KWIK-STIK, Quality control kit for culture media, sold as: KWIK-STIK (single-pack) Enterococcus faecalis derived from ATCC 51299
Class I - DangerousWhat Should You Do?
- Check if you have this product: Catalog Number HL163263 OEM Private Label with Healthlink; UDI: N/A (OEM agreement with Healthlink ended in Q1 2016 before UDI was required for class 1 IVDs) Lot Number (Expiration Date): 959-74-1 (10/08/2014), 959-75-2 (04/22/2015), 959-75-3 (04/22/2015), 959-75-6 (04/22/2015), 959-75-7 (04/22/2015), 959-75-10 (04/22/2015), and 959-76-2 (07/08/2015)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Microbiologics Inc
- Reason for Recall:
- This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
KWIK-STIK, Quality control kit for culture media, sold as: KWIK-STIK (single-pack) Enterococcus faecalis derived from ATCC 51299
Product Codes/Lot Numbers:
Catalog Number HL163263 OEM Private Label with Healthlink; UDI: N/A (OEM agreement with Healthlink ended in Q1 2016 before UDI was required for class 1 IVDs) Lot Number (Expiration Date): 959-74-1 (10/08/2014), 959-75-2 (04/22/2015), 959-75-3 (04/22/2015), 959-75-6 (04/22/2015), 959-75-7 (04/22/2015), 959-75-10 (04/22/2015), and 959-76-2 (07/08/2015)
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0714-2022
Related Recalls
KWIK-STIK, Quality control kit for culture media, Vibrio parahaemolyticus derived from ATCC 17802, Catalog Number 0818K
Microbiologics
Some products packaged for lot 818-111-7 may have been packaged in pouches labeled as lot 857-53-10.
LYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF 0894P (2-pack)
Microbiologics
KWIK-STIK REF numbers: 0894P, 0894K and LYFO DISK" 0894L may result in failure to recover the target microorganism. All lots of REF 894-76 are impacted.
KWIK-STIK REF numbers: 0894P, 0894K and LYFO DISK" 0894L may result in failure to recover the target microorganism. All lots of REF 894-76 are impacted.