Product Code 0847, MEGADYNE MEGA SOFT Universal Plus Patient Return Electrode, Reusable Patient Return Electrode, Mega Soft Universal Plus

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI 10614559104842 All units in expiry
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Megadyne Medical Products, Inc.
Reason for Recall:
Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Product Code 0847, MEGADYNE MEGA SOFT Universal Plus Patient Return Electrode, Reusable Patient Return Electrode, Mega Soft Universal Plus

Product Codes/Lot Numbers:

UDI-DI 10614559104842 All units in expiry

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0599-2024

Related Recalls

To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.

Jun 17, 2024 Other Medical Devices Nationwide View Details →

To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.

Jun 17, 2024 Other Medical Devices Nationwide View Details →

To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.

Jun 17, 2024 Other Medical Devices Nationwide View Details →