30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger, REF: CL3011; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL3011T; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3020; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, Drop-In Red Cap, REF: CL3028; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3028T; 42" (107 cm) Appx 5.2 ml, Admin Set w/20 Drop In-Line Chamber, ChemoLock w/Red Cap, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3361; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock; Spinning Spiros w/Red Cap, REF: CL3538; Oncology Kit w/31" (79 cm) 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros; ChemoLock; ChemoLock Universal Vented Vial Spike; Bag Hanger, REF CL3929; 8.5" (22 cm) Appx 1.0 ml, Ext Set w/MicroClave Clear, ChemoLock Port, Spiros w/Red Cap, Clamp, REF: CL3931; 7.5" (19 cm) Appx 1.1 ml, Bifuse Ext Set w/ChemoLock Port, ChemoClave, Spiros, 2 Clamps, REF: CL3932; 30 IN(76cm)APPX 2.1ml, EXT SET w/ChemoLock w/RED CAP, Spiros RED CAP, REF: CL3935; Oncology Kit w/ChemoLock w/ChemoClave; Spinning Spiros w/ Red Cap, REF: CL3941; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock w/Red Cap, Vented Cap; ChemoLock Port w/Bag Spike; Spiros w/Red Cap, REF: CL3946; Oncology Kit, ChemoLock Bag Spike w/Additive Port, Vented Cap, Spiros, REF: CL3948; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960T; 60" (152 cm) Appx 7.4 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL4100; 30" (76 cm) Appx 3.6 ml, Admin Set w/20 Drop In-Line Drip Chamber, ChemoLock Additive Port, Spiros w/Red Cap, REF: CL4131; 45" (114 cm) Appx 1.7 ml, Smallbore Ext Set w/Spiros w/Red Cap, Anti-Siphon Valve, Clamp, Y-Connector w/ChemoLock Port, Rotating Luer, REF: CL4151; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros w/Red Cap, Drop-In ChemoLock Injector, Bag Hanger, REF: CL3978
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lots: 4720479, 4724743, 4735699, 4775051, 4803047, 4808098, 4868079, 4868080, 4735696, 4763384, 4800367, 4866943, 4908482, 4724700, 4742639, 4758414, 4775030, 4787228, 4808060, 4848204, 4712385, 4724701, 4605106, 4733557, 4757782, 4775514, 4800370, 4724728, 4765922, 4867983, 4724703, 4749489, 4803140, 4878873, 4742641, 4758385, 4794501, 4720364, 4752843, 4799894, 4850394, 4866135, 4790865, 4735669, 4742630, 4758441, 4764898, 4787232, 4848213, 4867995, 4743901, 4765927, 4775070, 4777973, 4794503, 4606697, 4735662, 4803124, 4808069, 4867961, 4897098, 4898098, 4899830, 4573189, 4606684, 4749454, 4756268, 4803136, 4808059, 4848185, 4878880, 4726940, 4742091, 4738028, 4749474, 4560357, 4735675, 4714539, 4763384, 4800367, 4866943, 4908482, 4742641
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- ICU Medical, Inc.
- Reason for Recall:
- Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger, REF: CL3011; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL3011T; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3020; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, Drop-In Red Cap, REF: CL3028; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3028T; 42" (107 cm) Appx 5.2 ml, Admin Set w/20 Drop In-Line Chamber, ChemoLock w/Red Cap, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3361; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock; Spinning Spiros w/Red Cap, REF: CL3538; Oncology Kit w/31" (79 cm) 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros; ChemoLock; ChemoLock Universal Vented Vial Spike; Bag Hanger, REF CL3929; 8.5" (22 cm) Appx 1.0 ml, Ext Set w/MicroClave Clear, ChemoLock Port, Spiros w/Red Cap, Clamp, REF: CL3931; 7.5" (19 cm) Appx 1.1 ml, Bifuse Ext Set w/ChemoLock Port, ChemoClave, Spiros, 2 Clamps, REF: CL3932; 30 IN(76cm)APPX 2.1ml, EXT SET w/ChemoLock w/RED CAP, Spiros RED CAP, REF: CL3935; Oncology Kit w/ChemoLock w/ChemoClave; Spinning Spiros w/ Red Cap, REF: CL3941; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock w/Red Cap, Vented Cap; ChemoLock Port w/Bag Spike; Spiros w/Red Cap, REF: CL3946; Oncology Kit, ChemoLock Bag Spike w/Additive Port, Vented Cap, Spiros, REF: CL3948; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960T; 60" (152 cm) Appx 7.4 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL4100; 30" (76 cm) Appx 3.6 ml, Admin Set w/20 Drop In-Line Drip Chamber, ChemoLock Additive Port, Spiros w/Red Cap, REF: CL4131; 45" (114 cm) Appx 1.7 ml, Smallbore Ext Set w/Spiros w/Red Cap, Anti-Siphon Valve, Clamp, Y-Connector w/ChemoLock Port, Rotating Luer, REF: CL4151; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros w/Red Cap, Drop-In ChemoLock Injector, Bag Hanger, REF: CL3978
Product Codes/Lot Numbers:
Lots: 4720479, 4724743, 4735699, 4775051, 4803047, 4808098, 4868079, 4868080, 4735696, 4763384, 4800367, 4866943, 4908482, 4724700, 4742639, 4758414, 4775030, 4787228, 4808060, 4848204, 4712385, 4724701, 4605106, 4733557, 4757782, 4775514, 4800370, 4724728, 4765922, 4867983, 4724703, 4749489, 4803140, 4878873, 4742641, 4758385, 4794501, 4720364, 4752843, 4799894, 4850394, 4866135, 4790865, 4735669, 4742630, 4758441, 4764898, 4787232, 4848213, 4867995, 4743901, 4765927, 4775070, 4777973, 4794503, 4606697, 4735662, 4803124, 4808069, 4867961, 4897098, 4898098, 4899830, 4573189, 4606684, 4749454, 4756268, 4803136, 4808059, 4848185, 4878880, 4726940, 4742091, 4738028, 4749474, 4560357, 4735675, 4714539, 4763384, 4800367, 4866943, 4908482, 4742641
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0703-2021
Related Recalls
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Port weld of drug transfer device may separate or break during use and potentially result in a leak.