Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced during the same time period in which a main board was replaced using replacement main board part numbers G6000361, G6000435, G6001260, G6001560, G6001562 and G6002728.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Medfusion 3500 and 4000 Syringe Infusion Pumps manufactured between April 2012 and June 2015; and Medfusion 3010 and 3010A (and pump models 3500 and 40000 serviced during April 20121 and June 2015 in which a main board was replaced.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Smiths Medical ASD Inc.
- Reason for Recall:
- There have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall occurred in 2015.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced during the same time period in which a main board was replaced using replacement main board part numbers G6000361, G6000435, G6001260, G6001560, G6001562 and G6002728.
Product Codes/Lot Numbers:
Medfusion 3500 and 4000 Syringe Infusion Pumps manufactured between April 2012 and June 2015; and Medfusion 3010 and 3010A (and pump models 3500 and 40000 serviced during April 20121 and June 2015 in which a main board was replaced.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0653-2022
Related Recalls
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.