Immunoradiometric assay. Device is an immunoradiometric assay designed for the quantitative measurement of thyroid stimulating hormone (thyrotropin, TSH) in serum. It is intended strictly for in vitro diagnostic. Used as an aid in the assessment of thyroid status.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lots 742 and 743
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Healthcare Diagnostics
Reason for Recall:
The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 (level G) lot 035. The affected calibrator vials can be identified by a darker yellow color than normal. The TSI3 (level A), TSI7 (level E), and TSIX (level H)
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Immunoradiometric assay. Device is an immunoradiometric assay designed for the quantitative measurement of thyroid stimulating hormone (thyrotropin, TSH) in serum. It is intended strictly for in vitro diagnostic. Used as an aid in the assessment of thyroid status.

Product Codes/Lot Numbers:

Lots 742 and 743

Distribution:

Distributed in: CA, GA, IL, IN, MI, MN, NY

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0648-2013

Related Recalls

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Oct 6, 2025 Other Medical Devices Nationwide View Details →

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Oct 6, 2025 Infusion Pumps Nationwide View Details →

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Oct 6, 2025 Other Medical Devices Nationwide View Details →