MICRO-X Rover Mobile X-ray System, # MXU-RV35
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 9357123000037, Serial Numbers: 348, 349, 361, 351, 365, 391
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Micro-X Ltd.
- Reason for Recall:
- Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
MICRO-X Rover Mobile X-ray System, # MXU-RV35
Product Codes/Lot Numbers:
UDI-DI: 9357123000037, Serial Numbers: 348, 349, 361, 351, 365, 391
Distribution:
Distributed in: US, TX, PR, MD, CA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0580-2025
Related Recalls
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.