Carousel MAS1025-01, part of the Brain Surgical Accessories Intra-operative Kit (KIT0295), accessory used with the Renaissance System (TPL0038). Indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of, high contrast objects.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Affected products: all Carousels under Level 01 were affected and therefore are applicable to this issue. The related lots include: WP065611, WP066389, WP066411, WP066412, WP066420. Explanation of the coding system: a part number (e.g. MAS1025-01) is composed of a generic part number related to the part (e.g. MAS1025) and the level of the part (e.g. 01).
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Mazor Robotics Ltd
- Reason for Recall:
- In some situations the carousel position pin (locker) may cause the carousel upper plate to pop out and the carousel may then inadvertently move.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Carousel MAS1025-01, part of the Brain Surgical Accessories Intra-operative Kit (KIT0295), accessory used with the Renaissance System (TPL0038). Indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of, high contrast objects.
Product Codes/Lot Numbers:
Affected products: all Carousels under Level 01 were affected and therefore are applicable to this issue. The related lots include: WP065611, WP066389, WP066411, WP066412, WP066420. Explanation of the coding system: a part number (e.g. MAS1025-01) is composed of a generic part number related to the part (e.g. MAS1025) and the level of the part (e.g. 01).
Distribution:
Distributed in: US
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0609-2019
Related Recalls
Software anomalies which can result in two separate issues: A lagging screen display during a navigated procedure; or the potential for a mismatch between the navigated screw and the screw information presented on the screen.
MAZOR X robotic guidance system, Model TPL0059, with software versions 4.2.2 or 5.0.1
Mazor Robotics
Software update
The firm has become aware that the Snapshot Trackers were not manufactured in accordance with specifications and could potentially lead to navigational inaccuracies, which may result in patient injury, e.g. nerve damage or neurological deficit/dysfunction