V6 Rear Case Assembly, Product Code 35701, a replacement part for the SIGMA Spectrum Infusion Pump -- SIGMA LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended. Product Code 35701 is a replacement assembly that is shipped to customers for self-repair of the Spectrum Infusion Pump. It is not a registered finished good.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Product Code 35701; distributed between 02/04/2015 - 06/02/2015.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Baxter Healthcare Corp.
Reason for Recall:
A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

V6 Rear Case Assembly, Product Code 35701, a replacement part for the SIGMA Spectrum Infusion Pump -- SIGMA LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended. Product Code 35701 is a replacement assembly that is shipped to customers for self-repair of the Spectrum Infusion Pump. It is not a registered finished good.

Product Codes/Lot Numbers:

Product Code 35701; distributed between 02/04/2015 - 06/02/2015.