ScopeValet Guardian Disposable Endoscope Valves Set Distributed By Ruhof Corporation, 393 Sagamore Avenue, Mineola, NY 11501 Product Code: 345SVVO1

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Numbers: 19122414, 20030401, 20030402 Unit UDI: 0810929040380 Box UDI: 10810929040387 Carton Box UDI: 20810929040384
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
GA HEALTH COMPANY LIMITED
Reason for Recall:
Suction button is reported sticky and/or broken during or after the procedure, may cause prolonged procedure
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

ScopeValet Guardian Disposable Endoscope Valves Set Distributed By Ruhof Corporation, 393 Sagamore Avenue, Mineola, NY 11501 Product Code: 345SVVO1

Product Codes/Lot Numbers:

Lot Numbers: 19122414, 20030401, 20030402 Unit UDI: 0810929040380 Box UDI: 10810929040387 Carton Box UDI: 20810929040384

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0648-2021

Related Recalls