Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination with the operating microscopes HS Hi-R NEO 900 (Ref no. 657 820) and HS Hi-R NEO 900A (Ref no. 657 821), with software REF 615 588 versions 2.0 to 3.3 - Product Usage: use in ophthalmology and in other disciplines with vertical direction of view

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Floor stands FS 2-11, FS 2-15 with the software REF 615 588 in versions 2.0 to 3.3. FS 2-11 Ref. # 615H511 Serial Numbers:From 1001; FS 2-15 Ref. #615H515 Serial Numbers From 1001; HS Hi-R NEO 900 Ref. # 657 Serial Numbers 820 from 101; HS Hi-R NEO 900A Ref # 657 821 Serial Numbers from 101 Unique Device Identifier (UDI) HS Hi-R NEO 900 +EMWM657 820 0/$+serial no. HS Hi-R NEO 900A +EMWM657 821 0/$+serial no. FS 2 11 +EMWM 615H5110/$+serial no. FS 2-15 +EMWM 615H515 0/$+serial no.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Haag-Streit USA Inc
Reason for Recall:
Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing the button, result in potential risk to the patient's eye
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination with the operating microscopes HS Hi-R NEO 900 (Ref no. 657 820) and HS Hi-R NEO 900A (Ref no. 657 821), with software REF 615 588 versions 2.0 to 3.3 - Product Usage: use in ophthalmology and in other disciplines with vertical direction of view

Product Codes/Lot Numbers:

Floor stands FS 2-11, FS 2-15 with the software REF 615 588 in versions 2.0 to 3.3. FS 2-11 Ref. # 615H511 Serial Numbers:From 1001; FS 2-15 Ref. #615H515 Serial Numbers From 1001; HS Hi-R NEO 900 Ref. # 657 Serial Numbers 820 from 101; HS Hi-R NEO 900A Ref # 657 821 Serial Numbers from 101 Unique Device Identifier (UDI) HS Hi-R NEO 900 +EMWM657 820 0/$+serial no. HS Hi-R NEO 900A +EMWM657 821 0/$+serial no. FS 2 11 +EMWM 615H5110/$+serial no. FS 2-15 +EMWM 615H515 0/$+serial no.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0697-2021

Related Recalls