Recall Details

Company:
Draeger Medical Systems, Inc.
Reason for Recall:
After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Draeger Infinity Acute Care System Monitoring Solution

Product Codes/Lot Numbers:

with software versions VG2.0.3 and higher.

Distribution:

Distributed in: US, MA, NH, IN, CA, TX, NY, OK, IL