SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Models UDI/GTIN TFL-FBX150S 00821925043916 TFL-FBX200S 00821925043930 TFL-FBX365S 00821925043992 TFL-FBX550S 00821925044036 TFL-FBX940S 00821925044074 TFL-FBX150BS 00821925043879 TFL-FBX200BS 00821925043978 All lots affected
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Olympus Corporation of the Americas
Reason for Recall:
Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS

Product Codes/Lot Numbers:

Models UDI/GTIN TFL-FBX150S 00821925043916 TFL-FBX200S 00821925043930 TFL-FBX365S 00821925043992 TFL-FBX550S 00821925044036 TFL-FBX940S 00821925044074 TFL-FBX150BS 00821925043879 TFL-FBX200BS 00821925043978 All lots affected

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0748-2024

Related Recalls

Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.

Oct 14, 2025 Diagnostic Equipment Nationwide View Details →

Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit.

Oct 17, 2025 Surgical Instruments View Details →