iTotal Hip Replacement System, Total Hip Replacement System, Catalog Number HAA-050-D001-020101

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Number 0424647
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Conformis, Inc.
Reason for Recall:
The stage 1 and 2 reamer instruments used to prepare fixation for the cup in the acetabulum had a -1mm interference fit with acetabulum bone. The expected interference fit is -2mm.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

iTotal Hip Replacement System, Total Hip Replacement System, Catalog Number HAA-050-D001-020101

Product Codes/Lot Numbers:

Serial Number 0424647

Distribution:

Distributed in: US, TN

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0731-2019