MOSAIQ Oncology Information System Version 2.64. MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software Build(s): 2.64 SP9 and higher; UDI (01)00858164002091(10)2.64.235
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Elekta, Inc.
- Reason for Recall:
- If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter for imaging when it is sent to the machine.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- completed
Product Information
Full Description:
MOSAIQ Oncology Information System Version 2.64. MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery
Product Codes/Lot Numbers:
Software Build(s): 2.64 SP9 and higher; UDI (01)00858164002091(10)2.64.235
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0686-2019
Related Recalls
Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.
Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
Due to adapter not being properly aligned or locked in place.