Arrow Berman Angiographic Catheter 60cm Item Number: AI-07130-J

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Number: 16F19H0093
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Arrow International Inc
Reason for Recall:
Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Arrow Berman Angiographic Catheter 60cm Item Number: AI-07130-J

Product Codes/Lot Numbers:

Lot Number: 16F19H0093

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0716-2020

Related Recalls

Arrow Arterial Kits and Trays as follows: (1) Arrow Arterial Access Tray with 0.025" Spring-Wire Guide (Catheter Not Included), REF ASK-04001-BW1; (2) Arrow Arterial Access Tray with 0.021" Spring-Wire Guide (Catheter Not Included), REF ASK-04001-CK; (3) Arrow Arterial Access Kit, .025 inch dia. spring-wire guide, REF ASK-04001-VAMC1; (4) Arrow Arterial Catheterization Kit, 20 Ga., 4.45 cm catheter length, .018 inch dia. spring-wire guide, REF ASK-04020-BHS; (5) Arrow Arterial Catheterization Kit, 20 Ga., 4.45 cm catheter length, .025 inch dia. spring-wire guide, REF ASK-04020-BID; (6) Arrow Arterial Line Kit, 20 Ga., 12 cm. catheter length, .025 inch dia. spring-wire guide, REF ASK-04020-HMC; (7) Arrow Arterial Catheterization Kit, 20 Ga. 4.45 cm catheter length, .018 inch dia. spring-wire guide, REF ASK-04020-HOA; (8) Arrow Arterial Catheterization Kit, 20 Ga., 4.45 cm. catheter length, .018 inch dia. spring-wire guide, REF ASK-04020-KSD; (9) Arrow Arterial Catheterization Kit, 20 Ga. 4.45 cm catheter length, .018 inch dia. spring-wire guide, REF ASK-04020-LOL; (10) Arrow Radial Catheterization, 20 Ga. 4.45 cm catheter length, .018 inch dia. spring-wire guide, REF ASK-04020-MHS; (11) Arrow Arterial Catheterization Kit, 20 Ga., 4.45 cm catheter length, 018 inch dia. spring-wire guide, REF ASK-04020-MIB3; (12) Arrow Arterial Catheterization Kit, 20 Ga., 4.45 cm catheter length, .018 dia. spring- wire guide, REF ASK-04020-PRH; (13) Arrow Arterial Catheterization Kit, 20 Ga., 4.45 cm catheter length, .018 dia. spring-wire guide, REF ASK-04020-UOIL; (14) Arrow QuickFlash Radial Artery Catheterization Kit, 20 Ga., 3.81 cm catheter length, .018 inch dia. spring-wire guide, REF ASK-04220-KSP; and (15) Arrow Arterial Catheterization Kit, 20 Ga. 4.45/12 cm catheter length, .018 inch dia. spring-wire guide, REF ASK-04500-HFH4.

ARROW INTERNATIONAL

Class I - Dangerous

Reports have been received regarding open seals on the packaging.

Jul 26, 2024 Surgical Instruments Nationwide View Details →

Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.

Apr 29, 2024 Surgical Instruments Nationwide View Details →

Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.

Apr 29, 2024 Surgical Instruments Nationwide View Details →