12557 28.1.1 REV 1 16414 28.1.1 REV 1 16883 28.1.1 REV 1 20813 28.1.1 REV 1 23107 28.1.1 REV 1 23436 28.1.1 REV 1 23568 28.1.1 REV 1 23807 28.1.1 REV 1 24442 28.1.1 REV 1 24841 28.1.1 REV 1 27240 28.1.1 REV 1 27315 28.1.1 REV 1 27503 28.1.1 REV 1 12040719 28.1.1 REV 1 12041378 28.1.1 REV 1 12048170 28.1.1 REV 1 12052515 28.1.1 REV 1 15381 28.1.1 REV 1 18516 28.1.1 REV 1 19804 28.1.1 REV 1 20401 28.1.1 REV 1 21116 28.1.1 REV 1 21919 28.1.1 REV 1 23556 28.1.1 REV 1 25039 28.1.1 REV 1 12045207 28.1.1 REV 1 12052399 28.1.1 REV 1 12060735 28.1.1 REV 1 15425 28.1.1 REV 1 24726 28.1.1 REV 1 25626 28.1.1 REV 1 25867 28.1.1 REV 1 12041279 28.1.1 REV 1 12045195 28.1.1 REV 1 12055463 28.1.1 REV 1 12055792 28.1.1 REV 1 11353 28.1.1 REV 1 11522 28.1.1 REV 1 15442 28.1.1 REV 1 21039 28.1.1 REV 1 22231 28.1.1 REV 1 25314 28.1.1 REV 1 25674 28.1.1 REV 1 27923 28.1.1 REV 1 12042073 28.1.1 REV 1 12043911 28.1.1 REV 1 12048178 28.1.1 REV 1 12060662 28.1.1 REV 1 12450 28.1.1 REV 1 13729 28.1.1 REV 1 19585 28.1.1 REV 1 19676 28.1.1 REV 1 23102 28.1.1 REV 1 24243 28.1.1 REV 1 25429 28.1.1 REV 1 30900 28.1.1 REV 1 31047 28.1.1 REV 1 12040838 28.1.1 REV 1 12048258 28.1.1 REV 1 12049503 28.1.1 REV 1 12060330 28.1.1 REV 1 12368 28.1.1 REV 1 13923 28.1.1 REV 1 18182 28.1.1 REV 1 24012 28.1.1 REV 1 27796 28.1.1 REV 1 12045312 28.1.1 REV 1 12045441 28.1.1 REV 1 12045898 28.1.1 REV 1 12051269 28.1.1 REV 1 12060717 28.1.1 REV 1 11605 28.1.1 REV 1 16837 28.1.1 REV 1 23254 28.1.1 REV 1 23804 28.1.1 REV 1 26397 28.1.1 REV 1 26960 28.1.1 REV 1 27264 28.1.1 REV 1 12041407 28.1.1 REV 1 12042054 28.1.1 REV 1 12042785 28.1.1 REV 1 12044561 28.1.1 REV 1 12045037 28.1.1 REV 1 12045871 28.1.1 REV 1 12046951 28.1.1 REV 1 11933 28.1.1 REV 1 14587 28.1.1 REV 1 21096 28.1.1 REV 1 23592 28.1.1 REV 1 23737 28.1.1 REV 1 27637 28.1.1 REV 1 30806 28.1.1 REV 1 12042776 28.1.1 REV 1 12048406 28.1.1 REV 1 12048415 28.1.1 REV 1 12051380 28.1.1 REV 1 24574 28.1.1 REV 1 24285 28.1.1 REV 1 24607 28.1.1 REV 1 25984 28.1.1 REV 1 12045309 28.1.1 REV 1 11743 28.1.1 REV 1 12041962 28.1.1 REV 1 18595 28.1.1 REV 1 24462 28.1.1 REV 1 12041677 28.1.1 REV 1 12048389 28.1.1 REV 1 12048367 28.1.1 REV 1 12044021 28.1.1 REV 1 12052137 28.1.1 REV 1 19403 28.1.1 REV 1 12043949 28.1.1 REV 1 20034 28.1.1 REV 1 12055303 28.1.1 REV 1 24446 28.1.1 REV 1 16067 28.1.1 REV 1 19343 28.1.1 REV 1 25642 28.1.1 REV 1 12051376 28.1.1 REV 1 12044612 28.1.1 REV 1 25451 28.1.1 REV 1 11940 28.1.1 REV 1 18367 28.1.1 REV 1 12041227 28.1.1 REV 1 12055531 28.1.1 REV 1 24761 28.1.2 REV 1 24803 28.1.2 REV 1 12044767 28.1.2 REV 1 12042423 28.1.2 REV 1 14717 28.1.2 REV 1 12044309 28.1.2 REV 1 12044328 28.1.2 REV 1 12044341 28.1.2 REV 1 12044342 28.1.2 REV 1 15883 28.1.2 REV 1 19744 28.1.2 REV 1 22543 28.1.2 REV 1 22880 28.1.2 REV 1 22915 28.1.2 REV 1 13360 28.1.2 REV 1 14168 28.1.2 REV 1 14635 28.1.2 REV 1 14647 28.1.2 REV 1 14674 28.1.2 REV 1 15498 28.1.2 REV 1 15864 28.1.2 REV 1 19795 28.1.2 REV 1 22531 28.1.2 REV 1 12042702 28.1.2 REV 1 12051962 28.1.2 REV 1 13681 28.1.2 REV 1 14927 28.1.2 REV 1 15218 28.1.2 REV 1 30507 28.1.2 REV 1 10764 28.1.2 REV 1 12044392 28.1.2 REV 1 12044867 28.1.2 REV 1 12045423 28.1.2 REV 1 12045913 28.1.2 REV 1 12046177 28.1.2 REV 1 12048591 28.1.2 REV 1 12048976 28.1.2 REV 1 12049068 28.1.2 REV 1 12051449 28.1.2 REV 1 12051467 28.1.2 REV 1 12051883 28.1.2 REV 1 12053138 28.1.2 REV 1 12053158 28.1.2 REV 1 12053160 28.1.2 REV 1 12053199 28.1.2 REV 1 12053347 28.1.2 REV 1 12053848 28.1.2 REV 1 12054009 28.1.2 REV 1 12054143 28.1.2 REV 1 12056128 28.1.2 REV 1 12056769 28.1.2 REV 1 12649 28.1.2 REV 1 12658 28.1.2 REV 1 12659 28.1.2 REV 1 12692 28.1.2 REV 1 12958 28.1.2 REV 1 12980 28.1.2 REV 1 13046 28.1.2 REV 1 13278 28.1.2 REV 1 13473 28.1.2 REV 1 13905 28.1.2 REV 1 13984 28.1.2 REV 1 14097 28.1.2 REV 1 14157 28.1.2 REV 1 14341 28.1.2 REV 1 14371 28.1.2 REV 1 14435 28.1.2 REV 1 14625 28.1.2 REV 1 14659 28.1.2 REV 1 14684 28.1.2 REV 1 14727 28.1.2 REV 1 14729 28.1.2 REV 1 14738 28.1.2 REV 1 14753 28.1.2 REV 1 14774 28.1.2 REV 1 14822 28.1.2 REV 1 14829 28.1.2 REV 1 14849 28.1.2 REV 1 14864 28.1.2 REV 1 14865 28.1.2 REV 1 14885 28.1.2 REV 1 15816 28.1.2 REV 1 15859 28.1.2 REV 1 16108 28.1.2 REV 1 16215 28.1.2 REV 1 16259 28.1.2 REV 1 16301 28.1.2 REV 1 18283 28.1.2 REV 1 18405 28.1.2 REV 1 18406 28.1.2 REV 1 18411 28.1.2 REV 1 18459 28.1.2 REV 1 19010 28.1.2 REV 1 19047 28.1.2 REV 1 19050 28.1.2 REV 1 19061 28.1.2 REV 1 19085 28.1.2 REV 1 19601 28.1.2 REV 1 19707 28.1.2 REV 1 19738 28.1.2 REV 1 20377 28.1.2 REV 1 21455 28.1.2 REV 1 21463 28.1.2 REV 1 21464 28.1.2 REV 1 21470 28.1.2 REV 1 21910 28.1.2 REV 1 22093 28.1.2 REV 1 22496 28.1.2 REV 1 22523 28.1.2 REV 1 22558 28.1.2 REV 1 22693 28.1.2 REV 1 22727 28.1.2 REV 1 22861 28.1.2 REV 1 22936 28.1.2 REV 1 22957 28.1.2 REV 1 12044620 28.1.2 REV 1 12047443 28.1.2 REV 1 13404 28.1.2 REV 1 16677 28.1.2 REV 1 23625 28.1.2 REV 1 12043772 28.1.2 REV 1 12049939 28.1.2 REV 1 14063 28.1.2 REV 1 12043726 28.1.2 REV 1 12053011 28.1.2 REV 1 12054373 28.1.2 REV 1 12051235 28.1.2 REV 1 14454 28.1.2 REV 1 17734 28.1.2 REV 1 18358 28.1.2 REV 1 25707 28.1.2 REV 1 30466 28.1.2 REV 1 12042305 28.1.2 REV 1 13000 28.1.2 REV 1 13013 28.1.2 REV 1 14303 28.1.2 REV 1 22716 28.1.2 REV 1 24734 28.1.2 REV 1 27478 28.1.2 REV 1 12045038 28.1.2 REV 1 12046723 28.1.2 REV 1 12049959 28.1.2 REV 1 18874 28.1.2 REV 1 12045951 28.1.2 REV 1 12998 28.1.2 REV 1 13253 28.1.2 REV 1 22481 28.1.2 REV 1 26688 28.1.2 REV 1 12049860 28.1.2 REV 1 12050777 28.1.2 REV 1 12053016 28.1.2 REV 1 22853 28.1.2 REV 1 26691 28.1.2 REV 1 30142 28.1.2 REV 1 12047605 28.1.2 REV 1 12049716 28.1.2 REV 1 12052915 28.1.2 REV 1 13248 28.1.2 REV 1 14240 28.1.2 REV 1 14870 28.1.2 REV 1 20532 28.1.2 REV 1 20600 28.1.2 REV 1 21353 28.1.2 REV 1 21459 28.1.2 REV 1 22533 28.1.2 REV 1 25519 28.1.2 REV 1 12046161 28.1.2 REV 1 12047428 28.1.2 REV 1 12047546 28.1.2 REV 1 12047790 28.1.2 REV 1 12049064 28.1.2 REV 1 12049827 28.1.2 REV 1 12049867 28.1.2 REV 1 12049969 28.1.2 REV 1 12050829 28.1.2 REV 1 12050854 28.1.2 REV 1 12052317 28.1.2 REV 1 12053039 28.1.2 REV 1 12053498 28.1.2 REV 1 12053593 28.1.2 REV 1 13309 28.1.2 REV 1 15278 28.1.2 REV 1 19736 28.1.2 REV 1 22436 28.1.2 REV 1 22799 28.1.2 REV 1 22823 28.1.2 REV 1 30506 28.1.2 REV 1 12043717 28.1.2 REV 1 12044863 28.1.2 REV 1 12047186 28.1.2 REV 1 12047606 28.1.2 REV 1 12048606 28.1.2 REV 1 12048617 28.1.2 REV 1 12048675 28.1.2 REV 1 12049880 28.1.2 REV 1 12052285 28.1.2 REV 1 12052464 28.1.2 REV 1 12052480 28.1.2 REV 1 12052911 28.1.2 REV 1 12053699 28.1.2 REV 1 12055967 28.1.2 REV 1 12056201 28.1.2 REV 1 12056450 28.1.2 REV 1 12062539 28.1.2 REV 1 12272 28.1.2 REV 1 12883 28.1.2 REV 1 12989 28.1.2 REV 1 13572 28.1.2 REV 1 14037 28.1.2 REV 1 14098 28.1.2 REV 1 14174 28.1.2 REV 1 14552 28.1.2 REV 1 14860 28.1.2 REV 1 14931 28.1.2 REV 1 14998 28.1.2 REV 1 15391 28.1.2 REV 1 15782 28.1.2 REV 1 17351 28.1.2 REV 1 17792 28.1.2 REV 1 19663 28.1.2 REV 1 22425 28.1.2 REV 1 22890 28.1.2 REV 1 23529 28.1.2 REV 1 24266 28.1.2 REV 1 24454 28.1.2 REV 1 28133 28.1.2 REV 1 12041092 28.1.2 REV 1 12041136 28.1.2 REV 1 12042283 28.1.2 REV 1 12042295 28.1.2 REV 1 12042440 28.1.2 REV 1 12042491 28.1.2 REV 1 12042548 28.1.2 REV 1 12042845 28.1.2 REV 1 12042891 28.1.2 REV 1 12043170 28.1.2 REV 1 12043299 28.1.2 REV 1 12044211 28.1.2 REV 1 12044222 28.1.2 REV 1 12044289 28.1.2 REV 1 12044357 28.1.2 REV 1 12044798 28.1.2 REV 1 12044858 28.1.2 REV 1 12044882 28.1.2 REV 1 12045948 28.1.2 REV 1 12046030 28.1.2 REV 1 12046183 28.1.2 REV 1 12046441 28.1.2 REV 1 12046528 28.1.2 REV 1 12046788 28.1.2 REV 1 12046838 28.1.2 REV 1 12046859 28.1.2 REV 1 12047125 28.1.2 REV 1 12047147 28.1.2 REV 1 12047259 28.1.2 REV 1 12047272 28.1.2 REV 1 12047289 28.1.2 REV 1 12047349 28.1.2 REV 1 12047372 28.1.2 REV 1 12047560 28.1.2 REV 1 12047567 28.1.2 REV 1 12047593 28.1.2 REV 1 12047597 28.1.2 REV 1 12047705 28.1.2 REV 1 12047710 28.1.2 REV 1 12047796 28.1.2 REV 1 12047808 28.1.2 REV 1 12048577 28.1.2 REV 1 12048667 28.1.2 REV 1 12048938 28.1.2 REV 1 12048982 28.1.2 REV 1 12049008 28.1.2 REV 1 12049010 28.1.2 REV 1 12049036 28.1.2 REV 1 12049044 28.1.2 REV 1 12049047 28.1.2 REV 1 12049051 28.1.2 REV 1 12049076 28.1.2 REV 1 12049658 28.1.2 REV 1 12049703 28.1.2 REV 1 12049748 28.1.2 REV 1 12049803 28.1.2 REV 1 12049872 28.1.2 REV 1 12049927 28.1.2 REV 1 12049967 28.1.2 REV 1 12050273 28.1.2 REV 1 12050278 28.1.2 REV 1 12050298 28.1.2 REV 1 12050305 28.1.2 REV 1 12050311 28.1.2 REV 1 12050359 28.1.2 REV 1 12050606 28.1.2 REV 1 12050738 28.1.2 REV 1 12050770 28.1.2 REV 1 12050799 28.1.2 REV 1 12050800 28.1.2 REV 1 12050831 28.1.2 REV 1 12050843 28.1.2 REV 1 12050867 28.1.2 REV 1 12051477 28.1.2 REV 1 12051479 28.1.2 REV 1 12051485 28.1.2 REV 1 12051512 28.1.2 REV 1 12051515 28.1.2 REV 1 12051533 28.1.2 REV 1 12052306 28.1.2 REV 1 12052404 28.1.2 REV 1 12052455 28.1.2 REV 1 12052894 28.1.2 REV 1 12052905 28.1.2 REV 1 12052917 28.1.2 REV 1 12053071 28.1.2 REV 1 12053183 28.1.2 REV 1 12053210 28.1.2 REV 1 12053314 28.1.2 REV 1 12053667 28.1.2 REV 1 12053674 28.1.2 REV 1 12054375 28.1.2 REV 1 12054379 28.1.2 REV 1 12054414 28.1.2 REV 1 12054415 28.1.2 REV 1 12054432 28.1.2 REV 1 12054464 28.1.2 REV 1 12054490 28.1.2 REV 1 12054985 28.1.2 REV 1 12055607 28.1.2 REV 1 12055626 28.1.2 REV 1 12055787 28.1.2 REV 1 12055818 28.1.2 REV 1 12055847 28.1.2 REV 1 12056066 28.1.2 REV 1 12056247 28.1.2 REV 1 12056372 28.1.2 REV 1 12056446 28.1.2 REV 1 12056457 28.1.2 REV 1 12056482 28.1.2 REV 1 12056551 28.1.2 REV 1 12056577 28.1.2 REV 1 12056578 28.1.2 REV 1 12056656 28.1.2 REV 1 12056690 28.1.2 REV 1 12056750 28.1.2 REV 1 12062166 28.1.2 REV 1 12590 28.1.2 REV 1 12769 28.1.2 REV 1 12792 28.1.2 REV 1 12823 28.1.2 REV 1 12832 28.1.2 REV 1 12882 28.1.2 REV 1 12895 28.1.2 REV 1 12915 28.1.2 REV 1 13011 28.1.2 REV 1 13037 28.1.2 REV 1 13038 28.1.2 REV 1 13262 28.1.2 REV 1 13301 28.1.2 REV 1 13303 28.1.2 REV 1 13306 28.1.2 REV 1 13312 28.1.2 REV 1 13324 28.1.2 REV 1 13325 28.1.2 REV 1 13334 28.1.2 REV 1 13338 28.1.2 REV 1 13350 28.1.2 REV 1 13388 28.1.2 REV 1 13432 28.1.2 REV 1 13436 28.1.2 REV 1 13585 28.1.2 REV 1 13605 28.1.2 REV 1 13634 28.1.2 REV 1 13641 28.1.2 REV 1 13692 28.1.2 REV 1 14162 28.1.2 REV 1 14178 28.1.2 REV 1 14186 28.1.2 REV 1 14212 28.1.2 REV 1 14217 28.1.2 REV 1 14237 28.1.2 REV 1 14268 28.1.2 REV 1 14378 28.1.2 REV 1 15027 28.1.2 REV 1 15638 28.1.2 REV 1 15686 28.1.2 REV 1 15691 28.1.2 REV 1 15697 28.1.2 REV 1 15861 28.1.2 REV 1 15890 28.1.2 REV 1 15907 28.1.2 REV 1 18395 28.1.2 REV 1 18570 28.1.2 REV 1 18593 28.1.2 REV 1 18597 28.1.2 REV 1 18598 28.1.2 REV 1 18610 28.1.2 REV 1 18611 28.1.2 REV 1 19023 28.1.2 REV 1 19055 28.1.2 REV 1 19734 28.1.2 REV 1 19735 28.1.2 REV 1 19746 28.1.2 REV 1 19791 28.1.2 REV 1 20455 28.1.2 REV 1 20494 28.1.2 REV 1 20495 28.1.2 REV 1 20512 28.1.2 REV 1 20570 28.1.2 REV 1 20576 28.1.2 REV 1 20587 28.1.2 REV 1 20588 28.1.2 REV 1 20632 28.1.2 REV 1 20672 28.1.2 REV 1 20675 28.1.2 REV 1 21378 28.1.2 REV 1 21475 28.1.2 REV 1 21476 28.1.2 REV 1 22367 28.1.2 REV 1 22376 28.1.2 REV 1 22403 28.1.2 REV 1 22420 28.1.2 REV 1 22427 28.1.2 REV 1 22429 28.1.2 REV 1 22445 28.1.2 REV 1 22449 28.1.2 REV 1 22480 28.1.2 REV 1 22524 28.1.2 REV 1 22537 28.1.2 REV 1 22556 28.1.2 REV 1 22698 28.1.2 REV 1 22699 28.1.2 REV 1 22722 28.1.2 REV 1 22749 28.1.2 REV 1 22790 28.1.2 REV 1 22807 28.1.2 REV 1 22808 28.1.2 REV 1 22813 28.1.2 REV 1 22843 28.1.2 REV 1 22860 28.1.2 REV 1 22894 28.1.2 REV 1 22895 28.1.2 REV 1 22900 28.1.2 REV 1 22927 28.1.2 REV 1 22962 28.1.2 REV 1 22965 28.1.2 REV 1 22966 28.1.2 REV 1 22977 28.1.2 REV 1 22979 28.1.2 REV 1 23005 28.1.2 REV 1 23044 28.1.2 REV 1 23046 28.1.2 REV 1 23059 28.1.2 REV 1 23958 28.1.2 REV 1 24783 28.1.2 REV 1 24995 28.1.2 REV 1 25530 28.1.2 REV 1 25591 28.1.2 REV 1 25612 28.1.2 REV 1 25618 28.1.2 REV 1 26113 28.1.2 REV 1 26142 28.1.2 REV 1 26868 28.1.2 REV 1 27075 28.1.2 REV 1 27095 28.1.2 REV 1 27113 28.1.2 REV 1 27116 28.1.2 REV 1 27127 28.1.2 REV 1 27134 28.1.2 REV 1 27150 28.1.2 REV 1 27182 28.1.2 REV 1 27193 28.1.2 REV 1 27618 28.1.2 REV 1 30051 28.1.2 REV 1 30077 28.1.2 REV 1 30086 28.1.2 REV 1 30089 28.1.2 REV 1 30095 28.1.2 REV 1 30097 28.1.2 REV 1 30102 28.1.2 REV 1 30143 28.1.2 REV 1 30205 28.1.2 REV 1 30217 28.1.2 REV 1 30244 28.1.2 REV 1 30249 28.1.2 REV 1 30251 28.1.2 REV 1 30324 28.1.2 REV 1 30359 28.1.2 REV 1 30441 28.1.2 REV 1 30465 28.1.2 REV 1 30488 28.1.2 REV 1 12063954 28.1.2 REV 1 12064191 28.1.2 REV 1 12064518 28.1.2 REV 1 12042321 28.1.2 REV 1 12043633 28.1.2 REV 1 12046735 28.1.2 REV 1 12046883 28.1.2 REV 1 12047373 28.1.2 REV 1 12048494 28.1.2 REV 1 12049053 28.1.2 REV 1 12049717 28.1.2 REV 1 12051932 28.1.2 REV 1 12052483 28.1.2 REV 1 12055744 28.1.2 REV 1 12056040 28.1.2 REV 1 12056245 28.1.2 REV 1 12056647 28.1.2 REV 1 12120 28.1.2 REV 1 13029 28.1.2 REV 1 13040 28.1.2 REV 1 14129 28.1.2 REV 1 14175 28.1.2 REV 1 14881 28.1.2 REV 1 15543 28.1.2 REV 1 15696 28.1.2 REV 1 15760 28.1.2 REV 1 16768 28.1.2 REV 1 17033 28.1.2 REV 1 17537 28.1.2 REV 1 17546 28.1.2 REV 1 17577 28.1.2 REV 1 17583 28.1.2 REV 1 17764 28.1.2 REV 1 18097 28.1.2 REV 1 18181 28.1.2 REV 1 18304 28.1.2 REV 1 18964 28.1.2 REV 1 19107 28.1.2 REV 1 22070 28.1.2 REV 1 22714 28.1.2 REV 1 22759 28.1.2 REV 1 22960 28.1.2 REV 1 23395 28.1.2 REV 1 24548 28.1.2 REV 1 25124 28.1.2 REV 1 25372 28.1.2 REV 1 25634 28.1.2 REV 1 26061 28.1.2 REV 1 12044878 28.1.2 REV 1 12044927 28.1.2 REV 1 12822 28.1.2 REV 1 13255 28.1.2 REV 1 14696 28.1.2 REV 1 26951 28.1.2 REV 1 12043027 28.1.2 REV 1 12045350 28.1.2 REV 1 12049091 28.1.2 REV 1 12050269 28.1.2 REV 1 12051420 28.1.2 REV 1 12053020 28.1.2 REV 1 12054239 28.1.2 REV 1 17081 28.1.2 REV 1 17145 28.1.2 REV 1 17996 28.1.2 REV 1 20
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Abbott Medical
Reason for Recall:
Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.

Product Codes/Lot Numbers:

Programmer Model Number: 3650 UDI-DI code: 05414734509725 Software Version Number / Serial Numbers: 28.0.1 REV 2 12995 28.0.1 REV 2 16103 28.0.1 REV 2 30831 28.0.1 REV 2 12040778 28.0.1 REV 2 12041825 28.0.1 REV 2 12048943 28.0.1 REV 2 12050154 28.0.1 REV 2 12051352 28.0.1 REV 2 12051615 28.0.1 REV 2 12056133 28.0.1 REV 2 12713 28.0.1 REV 2 14655 28.0.1 REV 2 23324 28.0.1 REV 2 24050 28.0.1 REV 2 12042174 28.0.1 REV 2 12042239 28.0.1 REV 2 12048730 28.0.1 REV 2 11959 28.0.1 REV 2 16116 28.0.1 REV 2 16571 28.0.1 REV 2 27499 28.0.1 REV 2 12056125 28.0.1 REV 2 15378 28.0.1 REV 2 18041 28.0.1 REV 2 27164 28.0.1 REV 2 12045425 28.0.1 REV 2 12045483 28.0.1 REV 2 12870 28.0.1 REV 2 14910 28.0.1 REV 2 18143 28.0.1 REV 2 21904 28.0.1 REV 2 23317 28.0.1 REV 2 12043799 28.0.1 REV 2 12044806 28.0.1 REV 2 12055973 28.0.1 REV 2 24842 28.0.1 REV 2 12040739 28.0.1 REV 2 12042686 28.0.1 REV 2 12044212 28.0.1 REV 2 12045527 28.0.1 REV 2 12052141 28.0.1 REV 2 12055384 28.0.1 REV 2 12052537 28.0.1 REV 2 19660 28.0.1 REV 2 21099 28.0.1 REV 2 24610 28.0.1 REV 2 30022 28.0.1 REV 2 12044540 28.0.1 REV 2 12047076 28.0.1 REV 2 12053304 28.0.1 REV 2 12706 28.0.1 REV 2 20207 28.0.1 REV 2 20933 28.0.1 REV 2 22212 28.0.1 REV 2 25022 28.0.1 REV 2 27371 28.0.1 REV 2 12044759 28.0.1 REV 2 12045223 28.0.1 REV 2 12052322 28.0.1 REV 2 12056132 28.0.1 REV 2 12060627 28.0.1 REV 2 12061640 28.0.1 REV 2 11628 28.0.1 REV 2 11891 28.0.1 REV 2 20861 28.0.1 REV 2 22163 28.0.1 REV 2 25164 28.0.1 REV 2 25942 28.0.1 REV 2 26189 28.0.1 REV 2 12040940 28.0.1 REV 2 12041991 28.0.1 REV 2 12045260 28.0.1 REV 2 12046359 28.0.1 REV 2 12051613 28.0.1 REV 2 12052588 28.0.1 REV 2 12055272 28.0.1 REV 2 12061523 28.0.1 REV 2 12073 28.0.1 REV 2 12386 28.0.1 REV 2 21011 28.0.1 REV 2 22744 28.0.1 REV 2 23629 28.0.1 REV 2 23881 28.0.1 REV 2 12048206 28.0.1 REV 2 12048470 28.0.1 REV 2 12138 28.0.1 REV 2 12330 28.0.1 REV 2 16714 28.0.1 REV 2 20209 28.0.1 REV 2 21538 28.0.1 REV 2 23327 28.0.1 REV 2 24046 28.0.1 REV 2 25837 28.0.1 REV 2 12050165 28.0.1 REV 2 12051284 28.0.1 REV 2 12055859 28.0.1 REV 2 12055880 28.0.1 REV 2 16005 28.0.1 REV 2 12044210 28.0.1 REV 2 12044518 28.0.1 REV 2 12045375 28.0.1 REV 2 15043 28.0.1 REV 2 16127 28.0.1 REV 2 20025 28.0.1 REV 2 20956 28.0.1 REV 2 23113 28.0.1 REV 2 12042772 28.0.1 REV 2 12043605 28.0.1 REV 2 12048109 28.0.1 REV 2 11809 28.0.1 REV 2 16010 28.0.1 REV 2 25882 28.0.1 REV 2 26257 28.0.1 REV 2 27159 28.0.1 REV 2 12042806 28.0.1 REV 2 12061445 28.0.1 REV 2 11991 28.0.1 REV 2 12554 28.0.1 REV 2 16954 28.0.1 REV 2 17842 28.0.1 REV 2 20400 28.0.1 REV 2 26405 28.0.1 REV 2 12041928 28.0.1 REV 2 12044597 28.0.1 REV 2 12048812 28.0.1 REV 2 12060584 28.0.1 REV 2 24146 28.0.1 REV 2 12046557 28.0.1 REV 2 12048311 28.0.1 REV 2 12060407 28.0.1 REV 2 15175 28.0.1 REV 2 19180 28.0.1 REV 2 12045434 28.0.1 REV 2 12052162 28.0.1 REV 2 12055217 28.0.1 REV 2 11577 28.0.1 REV 2 16703 28.0.1 REV 2 16840 28.0.1 REV 2 19150 28.0.1 REV 2 26198 28.0.1 REV 2 26493 28.0.1 REV 2 28038 28.0.1 REV 2 12044032 28.0.1 REV 2 12050131 28.0.1 REV 2 12052590 28.0.1 REV 2 12055295 28.0.1 REV 2 12060629 28.0.1 REV 2 15257 28.0.1 REV 2 17658 28.0.1 REV 2 30735 28.0.1 REV 2 12045895 28.0.1 REV 2 12046696 28.0.1 REV 2 12047207 28.0.1 REV 2 12050175 28.0.1 REV 2 12051144 28.0.1 REV 2 22419 28.0.1 REV 2 22943 28.0.1 REV 2 30704 28.0.1 REV 2 12040805 28.0.1 REV 2 12044618 28.0.1 REV 2 12046363 28.0.1 REV 2 25194 28.0.1 REV 2 12045242 28.0.1 REV 2 12045276 28.0.1 REV 2 12051266 28.0.1 REV 2 11373 28.0.1 REV 2 18152 28.0.1 REV 2 12040924 28.0.1 REV 2 12041295 28.0.1 REV 2 22299 28.0.1 REV 2 24386 28.0.1 REV 2 24492 28.0.1 REV 2 25569 28.0.1 REV 2 12048312 28.0.1 REV 2 12048422 28.0.1 REV 2 12051199 28.0.1 REV 2 12052525 28.0.1 REV 2 11635 28.0.1 REV 2 17295 28.0.1 REV 2 20479 28.0.1 REV 2 23627 28.0.1 REV 2 12044434 28.0.1 REV 2 12045558 28.0.1 REV 2 12055864 28.0.1 REV 2 16105 28.0.1 REV 2 24762 28.0.1 REV 2 25147 28.0.1 REV 2 26011 28.0.1 REV 2 12041951 28.0.1 REV 2 12051079 28.0.1 REV 2 12051168 28.0.1 REV 2 12060736 28.0.1 REV 2 15926 28.0.1 REV 2 21177 28.0.1 REV 2 30016 28.0.1 REV 2 12060805 28.0.1 REV 2 16966 28.0.1 REV 2 20250 28.0.1 REV 2 20886 28.0.1 REV 2 21876 28.0.1 REV 2 24760 28.0.1 REV 2 26628 28.0.1 REV 2 27295 28.0.1 REV 2 27700 28.0.1 REV 2 27933 28.0.1 REV 2 12047177 28.0.1 REV 2 12048506 28.0.1 REV 2 12049432 28.0.1 REV 2 12060785 28.0.1 REV 2 24999 28.0.1 REV 2 12048294 28.0.1 REV 2 16828 28.0.1 REV 2 16907 28.0.1 REV 2 22241 28.0.1 REV 2 24884 28.0.1 REV 2 25075 28.0.1 REV 2 23091 28.0.1 REV 2 24696 28.0.1 REV 2 25999 28.0.1 REV 2 30823 28.0.1 REV 2 12045679 28.0.1 REV 2 12052118 28.0.1 REV 2 12052305 28.0.1 REV 2 18554 28.0.1 REV 2 20197 28.0.1 REV 2 21939 28.0.1 REV 2 23161 28.0.1 REV 2 25940 28.0.1 REV 2 12042222 28.0.1 REV 2 12046932 28.0.1 REV 2 12051416 28.0.1 REV 2 12055874 28.0.1 REV 2 19968 28.0.1 REV 2 21014 28.0.1 REV 2 25308 28.0.1 REV 2 25876 28.0.1 REV 2 12049933 28.0.1 REV 2 12061366 28.0.1 REV 2 21140 28.0.1 REV 2 24552 28.0.1 REV 2 12044482 28.0.1 REV 2 23466 28.0.1 REV 2 23422 28.0.1 REV 2 11688 28.0.1 REV 2 12052631 28.0.1 REV 2 25422 28.0.1 REV 2 12045682 28.0.1 REV 2 24946 28.0.1 REV 2 20876 28.0.1 REV 2 12049422 28.0.1 REV 2 25962 28.0.1 REV 2 12043694 28.0.1 REV 2 19241 28.0.1 REV 2 20042 28.0.1 REV 2 23762 28.0.1 REV 2 12048510 28.0.1 REV 2 12051108 28.0.1 REV 2 12061482 28.0.1 REV 2 12048450 28.0.1 REV 2 19699 28.0.1 REV 2 12052609 28.0.1 REV 2 14964 28.0.1 REV 2 21909 28.0.1 REV 3 15024 28.0.1 REV 3 15071 28.0.1 REV 3 15394 28.0.1 REV 3 17875 28.0.1 REV 3 12041047 28.0.1 REV 3 12042155 28.0.1 REV 3 14785 28.0.1 REV 3 14882 28.0.1 REV 3 15272 28.0.1 REV 3 15645 28.0.1 REV 3 19453 28.0.1 REV 3 21665 28.0.1 REV 3 23850 28.0.1 REV 3 25024 28.0.1 REV 3 12044251 28.0.1 REV 3 12052328 28.0.1 REV 3 11971 28.0.1 REV 3 16152 28.0.1 REV 3 20827 28.0.1 REV 3 27705 28.0.1 REV 3 12044657 28.0.1 REV 3 12049919 28.0.1 REV 3 12056081 28.0.1 REV 3 12061517 28.0.1 REV 3 12061540 28.0.1 REV 3 26178 28.0.1 REV 3 12041349 28.0.1 REV 3 14075 28.0.1 REV 3 15373 28.0.1 REV 3 19105 28.0.1 REV 3 19228 28.0.1 REV 3 21087 28.0.1 REV 3 24744 28.0.1 REV 3 12044142 28.0.1 REV 3 14007 28.0.1 REV 3 14782 28.0.1 REV 3 15928 28.0.1 REV 3 22448 28.0.1 REV 3 24561 28.0.1 REV 3 12041404 28.0.1 REV 3 12042086 28.0.1 REV 3 12043580 28.0.1 REV 3 12045905 28.0.1 REV 3 12048844 28.0.1 REV 3 12454 28.0.1 REV 3 14740 28.0.1 REV 3 15261 28.0.1 REV 3 20202 28.0.1 REV 3 26332 28.0.1 REV 3 27934 28.0.1 REV 3 12041063 28.0.1 REV 3 12048271 28.0.1 REV 3 12045488 28.0.1 REV 3 15423 28.0.1 REV 3 27848 28.0.1 REV 3 12045216 28.0.1 REV 3 12045453 28.0.1 REV 3 12048228 28.0.1 REV 3 11646 28.0.1 REV 3 23911 28.0.1 REV 3 26264 28.0.1 REV 3 12041335 28.0.1 REV 3 12051341 28.0.1 REV 3 12051667 28.0.1 REV 3 12060801 28.0.1 REV 3 11364 28.0.1 REV 3 12135 28.0.1 REV 3 14455 28.0.1 REV 3 12040818 28.0.1 REV 3 12040953 28.0.1 REV 3 12043618 28.0.1 REV 3 12046946 28.0.1 REV 3 19381 28.0.1 REV 3 16386 28.0.1 REV 3 16901 28.0.1 REV 3 12040798 28.0.1 REV 3 12051705 28.0.1 REV 3 11682 28.0.1 REV 3 20450 28.0.1 REV 3 24039 28.0.1 REV 3 12048318 28.0.1 REV 3 12054689 28.0.1 REV 3 12055869 28.0.1 REV 3 11371 28.0.1 REV 3 11708 28.0.1 REV 3 17280 28.0.1 REV 3 18084 28.0.1 REV 3 19840 28.0.1 REV 3 20380 28.0.1 REV 3 20459 28.0.1 REV 3 24679 28.0.1 REV 3 12042574 28.0.1 REV 3 12045237 28.0.1 REV 3 12050182 28.0.1 REV 3 12055311 28.0.1 REV 3 15305 28.0.1 REV 3 16595 28.0.1 REV 3 20483 28.0.1 REV 3 24217 28.0.1 REV 3 25972 28.0.1 REV 3 27666 28.0.1 REV 3 12056061 28.0.1 REV 3 11975 28.0.1 REV 3 16928 28.0.1 REV 3 20496 28.0.1 REV 3 20867 28.0.1 REV 3 25426 28.0.1 REV 3 26454 28.0.1 REV 3 12045804 28.0.1 REV 3 12051437 28.0.1 REV 3 12060033 28.0.1 REV 3 13668 28.0.1 REV 3 14794 28.0.1 REV 3 14983 28.0.1 REV 3 20065 28.0.1 REV 3 24357 28.0.1 REV 3 12045468 28.0.1 REV 3 17058 28.0.1 REV 3 19680 28.0.1 REV 3 27643 28.0.1 REV 3 12045570 28.0.1 REV 3 12050171 28.0.1 REV 3 12050187 28.0.1 REV 3 20188 28.0.1 REV 3 20442 28.0.1 REV 3 24027 28.0.1 REV 3 26432 28.0.1 REV 3 27592 28.0.1 REV 3 27743 28.0.1 REV 3 12049304 28.0.1 REV 3 15994 28.0.1 REV 3 28107 28.0.1 REV 3 12048078 28.0.1 REV 3 12060766 28.0.1 REV 3 18229 28.0.1 REV 3 22233 28.0.1 REV 3 23497 28.0.1 REV 3 25921 28.0.1 REV 3 30275 28.0.1 REV 3 12043589 28.0.1 REV 3 12044455 28.0.1 REV 3 18130 28.0.1 REV 3 23380 28.0.1 REV 3 23707 28.0.1 REV 3 12051098 28.0.1 REV 3 12055293 28.0.1 REV 3 20896 28.0.1 REV 3 23664 28.0.1 REV 3 27681 28.0.1 REV 3 12041300 28.0.1 REV 3 12044249 28.0.1 REV 3 12044699 28.0.1 REV 3 12051081 28.0.1 REV 3 12054754 28.0.1 REV 3 12055190 28.0.1 REV 3 12055550 28.0.1 REV 3 12398 28.0.1 REV 3 14676 28.0.1 REV 3 15073 28.0.1 REV 3 18068 28.0.1 REV 3 25127 28.0.1 REV 3 12042011 28.0.1 REV 3 12045140 28.0.1 REV 3 12045872 28.0.1 REV 3 12046244 28.0.1 REV 3 12052159 28.0.1 REV 3 12061378 28.0.1 REV 3 11790 28.0.1 REV 3 22079 28.0.1 REV 3 22797 28.0.1 REV 3 24471 28.0.1 REV 3 25938 28.0.1 REV 3 25968 28.0.1 REV 3 12041308 28.0.1 REV 3 12042662 28.0.1 REV 3 12045013 28.0.1 REV 3 12045721 28.0.1 REV 3 12048202 28.0.1 REV 3 12052134 28.0.1 REV 3 12054714 28.0.1 REV 3 12060802 28.0.1 REV 3 12469 28.0.1 REV 3 23798 28.0.1 REV 3 24089 28.0.1 REV 3 30709 28.0.1 REV 3 12044994 28.0.1 REV 3 12049444 28.0.1 REV 3 12051334 28.0.1 REV 3 11854 28.0.1 REV 3 17005 28.0.1 REV 3 19258 28.0.1 REV 3 21172 28.0.1 REV 3 23439 28.0.1 REV 3 24825 28.0.1 REV 3 27910 28.0.1 REV 3 12041331 28.0.1 REV 3 12043679 28.0.1 REV 3 12044986 28.0.1 REV 3 12048439 28.0.1 REV 3 12052263 28.0.1 REV 3 16702 28.0.1 REV 3 12044642 28.0.1 REV 3 12048466 28.0.1 REV 3 12060668 28.0.1 REV 3 15187 28.0.1 REV 3 23268 28.0.1 REV 3 24507 28.0.1 REV 3 12044610 28.0.1 REV 3 12052087 28.0.1 REV 3 12060663 28.0.1 REV 3 11863 28.0.1 REV 3 13959 28.0.1 REV 3 17266 28.0.1 REV 3 17741 28.0.1 REV 3 20356 28.0.1 REV 3 23582 28.0.1 REV 3 24413 28.0.1 REV 3 24972 28.0.1 REV 3 25773 28.0.1 REV 3 26112 28.0.1 REV 3 28029 28.0.1 REV 3 30012 28.0.1 REV 3 12050063 28.0.1 REV 3 15955 28.0.1 REV 3 25761 28.0.1 REV 3 26739 28.0.1 REV 3 30793 28.0.1 REV 3 12049901 28.0.1 REV 3 12051664 28.0.1 REV 3 12055810 28.0.1 REV 3 25265 28.0.1 REV 3 12055902 28.0.1 REV 3 22214 28.0.1 REV 3 22314 28.0.1 REV 3 12045485 28.0.1 REV 3 20005 28.0.1 REV 3 12046761 28.0.1 REV 3 12041886 28.0.1 REV 3 15004 28.0.1 REV 3 14988 28.0.1 REV 3 15513 28.0.1 REV 3 12381 28.0.1 REV 3 27344 28.0.1 REV 3 17640 28.0.1 REV 3 20222 28.0.1 REV 3 12429 28.0.1 REV 3 12044445 28.0.1 REV 3 19395 28.0.1 REV 3 26455 28.0.1 REV 3 19154 28.0.1 REV 3 12055956 28.0.1 REV 3 12041978 28.0.1 REV 3 12056082 28.0.1 REV 3 12042687 28.0.1 REV 3 27966 28.0.1 REV 3 14911 28.0.1 REV 3 17551 28.0.1 REV 3 12047002 28.0.1 REV 3 17462 28.0.1 REV 3 12055923 28.1.1 REV 1 18003 28.1.1 REV 1 19608 28.1.1 REV 1 23031 28.1.1 REV 1 25892 28.1.1 REV 1 12047062 28.1.1 REV 1 12053048 28.1.1 REV 1 12060701 28.1.1 REV 1 12310 28.1.1 REV 1 13817 28.1.1 REV 1 18019 28.1.1 REV 1 12043395 28.1.1 REV 1 12044648 28.1.1 REV 1 12045608 28.1.1 REV 1 12052959 28.1.1 REV 1 12053251 28.1.1 REV 1 12054669 28.1.1 REV 1 11819 28.1.1 REV 1 12152 28.1.1 REV 1 12206 28.1.1 REV 1 15031 28.1.1 REV 1 15476 28.1.1 REV 1 17871 28.1.1 REV 1 17912 28.1.1 REV 1 22167 28.1.1 REV 1 23984 28.1.1 REV 1 25969 28.1.1 REV 1 26712 28.1.1 REV 1 12040705 28.1.1 REV 1 12040809 28.1.1 REV 1 12042039 28.1.1 REV 1 12044422 28.1.1 REV 1 12048164 28.1.1 REV 1 12051084 28.1.1 REV 1 12051382 28.1.1 REV 1 12052034 28.1.1 REV 1 12052050 28.1.1 REV 1 12053203 28.1.1 REV 1 12005 28.1.1 REV 1 12564 28.1.1 REV 1 25820 28.1.1 REV 1 26419 28.1.1 REV 1 12042211 28.1.1 REV 1 12045204 28.1.1 REV 1 12045683 28.1.1 REV 1 12052633 28.1.1 REV 1 12256 28.1.1 REV 1 13627 28.1.1 REV 1 15312 28.1.1 REV 1 17877 28.1.1 REV 1 18020 28.1.1 REV 1 21117 28.1.1 REV 1 27039 28.1.1 REV 1 27673 28.1.1 REV 1 27913 28.1.1 REV 1 12041851 28.1.1 REV 1 12042084 28.1.1 REV 1 12046961 28.1.1 REV 1 12048566 28.1.1 REV 1 12051078 28.1.1 REV 1 12053189 28.1.1 REV 1 14952 28.1.1 REV 1 18057 28.1.1 REV 1 19093 28.1.1 REV 1 20076 28.1.1 REV 1 23687 28.1.1 REV 1 24797 28.1.1 REV 1 26345 28.1.1 REV 1 27152 28.1.1 REV 1 12042101 28.1.1 REV 1 12044565 28.1.1 REV 1 12048290 28.1.1 REV 1 20848 28.1.1 REV 1 23442 28.1.1 REV 1 25713 28.1.1 REV 1 30013 28.1.1 REV 1 12041867 28.1.1 REV 1 12044646 28.1.1 REV 1 12046545 28.1.1 REV 1 12047088 28.1.1 REV 1 12051391 28.1.1 REV 1 11886 28.1.1 REV 1 15176 28.1.1 REV 1 15323 28.1.1 REV 1 17288 28.1.1 REV 1 17499 28.1.1 REV 1 19537 28.1.1 REV 1 19577 28.1.1 REV 1 20317 28.1.1 REV 1 25692 28.1.1 REV 1 25849 28.1.1 REV 1 25878 28.1.1 REV 1 27500 28.1.1 REV 1 12041823 28.1.1 REV 1 12042668 28.1.1 REV 1 12043876 28.1.1 REV 1 12045277 28.1.1 REV 1 12045765 28.1.1 REV 1 12046291 28.1.1 REV 1 12048315 28.1.1 REV 1 12049278 28.1.1 REV 1 12051628 28.1.1 REV 1 12053335 28.1.1 REV 1 11746 28.1.1 REV 1 11760 28.1.1 REV 1 13842 28.1.1 REV 1 16730 28.1.1 REV 1 18573 28.1.1 REV 1 19363 28.1.1 REV 1 19484 28.1.1 REV 1 19638 28.1.1 REV 1 20012 28.1.1 REV 1 22306 28.1.1 REV 1 23950 28.1.1 REV 1 24820 28.1.1 REV 1 25833 28.1.1 REV 1 25872 28.1.1 REV 1 25924 28.1.1 REV 1 26393 28.1.1 REV 1 26656 28.1.1 REV 1 27018 28.1.1 REV 1 27101 28.1.1 REV 1 27141 28.1.1 REV 1 27322 28.1.1 REV 1 27631 28.1.1 REV 1 27803 28.1.1 REV 1 12040563 28.1.1 REV 1 12040847 28.1.1 REV 1 12043952 28.1.1 REV 1 12043959 28.1.1 REV 1 12045003 28.1.1 REV 1 12045238 28.1.1 REV 1 12047077 28.1.1 REV 1 12050195 28.1.1 REV 1 12051121 28.1.1 REV 1 12052022 28.1.1 REV 1 12311 28.1.1 REV 1 18894 28.1.1 REV 1 19440 28.1.1 REV 1 19980 28.1.1 REV 1 20870 28.1.1 REV 1 23573 28.1.1 REV 1 24194 28.1.1 REV 1 24455 28.1.1 REV 1 25129 28.1.1 REV 1 26453 28.1.1 REV 1 26611 28.1.1 REV 1 27038 28.1.1 REV 1 12042880 28.1.1 REV 1 12043603 28.1.1 REV 1 12044984 28.1.1 REV 1 12045047 28.1.1 REV 1 12048247 28.1.1 REV 1 12051126 28.1.1 REV 1 12052616 28.1.1 REV 1 14692 28.1.1 REV 1 18111 28.1.1 REV 1 20291 28.1.1 REV 1 20809 28.1.1 REV 1 21243 28.1.1 REV 1 22358 28.1.1 REV 1 23366 28.1.1 REV 1 26337 28.1.1 REV 1 12044069 28.1.1 REV 1 12061618 28.1.1 REV 1 11637 28.1.1 REV 1 12121 28.1.1 REV 1 12786 28.1.1 REV 1 13670 28.1.1 REV 1 19366 28.1.1 REV 1 20236 28.1.1 REV 1 21165 28.1.1 REV 1 22287 28.1.1 REV 1 24110 28.1.1 REV 1 12047215 28.1.1 REV 1 12052591 28.1.1 REV 1 13742 28.1.1 REV 1 19175 28.1.1 REV 1 19600 28.1.1 REV 1 19634 28.1.1 REV 1 23139 28.1.1 REV 1 23168 28.1.1 REV 1 24086 28.1.1 REV 1 24181 28.1.1 REV 1 26365 28.1.1 REV 1 27645 28.1.1 REV 1 12045810 28.1.1 REV 1 12047217 28.1.1 REV 1 12050191 28.1.1 REV 1 11513 28.1.1 REV 1 12234 28.1.1 REV 1 16330 28.1.1 REV 1 18096 28.1.1 REV 1 19503 28.1.1 REV 1 23715 28.1.1 REV 1 24690 28.1.1 REV 1 12042468 28.1.1 REV 1 12045677 28.1.1 REV 1 12045862 28.1.1 REV 1 11700 28.1.1 REV 1 13506 28.1.1 REV 1 13653 28.1.1 REV 1 15628 28.1.1 REV 1 17930 28.1.1 REV 1 20758 28.1.1 REV 1 21218 28.1.1 REV 1 22579 28.1.1 REV 1 23738 28.1.1 REV 1 25757 28.1.1 REV 1 25908 28.1.1 REV 1 26765 28.1.1 REV 1 26988 28.1.1 REV 1 12041700 28.1.1 REV 1 12041906 28.1.1 REV 1 12042654 28.1.1 REV 1 12042788 28.1.1 REV 1 12044245 28.1.1 REV 1 12050089 28.1.1 REV 1 12052036 28.1.1 REV 1 11387 28.1.1 REV 1 12149 28.1.1 REV 1 16416 28.1.1 REV 1 17843 28.1.1 REV 1 19470 28.1.1 REV 1 20003 28.1.1 REV 1 20205 28.1.1 REV 1 24233 28.1.1 REV 1 24568 28.1.1 REV 1 24883 28.1.1 REV 1 27372 28.1.1 REV 1 12044421 28.1.1 REV 1 12044602 28.1.1 REV 1 12046643 28.1.1 REV 1 12048902 28.1.1 REV 1 11862 28.1.1 REV 1 13999 28.1.1 REV 1 16198 28.1.1 REV 1 16266 28.1.1 REV 1 17634 28.1.1 REV 1 19912 28.1.1 REV 1 21112 28.1.1 REV 1 22996 28.1.1 REV 1 23109 28.1.1 REV 1 23188 28.1.1 REV 1 23204 28.1.1 REV 1 23944 28.1.1 REV 1 25385 28.1.1 REV 1 27845 28.1.1 REV 1 12043699 28.1.1 REV 1 12044203 28.1.1 REV 1 12046468 28.1.1 REV 1 12047639 28.1.1 REV 1 12048323 28.1.1 REV 1 12052358 28.1.1 REV 1 11653 28.1.1 REV 1 11756 28.1.1 REV 1 16323 28.1.1 REV 1 17751 28.1.1 REV 1 25023 28.1.1 REV 1 12041344 28.1.1 REV 1 12041989 28.1.1 REV 1 12043925 28.1.1 REV 1 12051648 28.1.1 REV 1 12132 28.1.1 REV 1 14748 28.1.1 REV 1 14771 28.1.1 REV 1 16442 28.1.1 REV 1 19263 28.1.1 REV 1 20473 28.1.1 REV 1 21823 28.1.1 REV 1 21912 28.1.1 REV 1 22507 28.1.1 REV 1 23192 28.1.1 REV 1 25342 28.1.1 REV 1 25817 28.1.1 REV 1 26002 28.1.1 REV 1 30678 28.1.1 REV 1 31052 28.1.1 REV 1 12040399 28.1.1 REV 1 12044417 28.1.1 REV 1 12046553 28.1.1 REV 1 12052095 28.1.1 REV 1 12054725 28.1.1 REV 1 12055198 28.1.1 REV 1 12437 28.1.1 REV 1 15060 28.1.1 REV 1 15199 28.1.1 REV 1 15277 28.1.1 REV 1 16244 28.1.1 REV 1 16290 28.1.1 REV 1 17011 28.1.1 REV 1 19223 28.1.1 REV 1 19433 28.1.1 REV 1 20314 28.1.1 REV 1 20379 28.1.1 REV 1 21849 28.1.1 REV 1 25509 28.1.1 REV 1 30799 28.1.1 REV 1 12045775 28.1.1 REV 1 12049483 28.1.1 REV 1 12052116 28.1.1 REV 1 11820 28.1.1 REV 1 12605 28.1.1 REV 1 14298 28.1.1 REV 1 15826 28.1.1 REV 1 17616 28.1.1 REV 1 22708 28.1.1 REV 1 23410 28.1.1 REV 1 23764 28.1.1 REV 1 26880 28.1.1 REV 1 12040678 28.1.1 REV 1 12041743 28.1.1 REV 1 12042022 28.1.1 REV 1 12044945 28.1.1 REV 1 12045827 28.1.1 REV 1 12046237 28.1.1 REV 1 12049456 28.1.1 REV 1 12052064 28.1.1 REV 1 12052163 28.1.1 REV 1 12052550 28.1.1 REV 1 12055900 28.1.1 REV 1 12061598 28.1.1 REV 1 11785 28.1.1 REV 1 12196 28.1.1 REV 1 12683 28.1.1 REV 1 15439 28.1.1 REV 1 20312 28.1.1 REV 1 20847 28.1.1 REV 1 20932 28.1.1 REV 1 22378 28.1.1 REV 1 24186 28.1.1 REV 1 25030 28.1.1 REV 1 25879 28.1.1 REV 1 25956 28.1.1 REV 1 26339 28.1.1 REV 1 26649 28.1.1 REV 1 27215 28.1.1 REV 1 30968 28.1.1 REV 1 12041299 28.1.1 REV 1 12041805 28.1.1 REV 1 12042670 28.1.1 REV 1 12042720 28.1.1 REV 1 12044147 28.1.1 REV 1 12046106 28.1.1 REV 1 12048257 28.1.1 REV 1 14675 28.1.1 REV 1 15374 28.1.1 REV 1 16785 28.1.1 REV 1 19505 28.1.1 REV 1 20386 28.1.1 REV 1 21758 28.1.1 REV 1 24526 28.1.1 REV 1 26424 28.1.1 REV 1 12040876 28.1.1 REV 1 12041031 28.1.1 REV 1 12042122 28.1.1 REV 1 12042220 28.1.1 REV 1 12044611 28.1.1 REV 1 12045669 28.1.1 REV 1 12061546 28.1.1 REV 1 11824 28.1.1 REV 1 11913 28.1.1 REV 1 12194 28.1.1 REV 1 12211 28.1.1 REV 1 13702 28.1.1 REV 1 17959 28.1.1 REV 1 19636 28.1.1 REV 1 20335 28.1.1 REV 1 20802 28.1.1 REV 1 21316 28.1.1 REV 1 21946 28.1.1 REV 1 22201 28.1.1 REV 1 23025 28.1.1 REV 1 23088 28.1.1 REV 1 23677 28.1.1 REV 1 24971 28.1.1 REV 1 26592 28.1.1 REV 1 27419 28.1.1 REV 1 27775 28.1.1 REV 1 12040943 28.1.1 REV 1 12042634 28.1.1 REV 1 12043593 28.1.1 REV 1 12044183 28.1.1 REV 1 12045728 28.1.1 REV 1 12057 28.1.1 REV 1 17549 28.1.1 REV 1 17964 28.1.1 REV 1 25781 28.1.1 REV 1 12043807 28.1.1 REV 1 12045435 28.1.1 REV 1 12050014 28.1.1 REV 1 12065 28.1.1 REV 1 17116 28.1.1 REV 1 17286 28.1.1 REV 1 19686 28.1.1 REV 1 20824 28.1.1 REV 1 23444 28.1.1 REV 1 24756 28.1.1 REV 1 24871 28.1.1 REV 1 24873 28.1.1 REV 1 12042204 28.1.1 REV 1 12043920 28.1.1 REV 1 12043937 28.1.1 REV 1 12045700 28.1.1 REV 1 12049429 28.1.1 REV 1 12051285 28.1.1 REV 1 12052131 28.1.1 REV 1 12053213 28.1.1 REV 1 12348 28.1.1 REV 1 12475 28.1.1 REV 1 12557 28.1.1 REV 1 16414 28.1.1 REV 1 16883 28.1.1 REV 1 20813 28.1.1 REV 1 23107 28.1.1 REV 1 23436 28.1.1 REV 1 23568 28.1.1 REV 1 23807 28.1.1 REV 1 24442 28.1.1 REV 1 24841 28.1.1 REV 1 27240 28.1.1 REV 1 27315 28.1.1 REV 1 27503 28.1.1 REV 1 12040719 28.1.1 REV 1 12041378 28.1.1 REV 1 12048170 28.1.1 REV 1 12052515 28.1.1 REV 1 15381 28.1.1 REV 1 18516 28.1.1 REV 1 19804 28.1.1 REV 1 20401 28.1.1 REV 1 21116 28.1.1 REV 1 21919 28.1.1 REV 1 23556 28.1.1 REV 1 25039 28.1.1 REV 1 12045207 28.1.1 REV 1 12052399 28.1.1 REV 1 12060735 28.1.1 REV 1 15425 28.1.1 REV 1 24726 28.1.1 REV 1 25626 28.1.1 REV 1 25867 28.1.1 REV 1 12041279 28.1.1 REV 1 12045195 28.1.1 REV 1 12055463 28.1.1 REV 1 12055792 28.1.1 REV 1 11353 28.1.1 REV 1 11522 28.1.1 REV 1 15442 28.1.1 REV 1 21039 28.1.1 REV 1 22231 28.1.1 REV 1 25314 28.1.1 REV 1 25674 28.1.1 REV 1 27923 28.1.1 REV 1 12042073 28.1.1 REV 1 12043911 28.1.1 REV 1 12048178 28.1.1 REV 1 12060662 28.1.1 REV 1 12450 28.1.1 REV 1 13729 28.1.1 REV 1 19585 28.1.1 REV 1 19676 28.1.1 REV 1 23102 28.1.1 REV 1 24243 28.1.1 REV 1 25429 28.1.1 REV 1 30900 28.1.1 REV 1 31047 28.1.1 REV 1 12040838 28.1.1 REV 1 12048258 28.1.1 REV 1 12049503 28.1.1 REV 1 12060330 28.1.1 REV 1 12368 28.1.1 REV 1 13923 28.1.1 REV 1 18182 28.1.1 REV 1 24012 28.1.1 REV 1 27796 28.1.1 REV 1 12045312 28.1.1 REV 1 12045441 28.1.1 REV 1 12045898 28.1.1 REV 1 12051269 28.1.1 REV 1 12060717 28.1.1 REV 1 11605 28.1.1 REV 1 16837 28.1.1 REV 1 23254 28.1.1 REV 1 23804 28.1.1 REV 1 26397 28.1.1 REV 1 26960 28.1.1 REV 1 27264 28.1.1 REV 1 12041407 28.1.1 REV 1 12042054 28.1.1 REV 1 12042785 28.1.1 REV 1 12044561 28.1.1 REV 1 12045037 28.1.1 REV 1 12045871 28.1.1 REV 1 12046951 28.1.1 REV 1 11933 28.1.1 REV 1 14587 28.1.1 REV 1 21096 28.1.1 REV 1 23592 28.1.1 REV 1 23737 28.1.1 REV 1 27637 28.1.1 REV 1 30806 28.1.1 REV 1 12042776 28.1.1 REV 1 12048406 28.1.1 REV 1 12048415 28.1.1 REV 1 12051380 28.1.1 REV 1 24574 28.1.1 REV 1 24285 28.1.1 REV 1 24607 28.1.1 REV 1 25984 28.1.1 REV 1 12045309 28.1.1 REV 1 11743 28.1.1 REV 1 12041962 28.1.1 REV 1 18595 28.1.1 REV 1 24462 28.1.1 REV 1 12041677 28.1.1 REV 1 12048389 28.1.1 REV 1 12048367 28.1.1 REV 1 12044021 28.1.1 REV 1 12052137 28.1.1 REV 1 19403 28.1.1 REV 1 12043949 28.1.1 REV 1 20034 28.1.1 REV 1 12055303 28.1.1 REV 1 24446 28.1.1 REV 1 16067 28.1.1 REV 1 19343 28.1.1 REV 1 25642 28.1.1 REV 1 12051376 28.1.1 REV 1 12044612 28.1.1 REV 1 25451 28.1.1 REV 1 11940 28.1.1 REV 1 18367 28.1.1 REV 1 12041227 28.1.1 REV 1 12055531 28.1.1 REV 1 24761 28.1.2 REV 1 24803 28.1.2 REV 1 12044767 28.1.2 REV 1 12042423 28.1.2 REV 1 14717 28.1.2 REV 1 12044309 28.1.2 REV 1 12044328 28.1.2 REV 1 12044341 28.1.2 REV 1 12044342 28.1.2 REV 1 15883 28.1.2 REV 1 19744 28.1.2 REV 1 22543 28.1.2 REV 1 22880 28.1.2 REV 1 22915 28.1.2 REV 1 13360 28.1.2 REV 1 14168 28.1.2 REV 1 14635 28.1.2 REV 1 14647 28.1.2 REV 1 14674 28.1.2 REV 1 15498 28.1.2 REV 1 15864 28.1.2 REV 1 19795 28.1.2 REV 1 22531 28.1.2 REV 1 12042702 28.1.2 REV 1 12051962 28.1.2 REV 1 13681 28.1.2 REV 1 14927 28.1.2 REV 1 15218 28.1.2 REV 1 30507 28.1.2 REV 1 10764 28.1.2 REV 1 12044392 28.1.2 REV 1 12044867 28.1.2 REV 1 12045423 28.1.2 REV 1 12045913 28.1.2 REV 1 12046177 28.1.2 REV 1 12048591 28.1.2 REV 1 12048976 28.1.2 REV 1 12049068 28.1.2 REV 1 12051449 28.1.2 REV 1 12051467 28.1.2 REV 1 12051883 28.1.2 REV 1 12053138 28.1.2 REV 1 12053158 28.1.2 REV 1 12053160 28.1.2 REV 1 12053199 28.1.2 REV 1 12053347 28.1.2 REV 1 12053848 28.1.2 REV 1 12054009 28.1.2 REV 1 12054143 28.1.2 REV 1 12056128 28.1.2 REV 1 12056769 28.1.2 REV 1 12649 28.1.2 REV 1 12658 28.1.2 REV 1 12659 28.1.2 REV 1 12692 28.1.2 REV 1 12958 28.1.2 REV 1 12980 28.1.2 REV 1 13046 28.1.2 REV 1 13278 28.1.2 REV 1 13473 28.1.2 REV 1 13905 28.1.2 REV 1 13984 28.1.2 REV 1 14097 28.1.2 REV 1 14157 28.1.2 REV 1 14341 28.1.2 REV 1 14371 28.1.2 REV 1 14435 28.1.2 REV 1 14625 28.1.2 REV 1 14659 28.1.2 REV 1 14684 28.1.2 REV 1 14727 28.1.2 REV 1 14729 28.1.2 REV 1 14738 28.1.2 REV 1 14753 28.1.2 REV 1 14774 28.1.2 REV 1 14822 28.1.2 REV 1 14829 28.1.2 REV 1 14849 28.1.2 REV 1 14864 28.1.2 REV 1 14865 28.1.2 REV 1 14885 28.1.2 REV 1 15816 28.1.2 REV 1 15859 28.1.2 REV 1 16108 28.1.2 REV 1 16215 28.1.2 REV 1 16259 28.1.2 REV 1 16301 28.1.2 REV 1 18283 28.1.2 REV 1 18405 28.1.2 REV 1 18406 28.1.2 REV 1 18411 28.1.2 REV 1 18459 28.1.2 REV 1 19010 28.1.2 REV 1 19047 28.1.2 REV 1 19050 28.1.2 REV 1 19061 28.1.2 REV 1 19085 28.1.2 REV 1 19601 28.1.2 REV 1 19707 28.1.2 REV 1 19738 28.1.2 REV 1 20377 28.1.2 REV 1 21455 28.1.2 REV 1 21463 28.1.2 REV 1 21464 28.1.2 REV 1 21470 28.1.2 REV 1 21910 28.1.2 REV 1 22093 28.1.2 REV 1 22496 28.1.2 REV 1 22523 28.1.2 REV 1 22558 28.1.2 REV 1 22693 28.1.2 REV 1 22727 28.1.2 REV 1 22861 28.1.2 REV 1 22936 28.1.2 REV 1 22957 28.1.2 REV 1 12044620 28.1.2 REV 1 12047443 28.1.2 REV 1 13404 28.1.2 REV 1 16677 28.1.2 REV 1 23625 28.1.2 REV 1 12043772 28.1.2 REV 1 12049939 28.1.2 REV 1 14063 28.1.2 REV 1 12043726 28.1.2 REV 1 12053011 28.1.2 REV 1 12054373 28.1.2 REV 1 12051235 28.1.2 REV 1 14454 28.1.2 REV 1 17734 28.1.2 REV 1 18358 28.1.2 REV 1 25707 28.1.2 REV 1 30466 28.1.2 REV 1 12042305 28.1.2 REV 1 13000 28.1.2 REV 1 13013 28.1.2 REV 1 14303 28.1.2 REV 1 22716 28.1.2 REV 1 24734 28.1.2 REV 1 27478 28.1.2 REV 1 12045038 28.1.2 REV 1 12046723 28.1.2 REV 1 12049959 28.1.2 REV 1 18874 28.1.2 REV 1 12045951 28.1.2 REV 1 12998 28.1.2 REV 1 13253 28.1.2 REV 1 22481 28.1.2 REV 1 26688 28.1.2 REV 1 12049860 28.1.2 REV 1 12050777 28.1.2 REV 1 12053016 28.1.2 REV 1 22853 28.1.2 REV 1 26691 28.1.2 REV 1 30142 28.1.2 REV 1 12047605 28.1.2 REV 1 12049716 28.1.2 REV 1 12052915 28.1.2 REV 1 13248 28.1.2 REV 1 14240 28.1.2 REV 1 14870 28.1.2 REV 1 20532 28.1.2 REV 1 20600 28.1.2 REV 1 21353 28.1.2 REV 1 21459 28.1.2 REV 1 22533 28.1.2 REV 1 25519 28.1.2 REV 1 12046161 28.1.2 REV 1 12047428 28.1.2 REV 1 12047546 28.1.2 REV 1 12047790 28.1.2 REV 1 12049064 28.1.2 REV 1 12049827 28.1.2 REV 1 12049867 28.1.2 REV 1 12049969 28.1.2 REV 1 12050829 28.1.2 REV 1 12050854 28.1.2 REV 1 12052317 28.1.2 REV 1 12053039 28.1.2 REV 1 12053498 28.1.2 REV 1 12053593 28.1.2 REV 1 13309 28.1.2 REV 1 15278 28.1.2 REV 1 19736 28.1.2 REV 1 22436 28.1.2 REV 1 22799 28.1.2 REV 1 22823 28.1.2 REV 1 30506 28.1.2 REV 1 12043717 28.1.2 REV 1 12044863 28.1.2 REV 1 12047186 28.1.2 REV 1 12047606 28.1.2 REV 1 12048606 28.1.2 REV 1 12048617 28.1.2 REV 1 12048675 28.1.2 REV 1 12049880 28.1.2 REV 1 12052285 28.1.2 REV 1 12052464 28.1.2 REV 1 12052480 28.1.2 REV 1 12052911 28.1.2 REV 1 12053699 28.1.2 REV 1 12055967 28.1.2 REV 1 12056201 28.1.2 REV 1 12056450 28.1.2 REV 1 12062539 28.1.2 REV 1 12272 28.1.2 REV 1 12883 28.1.2 REV 1 12989 28.1.2 REV 1 13572 28.1.2 REV 1 14037 28.1.2 REV 1 14098 28.1.2 REV 1 14174 28.1.2 REV 1 14552 28.1.2 REV 1 14860 28.1.2 REV 1 14931 28.1.2 REV 1 14998 28.1.2 REV 1 15391 28.1.2 REV 1 15782 28.1.2 REV 1 17351 28.1.2 REV 1 17792 28.1.2 REV 1 19663 28.1.2 REV 1 22425 28.1.2 REV 1 22890 28.1.2 REV 1 23529 28.1.2 REV 1 24266 28.1.2 REV 1 24454 28.1.2 REV 1 28133 28.1.2 REV 1 12041092 28.1.2 REV 1 12041136 28.1.2 REV 1 12042283 28.1.2 REV 1 12042295 28.1.2 REV 1 12042440 28.1.2 REV 1 12042491 28.1.2 REV 1 12042548 28.1.2 REV 1 12042845 28.1.2 REV 1 12042891 28.1.2 REV 1 12043170 28.1.2 REV 1 12043299 28.1.2 REV 1 12044211 28.1.2 REV 1 12044222 28.1.2 REV 1 12044289 28.1.2 REV 1 12044357 28.1.2 REV 1 12044798 28.1.2 REV 1 12044858 28.1.2 REV 1 12044882 28.1.2 REV 1 12045948 28.1.2 REV 1 12046030 28.1.2 REV 1 12046183 28.1.2 REV 1 12046441 28.1.2 REV 1 12046528 28.1.2 REV 1 12046788 28.1.2 REV 1 12046838 28.1.2 REV 1 12046859 28.1.2 REV 1 12047125 28.1.2 REV 1 12047147 28.1.2 REV 1 12047259 28.1.2 REV 1 12047272 28.1.2 REV 1 12047289 28.1.2 REV 1 12047349 28.1.2 REV 1 12047372 28.1.2 REV 1 12047560 28.1.2 REV 1 12047567 28.1.2 REV 1 12047593 28.1.2 REV 1 12047597 28.1.2 REV 1 12047705 28.1.2 REV 1 12047710 28.1.2 REV 1 12047796 28.1.2 REV 1 12047808 28.1.2 REV 1 12048577 28.1.2 REV 1 12048667 28.1.2 REV 1 12048938 28.1.2 REV 1 12048982 28.1.2 REV 1 12049008 28.1.2 REV 1 12049010 28.1.2 REV 1 12049036 28.1.2 REV 1 12049044 28.1.2 REV 1 12049047 28.1.2 REV 1 12049051 28.1.2 REV 1 12049076 28.1.2 REV 1 12049658 28.1.2 REV 1 12049703 28.1.2 REV 1 12049748 28.1.2 REV 1 12049803 28.1.2 REV 1 12049872 28.1.2 REV 1 12049927 28.1.2 REV 1 12049967 28.1.2 REV 1 12050273 28.1.2 REV 1 12050278 28.1.2 REV 1 12050298 28.1.2 REV 1 12050305 28.1.2 REV 1 12050311 28.1.2 REV 1 12050359 28.1.2 REV 1 12050606 28.1.2 REV 1 12050738 28.1.2 REV 1 12050770 28.1.2 REV 1 12050799 28.1.2 REV 1 12050800 28.1.2 REV 1 12050831 28.1.2 REV 1 12050843 28.1.2 REV 1 12050867 28.1.2 REV 1 12051477 28.1.2 REV 1 12051479 28.1.2 REV 1 12051485 28.1.2 REV 1 12051512 28.1.2 REV 1 12051515 28.1.2 REV 1 12051533 28.1.2 REV 1 12052306 28.1.2 REV 1 12052404 28.1.2 REV 1 12052455 28.1.2 REV 1 12052894 28.1.2 REV 1 12052905 28.1.2 REV 1 12052917 28.1.2 REV 1 12053071 28.1.2 REV 1 12053183 28.1.2 REV 1 12053210 28.1.2 REV 1 12053314 28.1.2 REV 1 12053667 28.1.2 REV 1 12053674 28.1.2 REV 1 12054375 28.1.2 REV 1 12054379 28.1.2 REV 1 12054414 28.1.2 REV 1 12054415 28.1.2 REV 1 12054432 28.1.2 REV 1 12054464 28.1.2 REV 1 12054490 28.1.2 REV 1 12054985 28.1.2 REV 1 12055607 28.1.2 REV 1 12055626 28.1.2 REV 1 12055787 28.1.2 REV 1 12055818 28.1.2 REV 1 12055847 28.1.2 REV 1 12056066 28.1.2 REV 1 12056247 28.1.2 REV 1 12056372 28.1.2 REV 1 12056446 28.1.2 REV 1 12056457 28.1.2 REV 1 12056482 28.1.2 REV 1 12056551 28.1.2 REV 1 12056577 28.1.2 REV 1 12056578 28.1.2 REV 1 12056656 28.1.2 REV 1 12056690 28.1.2 REV 1 12056750 28.1.2 REV 1 12062166 28.1.2 REV 1 12590 28.1.2 REV 1 12769 28.1.2 REV 1 12792 28.1.2 REV 1 12823 28.1.2 REV 1 12832 28.1.2 REV 1 12882 28.1.2 REV 1 12895 28.1.2 REV 1 12915 28.1.2 REV 1 13011 28.1.2 REV 1 13037 28.1.2 REV 1 13038 28.1.2 REV 1 13262 28.1.2 REV 1 13301 28.1.2 REV 1 13303 28.1.2 REV 1 13306 28.1.2 REV 1 13312 28.1.2 REV 1 13324 28.1.2 REV 1 13325 28.1.2 REV 1 13334 28.1.2 REV 1 13338 28.1.2 REV 1 13350 28.1.2 REV 1 13388 28.1.2 REV 1 13432 28.1.2 REV 1 13436 28.1.2 REV 1 13585 28.1.2 REV 1 13605 28.1.2 REV 1 13634 28.1.2 REV 1 13641 28.1.2 REV 1 13692 28.1.2 REV 1 14162 28.1.2 REV 1 14178 28.1.2 REV 1 14186 28.1.2 REV 1 14212 28.1.2 REV 1 14217 28.1.2 REV 1 14237 28.1.2 REV 1 14268 28.1.2 REV 1 14378 28.1.2 REV 1 15027 28.1.2 REV 1 15638 28.1.2 REV 1 15686 28.1.2 REV 1 15691 28.1.2 REV 1 15697 28.1.2 REV 1 15861 28.1.2 REV 1 15890 28.1.2 REV 1 15907 28.1.2 REV 1 18395 28.1.2 REV 1 18570 28.1.2 REV 1 18593 28.1.2 REV 1 18597 28.1.2 REV 1 18598 28.1.2 REV 1 18610 28.1.2 REV 1 18611 28.1.2 REV 1 19023 28.1.2 REV 1 19055 28.1.2 REV 1 19734 28.1.2 REV 1 19735 28.1.2 REV 1 19746 28.1.2 REV 1 19791 28.1.2 REV 1 20455 28.1.2 REV 1 20494 28.1.2 REV 1 20495 28.1.2 REV 1 20512 28.1.2 REV 1 20570 28.1.2 REV 1 20576 28.1.2 REV 1 20587 28.1.2 REV 1 20588 28.1.2 REV 1 20632 28.1.2 REV 1 20672 28.1.2 REV 1 20675 28.1.2 REV 1 21378 28.1.2 REV 1 21475 28.1.2 REV 1 21476 28.1.2 REV 1 22367 28.1.2 REV 1 22376 28.1.2 REV 1 22403 28.1.2 REV 1 22420 28.1.2 REV 1 22427 28.1.2 REV 1 22429 28.1.2 REV 1 22445 28.1.2 REV 1 22449 28.1.2 REV 1 22480 28.1.2 REV 1 22524 28.1.2 REV 1 22537 28.1.2 REV 1 22556 28.1.2 REV 1 22698 28.1.2 REV 1 22699 28.1.2 REV 1 22722 28.1.2 REV 1 22749 28.1.2 REV 1 22790 28.1.2 REV 1 22807 28.1.2 REV 1 22808 28.1.2 REV 1 22813 28.1.2 REV 1 22843 28.1.2 REV 1 22860 28.1.2 REV 1 22894 28.1.2 REV 1 22895 28.1.2 REV 1 22900 28.1.2 REV 1 22927 28.1.2 REV 1 22962 28.1.2 REV 1 22965 28.1.2 REV 1 22966 28.1.2 REV 1 22977 28.1.2 REV 1 22979 28.1.2 REV 1 23005 28.1.2 REV 1 23044 28.1.2 REV 1 23046 28.1.2 REV 1 23059 28.1.2 REV 1 23958 28.1.2 REV 1 24783 28.1.2 REV 1 24995 28.1.2 REV 1 25530 28.1.2 REV 1 25591 28.1.2 REV 1 25612 28.1.2 REV 1 25618 28.1.2 REV 1 26113 28.1.2 REV 1 26142 28.1.2 REV 1 26868 28.1.2 REV 1 27075 28.1.2 REV 1 27095 28.1.2 REV 1 27113 28.1.2 REV 1 27116 28.1.2 REV 1 27127 28.1.2 REV 1 27134 28.1.2 REV 1 27150 28.1.2 REV 1 27182 28.1.2 REV 1 27193 28.1.2 REV 1 27618 28.1.2 REV 1 30051 28.1.2 REV 1 30077 28.1.2 REV 1 30086 28.1.2 REV 1 30089 28.1.2 REV 1 30095 28.1.2 REV 1 30097 28.1.2 REV 1 30102 28.1.2 REV 1 30143 28.1.2 REV 1 30205 28.1.2 REV 1 30217 28.1.2 REV 1 30244 28.1.2 REV 1 30249 28.1.2 REV 1 30251 28.1.2 REV 1 30324 28.1.2 REV 1 30359 28.1.2 REV 1 30441 28.1.2 REV 1 30465 28.1.2 REV 1 30488 28.1.2 REV 1 12063954 28.1.2 REV 1 12064191 28.1.2 REV 1 12064518 28.1.2 REV 1 12042321 28.1.2 REV 1 12043633 28.1.2 REV 1 12046735 28.1.2 REV 1 12046883 28.1.2 REV 1 12047373 28.1.2 REV 1 12048494 28.1.2 REV 1 12049053 28.1.2 REV 1 12049717 28.1.2 REV 1 12051932 28.1.2 REV 1 12052483 28.1.2 REV 1 12055744 28.1.2 REV 1 12056040 28.1.2 REV 1 12056245 28.1.2 REV 1 12056647 28.1.2 REV 1 12120 28.1.2 REV 1 13029 28.1.2 REV 1 13040 28.1.2 REV 1 14129 28.1.2 REV 1 14175 28.1.2 REV 1 14881 28.1.2 REV 1 15543 28.1.2 REV 1 15696 28.1.2 REV 1 15760 28.1.2 REV 1 16768 28.1.2 REV 1 17033 28.1.2 REV 1 17537 28.1.2 REV 1 17546 28.1.2 REV 1 17577 28.1.2 REV 1 17583 28.1.2 REV 1 17764 28.1.2 REV 1 18097 28.1.2 REV 1 18181 28.1.2 REV 1 18304 28.1.2 REV 1 18964 28.1.2 REV 1 19107 28.1.2 REV 1 22070 28.1.2 REV 1 22714 28.1.2 REV 1 22759 28.1.2 REV 1 22960 28.1.2 REV 1 23395 28.1.2 REV 1 24548 28.1.2 REV 1 25124 28.1.2 REV 1 25372 28.1.2 REV 1 25634 28.1.2 REV 1 26061 28.1.2 REV 1 12044878 28.1.2 REV 1 12044927 28.1.2 REV 1 12822 28.1.2 REV 1 13255 28.1.2 REV 1 14696 28.1.2 REV 1 26951 28.1.2 REV 1 12043027 28.1.2 REV 1 12045350 28.1.2 REV 1 12049091 28.1.2 REV 1 12050269 28.1.2 REV 1 12051420 28.1.2 REV 1 12053020 28.1.2 REV 1 12054239 28.1.2 REV 1 17081 28.1.2 REV 1 17145 28.1.2 REV 1 17996 28.1.2 REV 1 20

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0753-2025

Related Recalls

There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).

Oct 6, 2025 Implants & Prosthetics View Details →

There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).

Oct 6, 2025 Implants & Prosthetics View Details →

Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.

Feb 15, 2025 Implants & Prosthetics Nationwide View Details →