BRASSELER KM166-310-35 Twist Drill Quick Connect Diameter 3.5 mm, Total Length 112mm, Effective Length, 86mm - Product Usage: Twist drill intended use: to drill into bone in hip or knee surgeries.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Number NR1VM
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Brasseler USA I Lp
- Reason for Recall:
- One lot of 3.5 mm twist drills were incorrectly laser etched as 2.5 mm the shaft of the device.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
BRASSELER KM166-310-35 Twist Drill Quick Connect Diameter 3.5 mm, Total Length 112mm, Effective Length, 86mm - Product Usage: Twist drill intended use: to drill into bone in hip or knee surgeries.
Product Codes/Lot Numbers:
Lot Number NR1VM
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0591-2020
Related Recalls
Brasseler 8S RA SHP RD (Super-Sharp Round) H1S.21.023, REF 001156T0), carbide dental bur, 10 burs/package, Rx.
Brasseler USA I
The device has a grip detail (right angle latch) out of specification causing the bur to not adequately latch in the handpiece.
103 packaged and labeled tubes (103 product tubes with 6 pins per tube) of KM168-39-76 Single Diamond Smooth Steinmann Pins were labeled with the incorrect product labels. The tubes were incorrectly labeled as KM169-39-76 Single Diamond Threaded Steinmann Pins.