7% BSA (Bovine Serum Albumin), Product Code: 8262487 - Product Usage: is used to dilute samples when assay values exceed the reportable (dynamic) range using the VITROS 250/350/5,1 FS/4600 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot with Expiry Date: D6289 2019-11-20 E6382 2020-01-15 F6602 2020-05-02 G6679 2020-08-13 H6897 2020-10-01 J7129 2021-01-23 K7421 2021-07-01 L7592 2021-08-26
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Ortho Clinical Diagnostics Inc
Reason for Recall:
Potential for the instability of the diluents VITROS 7% BSA (Product Code 8262487) and VITROS Urine Electrolyte Diluent (UED) (Product Code 1110352) when stored in a 2mL cup for 7 days
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

7% BSA (Bovine Serum Albumin), Product Code: 8262487 - Product Usage: is used to dilute samples when assay values exceed the reportable (dynamic) range using the VITROS 250/350/5,1 FS/4600 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Product Codes/Lot Numbers:

Lot with Expiry Date: D6289 2019-11-20 E6382 2020-01-15 F6602 2020-05-02 G6679 2020-08-13 H6897 2020-10-01 J7129 2021-01-23 K7421 2021-07-01 L7592 2021-08-26

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0739-2020

Related Recalls

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

Jul 29, 2025 Diagnostic Equipment Nationwide View Details →

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

Jul 29, 2025 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive "spikes" (imperfections) in slide images and generates the TH4-63J condition code when the threshold of 600 spikes is exceeded. Updated analysis revealed a 19-fold increase in failure rates from 0.0043% (2024 baseline) to 0.082% for Generation 67+ slides in 2025, with 100% of calibrator failures occurring at the lowest concentration level. This can result in delayed diagnosis and treatment of conditions such as hypercalcemic crisis, parathyroid disorders, and other calcium-related emergencies requiring urgent intervention.