Natus Quantum System with NeuroWorks Software. Catalog /Part Numbers: 013926. The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks /SleepWorks software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Number 013926020008 013926020009 013926020006 013926020016 013926020014 013926030024 013926030031 013926020001 013926020004 013926030059 013926030060 013926030074 013926030075 013926030052 013926030017 013926030018 013926030021 013926030053 013926030101 013926030058 013926030019 013926030038 013926030039 013926030044 013926020015 013926030097 013926020013 013926030055 013926030056 013926030081 013926030096 013926030078 013926030048 013926030057 013926030064 013926030047 013926030027 013926030025 013926030042 013926030023 013926030030 013926030036 013926030026 013926030050 013926030080 013926030099 013926030035 013926030046 013926030049 013926030073 013926030020 013926030092 013926030040 013926030041 013926030061 013926030062
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
Reason for Recall:
During an internal testing, Natus Medical Incorporated, identified that in Neuro Works 8.1 with Quantum hardware, the incoming signal is displayed with reversed polarity.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Natus Quantum System with NeuroWorks Software. Catalog /Part Numbers: 013926. The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks /SleepWorks software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals.

Product Codes/Lot Numbers:

Serial Number 013926020008 013926020009 013926020006 013926020016 013926020014 013926030024 013926030031 013926020001 013926020004 013926030059 013926030060 013926030074 013926030075 013926030052 013926030017 013926030018 013926030021 013926030053 013926030101 013926030058 013926030019 013926030038 013926030039 013926030044 013926020015 013926030097 013926020013 013926030055 013926030056 013926030081 013926030096 013926030078 013926030048 013926030057 013926030064 013926030047 013926030027 013926030025 013926030042 013926030023 013926030030 013926030036 013926030026 013926030050 013926030080 013926030099 013926030035 013926030046 013926030049 013926030073 013926030020 013926030092 013926030040 013926030041 013926030061 013926030062

Distribution:

Distributed in: US, AL, AZ, AR, CO, FL, GA, MD, MA, MN, MS, NJ, NY, OH, SC, TN, TX, UT

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0581-2016

Related Recalls

Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.

May 7, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Natus has become aware that its third-party supplier Parker Laboratories Inc. is conducting a voluntary recall related to their Tensive Conductive Adhesive Gel, Ref. 22-60 Lot Number A0623009. They have received five complaints. These customers reported that the gel is too thick, is discolored (darker than usual), and in one instance, patients reported irritation after procedures using the gel.

Feb 7, 2024 Infusion Pumps Nationwide View Details →