Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Multiport, Models: 1) Model # WC77, 2) Model # WC77D (Demo), 3) Model # WC77R (Refurbished); used in conjunction with Warming Blankets and Mattresses.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 1) UDI-DI 00850011479025, Serial Numbers: 20080060-24060065; 2) UDI/DI 00850011479292, Serial Numbers: 20080060-24060065; 3) UDI-DI 00850011479308, Serial Numbers: 20080060-24060065
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Augustine Temperature Management, LLC
- Reason for Recall:
- Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Multiport, Models: 1) Model # WC77, 2) Model # WC77D (Demo), 3) Model # WC77R (Refurbished); used in conjunction with Warming Blankets and Mattresses.
Product Codes/Lot Numbers:
1) UDI-DI 00850011479025, Serial Numbers: 20080060-24060065; 2) UDI/DI 00850011479292, Serial Numbers: 20080060-24060065; 3) UDI-DI 00850011479308, Serial Numbers: 20080060-24060065
Distribution:
Distributed in: US
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0701-2025
Related Recalls
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia
Augustine Temperature Management
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia
Augustine Temperature Management
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses.
Augustine Temperature Management
Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .