Recall Details
- Company:
- The Anspach Effort, Inc.
- Reason for Recall:
- The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the hand piece's outer hose can rupture without warning if the air flow is blocked during use..
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
***REF IMRI-MOTOR***iMRI Handpiece. Use with iMRI System Only***Rx Only***Manufacturer: The Anspach Effort, Inc. Intended for cutting and shaping bone.
Product Codes/Lot Numbers:
Serial Number F37306977201
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0657-2014
Related Recalls
Pediatric Craniotome Ref:CRANI-P-G1
The Anspach Effort
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
Adult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium
The Anspach Effort
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury