3mL Syringe w/Spinning Spiros, Red Cap, REF: CH2003; 5mL Syringe w/Spinning Spiros, Red Cap, REF: CH2005; 10mL Syringe w/Spinning Spiros, Red Cap, REF: CH2010; 20mL Syringe w/Spinning Spiros, Red Cap, REF: CH2020; 30mL Syringe w/Spinning Spiros, Red Cap, REF: CH2030; 60mL Syringe w/Spinning Spiros, Red Cap, REF: CH2060; 30" (76 cm) Appx 3.3 mL, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CH3011; 31" (79 cm) Appx 3.3 ml, Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros w/Red Cap, Hanger, REF: CH3011-C; 40 IN (102 cm) App 4.8 ml, Admin Set, 20 Drop, 2 Spiros, Bag Hanger, REF: CH3020; 14 IN (36 cm) App 4.8 ml, Bifuse Add-On Set, Spiros w/Red Cap, Bag Spike, REF: CH3033; 5" (13 cm) Bag Spike Adapter w/Spiros w/Red Cap, Vented Cap, REF: CH3034; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, 0.2 Micron Filter, Spiros w/Red Cap, Hanger, REF: CH3128; Oncology Kit w/60" (152 cm) Appx 2.3 ml, Smallbore Ext Set w/Spiros, Anti-Siphon Valve, ChemoClave, Clamp, Rotating Luer, Double ChemoClave Syringe Transfer Set, Spiros w/Red Cap, REF: CH3606; 5" (13 cm) Appx 0.50 ml, Smallbore Trifuse Ext Set w/3 ChemoClave (Purple, Green Rings), Spiros w/Red Cap, 3 Clamps, REF: CH3625; 86" (218 cm) Appx 10.4 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, REF: CH3642; Oncology Kit w/40" (102 cm) Appx 4.7 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, Bag Hanger, Double ChemoClave Syringe Transfer Set, REF: CH3677; 5.5" (14 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 ChemoClave, Spiros w/Red Cap, 3 Clamps, REF: CH3700; 50 IN (127 cm) Appx 6.4 ml, Ext Set w/ChemoClave, Spiros w/Red Cap, Clamp, REF: CH3703; 16 IN (41 cm) Appx 1.6 ml, Ext Set, ChemoClave, Spiros w/Red Cap, REF: CH3715; Oncology Kit, Ext w/Spiros, Grad Conn, 2 Vial Spikes w/Clave, Spinning Spiros , REF: CH3929; Oncology Kit, w/Ext, MicroClave, Clave, Grad Conn, Transfer, Spiros, Vial Spike, Clave, REF: CH3932; Oncology Kit, 12 IN Ext, Spiros, Graduated Adapter, 2 ChemoClave w/Vial Spike, Spiros, REF: CH3943; Oncology Kit w/12" Ext Set w/Spiros w/Red Cap, Clamp, Graduated Adapter, ChemoClave Vented Vial Spike, 20mm, ChemoClave Vented Vial Spike, 13mm, REF: CH3944; Oncology Kit, 16 IN (41 cm) Ext Set, ChemoClave, Graduated Conn, 2 ChemoClave w/Vial Spike, Spiros, REF: CH3946; Oncology Kit, 21" (53 cm) Appx 3.2 ml, Ext Set w/Spiros w/Red Cap, Graduated Adapter, Locking Universal Vented Vial Spike w/ChemoClave, Closed Vial Spike, 13mm w/ChemoClave, Spinning Spiros w/Red Cap, REF: CH3949; 46" (117 cm) 50 ml, Diluent Set for Channel 2 w/20 Drop Drip Chamber, Spiros, REF: CH4002
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lots: 4553702, 4605473, 4763191, 4775520, 4849205, 4549847, 4548779, 4572314, 4734247, 4763917, 4849208, 4741169, 4763921, 4560231, 4750962, 4549850, 4763200, 4734249, 4794307, 4806855, 4849220, 4861176, 4877472, 4605484, 4749780, 4795448, 4876942, 4719395, 4737146, 4764825, 4800210, 4866411, 4800214, 4734262, 4794308, 4795454, 4806862, 4849239, 4849248, 4849251, 4605504, 4719410, 4734273, 4757665, 4763982, 4799740, 4861174, 4605074, 4741282, 4751117, 4809117, 4877145, 4894538, 4557534, 4733509, 4741291, 4743763, 4774120, 4733512, 4865834, 4894550, 4573105, 4803126, 4808046, 4860027, 4806400, 4720476, 4724725, 4749443, 4786358, 4802674, 4849131, 4783296, 4730109, 4742662, 4759699, 4768019, 4794497, 4877852, 4897107, 4712378, 4749445, 4799839, 4806855, 4861176, 4719395, 4800210, 4764826, 4800214, 4719927, 4726732, 4733428, 4734262, 4741183, 4756802, 4763958, 4794308, 4794310, 4795452, 4806862, 4849230, 4849233, 4849239, 4719410, 4799839, 4859282
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- ICU Medical, Inc.
- Reason for Recall:
- Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
3mL Syringe w/Spinning Spiros, Red Cap, REF: CH2003; 5mL Syringe w/Spinning Spiros, Red Cap, REF: CH2005; 10mL Syringe w/Spinning Spiros, Red Cap, REF: CH2010; 20mL Syringe w/Spinning Spiros, Red Cap, REF: CH2020; 30mL Syringe w/Spinning Spiros, Red Cap, REF: CH2030; 60mL Syringe w/Spinning Spiros, Red Cap, REF: CH2060; 30" (76 cm) Appx 3.3 mL, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CH3011; 31" (79 cm) Appx 3.3 ml, Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros w/Red Cap, Hanger, REF: CH3011-C; 40 IN (102 cm) App 4.8 ml, Admin Set, 20 Drop, 2 Spiros, Bag Hanger, REF: CH3020; 14 IN (36 cm) App 4.8 ml, Bifuse Add-On Set, Spiros w/Red Cap, Bag Spike, REF: CH3033; 5" (13 cm) Bag Spike Adapter w/Spiros w/Red Cap, Vented Cap, REF: CH3034; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, 0.2 Micron Filter, Spiros w/Red Cap, Hanger, REF: CH3128; Oncology Kit w/60" (152 cm) Appx 2.3 ml, Smallbore Ext Set w/Spiros, Anti-Siphon Valve, ChemoClave, Clamp, Rotating Luer, Double ChemoClave Syringe Transfer Set, Spiros w/Red Cap, REF: CH3606; 5" (13 cm) Appx 0.50 ml, Smallbore Trifuse Ext Set w/3 ChemoClave (Purple, Green Rings), Spiros w/Red Cap, 3 Clamps, REF: CH3625; 86" (218 cm) Appx 10.4 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, REF: CH3642; Oncology Kit w/40" (102 cm) Appx 4.7 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, Bag Hanger, Double ChemoClave Syringe Transfer Set, REF: CH3677; 5.5" (14 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 ChemoClave, Spiros w/Red Cap, 3 Clamps, REF: CH3700; 50 IN (127 cm) Appx 6.4 ml, Ext Set w/ChemoClave, Spiros w/Red Cap, Clamp, REF: CH3703; 16 IN (41 cm) Appx 1.6 ml, Ext Set, ChemoClave, Spiros w/Red Cap, REF: CH3715; Oncology Kit, Ext w/Spiros, Grad Conn, 2 Vial Spikes w/Clave, Spinning Spiros , REF: CH3929; Oncology Kit, w/Ext, MicroClave, Clave, Grad Conn, Transfer, Spiros, Vial Spike, Clave, REF: CH3932; Oncology Kit, 12 IN Ext, Spiros, Graduated Adapter, 2 ChemoClave w/Vial Spike, Spiros, REF: CH3943; Oncology Kit w/12" Ext Set w/Spiros w/Red Cap, Clamp, Graduated Adapter, ChemoClave Vented Vial Spike, 20mm, ChemoClave Vented Vial Spike, 13mm, REF: CH3944; Oncology Kit, 16 IN (41 cm) Ext Set, ChemoClave, Graduated Conn, 2 ChemoClave w/Vial Spike, Spiros, REF: CH3946; Oncology Kit, 21" (53 cm) Appx 3.2 ml, Ext Set w/Spiros w/Red Cap, Graduated Adapter, Locking Universal Vented Vial Spike w/ChemoClave, Closed Vial Spike, 13mm w/ChemoClave, Spinning Spiros w/Red Cap, REF: CH3949; 46" (117 cm) 50 ml, Diluent Set for Channel 2 w/20 Drop Drip Chamber, Spiros, REF: CH4002
Product Codes/Lot Numbers:
Lots: 4553702, 4605473, 4763191, 4775520, 4849205, 4549847, 4548779, 4572314, 4734247, 4763917, 4849208, 4741169, 4763921, 4560231, 4750962, 4549850, 4763200, 4734249, 4794307, 4806855, 4849220, 4861176, 4877472, 4605484, 4749780, 4795448, 4876942, 4719395, 4737146, 4764825, 4800210, 4866411, 4800214, 4734262, 4794308, 4795454, 4806862, 4849239, 4849248, 4849251, 4605504, 4719410, 4734273, 4757665, 4763982, 4799740, 4861174, 4605074, 4741282, 4751117, 4809117, 4877145, 4894538, 4557534, 4733509, 4741291, 4743763, 4774120, 4733512, 4865834, 4894550, 4573105, 4803126, 4808046, 4860027, 4806400, 4720476, 4724725, 4749443, 4786358, 4802674, 4849131, 4783296, 4730109, 4742662, 4759699, 4768019, 4794497, 4877852, 4897107, 4712378, 4749445, 4799839, 4806855, 4861176, 4719395, 4800210, 4764826, 4800214, 4719927, 4726732, 4733428, 4734262, 4741183, 4756802, 4763958, 4794308, 4794310, 4795452, 4806862, 4849230, 4849233, 4849239, 4719410, 4799839, 4859282
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0704-2021
Related Recalls
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Port weld of drug transfer device may separate or break during use and potentially result in a leak.