Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1506, CRB-1508, CRB-1516, CRB-1517, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1527, CRB-1528, CRB-1529, CRB-1530, CRB-1532,
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Noah Medical
- Reason for Recall:
- Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
Product Codes/Lot Numbers:
UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1506, CRB-1508, CRB-1516, CRB-1517, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1527, CRB-1528, CRB-1529, CRB-1530, CRB-1532,
Distribution:
Distributed in: US, SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0742-2025
Related Recalls
Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.
Galaxy System (GAL-001)
Noah Medical
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.