MICRO-X Rover Mobile X-ray System, # MXU-RV35

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Micro-X Ltd.
Reason for Recall:
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Product Codes/Lot Numbers:

UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0664-2025

Related Recalls

Potential for internal fastening mechanism within generator of mobile x-Ray to fail.

Apr 4, 2025 Diagnostic Equipment View Details →

Potential for internal fastening mechanism within generator of mobile x-Ray to fail.

Apr 4, 2025 Diagnostic Equipment View Details →

Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.

Mar 28, 2025 Diagnostic Equipment Nationwide View Details →