Siemens CLINITEK Atlas Positive and Negative Control Strips for Urine Chemistry Catalog Number: 5019 When used with the CLINITEK ATLAS¿ Automated urine Chemistry Analyzer and CLINITEK ATLAS¿ 10 reagent Pak, CLINITEK ATLAS¿ Positive and Negative Control Strips provide defined results for color, clarity, glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogen, nitrite, and leukocytes. Each CLINITEK ATLAS Positive Control Strip is a firm plastic strip to which are affixed seven separate analyte areas.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Carton/Bottle Lot Number: S9093061A P9093061A S9094081A P9094081A S0097091C P0097091C S9093061A P9093061A S9094081A P9094081A S0097091C P0097091C S0100101B P0100101B S0103042A P0103042A S0103042H P0103042H S0106062A P0106062A S9101022A P9101022A S9101022B P9101022B S9105052A P9105052A
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics Inc
- Reason for Recall:
- When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Siemens CLINITEK Atlas Positive and Negative Control Strips for Urine Chemistry Catalog Number: 5019 When used with the CLINITEK ATLAS¿ Automated urine Chemistry Analyzer and CLINITEK ATLAS¿ 10 reagent Pak, CLINITEK ATLAS¿ Positive and Negative Control Strips provide defined results for color, clarity, glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogen, nitrite, and leukocytes. Each CLINITEK ATLAS Positive Control Strip is a firm plastic strip to which are affixed seven separate analyte areas.
Product Codes/Lot Numbers:
Carton/Bottle Lot Number: S9093061A P9093061A S9094081A P9094081A S0097091C P0097091C S9093061A P9093061A S9094081A P9094081A S0097091C P0097091C S0100101B P0100101B S0103042A P0103042A S0103042H P0103042H S0106062A P0106062A S9101022A P9101022A S9101022B P9101022B S9105052A P9105052A
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0636-2013
Related Recalls
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.