Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray: Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039. Allura Xper F010 {C+F). 722003 Allura Xper F010/10 722005 Allura Xper F020 722006 Allura Xper F020/10 and F020/20 722008 Allura Xper F010 722010 Altura Xper F010/10 722011 Allura Xper F020 722012 Allura Xper FD20 biplane 722013 Allura Xper FDlO OR Table 722014 Allura Xper FD20 OR Table 722015 Allura Xper FDl0/10 OR Table 722019 Allura Xper FD20 Biplane OR Table 722020 Allura Xper FD10 OR Table 722022 Allura Xper FD20 OR Table 722023 Allura Xper FD10/10 OR Table 722024 Altura Xper FD20 Biplane OR Table 722025 Allura Xper FD10 722026 Allura Xper FD10/10 722027 Altura Xper FD20/10 biplane 722029 Allura Xper FD10 OR Table 722033 Allura Xper FD10/10 OR Table 722034 Allura Xper FD20/10 OR Table 722036 Allure Xper F020/20 722038 Altura Xper FD20/20 biplane OR Table 722039
Class I - DangerousWhat Should You Do?
- Check if you have this product: Allura Xper FD10 722003/722026 (US) S/N: 24 40 71 162 191 266 283 294 319 347 406 485 825 959 965 967 985 990 1246 1353 1444 1944 1949 UDI: (01)00884838054189(21)825 (01)00884838054189(21)967 (01)00884838054189(21)1246 (01)00884838054189(21)965 (01)00884838054189(21)959 (01)00884838054189(21)990 (01)00884838054189(21)985 Allura Xper FD10/10 722005 S/N: 144 US OUS: 37 52 74 79 80 108 279 417 459 Allura Xper FD20 722006 S/N: US 44 72 234 246 263 268 327 333 429 469 471 498 514 535 550 554 588 604 718 754 766 861 1174 1274 1355 1400 1783 1831 1844 1878 1886 1931 2018 UDI: (01)00884838054202(21)893 OUS: 11 40 89 102 112 125 151 244 364 408 431 484 568 580 594 597 602 640 643 704 811 940 971 1262 1312 1402 1405 1424 1465 1481 1658 1691 1794 1796 1815 1852 1859 1879 1891 1948 1959 1974 1985 1988 2141 2311 2521 2554 Allura Xper FD20 Biplane 722008 S/N: 33, 82, 374 (US) OUS: 39 86 224 249 336 384 388 532 Allura Xper FD10 722010 S/N:OUS 163 340 1422 1429 1434 Allura Xper FD20 722012 S/N: OUS 395 463 927 1013 1518 2652 2854 2918 Allura Xper FD20 OR Table 722015 S/N: 30 (OUS) Allura Xper FD10 722026 S/N OUS: 315 701 748 884 888 943 948 960 971 975 978 984 989 995 Allura Xper FD10/10 722027 S/N: OUS 268 278 Allura Xper FD20/10 722029 S/N:OUS 9 Allura Xper FD10 722026 S/N : US 825 959 965 967 985 990 1246 OUS: 11 40 89 102 112 125 151 244 364 408 431 484 568 580 594 597 602 640 643 704 811 940 971 1262 1312 1402 1405 1424 1465 1481 1658 1691 1794 1796 1815 1852 1859 1879 1891 1948 1959 1974 1985 1988 2141 2311 2521 2554 Allura Xper FD10/10 266 722027 S/N: 266 Allura Xper FD20 722028 S/N: 893 UDI: (01)00884838054202(21)893 Allura Xper FD20/10 722029 S/N: 91, 94 US UDI: (01)00884838054219(21)94 OUS: S/N:6
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips North America Llc
- Reason for Recall:
- Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray: Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039. Allura Xper F010 {C+F). 722003 Allura Xper F010/10 722005 Allura Xper F020 722006 Allura Xper F020/10 and F020/20 722008 Allura Xper F010 722010 Altura Xper F010/10 722011 Allura Xper F020 722012 Allura Xper FD20 biplane 722013 Allura Xper FDlO OR Table 722014 Allura Xper FD20 OR Table 722015 Allura Xper FDl0/10 OR Table 722019 Allura Xper FD20 Biplane OR Table 722020 Allura Xper FD10 OR Table 722022 Allura Xper FD20 OR Table 722023 Allura Xper FD10/10 OR Table 722024 Altura Xper FD20 Biplane OR Table 722025 Allura Xper FD10 722026 Allura Xper FD10/10 722027 Altura Xper FD20/10 biplane 722029 Allura Xper FD10 OR Table 722033 Allura Xper FD10/10 OR Table 722034 Allura Xper FD20/10 OR Table 722036 Allure Xper F020/20 722038 Altura Xper FD20/20 biplane OR Table 722039
Product Codes/Lot Numbers:
Allura Xper FD10 722003/722026 (US) S/N: 24 40 71 162 191 266 283 294 319 347 406 485 825 959 965 967 985 990 1246 1353 1444 1944 1949 UDI: (01)00884838054189(21)825 (01)00884838054189(21)967 (01)00884838054189(21)1246 (01)00884838054189(21)965 (01)00884838054189(21)959 (01)00884838054189(21)990 (01)00884838054189(21)985 Allura Xper FD10/10 722005 S/N: 144 US OUS: 37 52 74 79 80 108 279 417 459 Allura Xper FD20 722006 S/N: US 44 72 234 246 263 268 327 333 429 469 471 498 514 535 550 554 588 604 718 754 766 861 1174 1274 1355 1400 1783 1831 1844 1878 1886 1931 2018 UDI: (01)00884838054202(21)893 OUS: 11 40 89 102 112 125 151 244 364 408 431 484 568 580 594 597 602 640 643 704 811 940 971 1262 1312 1402 1405 1424 1465 1481 1658 1691 1794 1796 1815 1852 1859 1879 1891 1948 1959 1974 1985 1988 2141 2311 2521 2554 Allura Xper FD20 Biplane 722008 S/N: 33, 82, 374 (US) OUS: 39 86 224 249 336 384 388 532 Allura Xper FD10 722010 S/N:OUS 163 340 1422 1429 1434 Allura Xper FD20 722012 S/N: OUS 395 463 927 1013 1518 2652 2854 2918 Allura Xper FD20 OR Table 722015 S/N: 30 (OUS) Allura Xper FD10 722026 S/N OUS: 315 701 748 884 888 943 948 960 971 975 978 984 989 995 Allura Xper FD10/10 722027 S/N: OUS 268 278 Allura Xper FD20/10 722029 S/N:OUS 9 Allura Xper FD10 722026 S/N : US 825 959 965 967 985 990 1246 OUS: 11 40 89 102 112 125 151 244 364 408 431 484 568 580 594 597 602 640 643 704 811 940 971 1262 1312 1402 1405 1424 1465 1481 1658 1691 1794 1796 1815 1852 1859 1879 1891 1948 1959 1974 1985 1988 2141 2311 2521 2554 Allura Xper FD10/10 266 722027 S/N: 266 Allura Xper FD20 722028 S/N: 893 UDI: (01)00884838054202(21)893 Allura Xper FD20/10 722029 S/N: 91, 94 US UDI: (01)00884838054219(21)94 OUS: S/N:6
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0724-2022
Related Recalls
Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Philips North America
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
IntelliVue MP90. Product Number: M8010A.
Philips North America
Potential issue where the IntelliVue monitors did not alarm.
IntelliVue MP30. Product Number: M8002A.
Philips North America
Potential issue where the IntelliVue monitors did not alarm.