1488 HD 3-Chip Inline Camera; Model Number 1488-710-105 - HD 3-Chip Inline Camera, C-Mount; Product Usage: High definition camera to view endoscopic surgical sites on video monitors and capture still and video images of endoscopic surgical applications. Indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy and plastic surgery wherever a laparoscope/endoscope/arthroscope is indicated for use.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model 1488-710-105: 1488 HD 3-Chip Camera head, C-Mount; Serial numbers: 12F016274 12F016294 12F055824 12F055844 12F055874 12F055884 12F055894 12F055904 12F055914 12F055924 12F055934 12F055954 12F055964 12F055974 12F055994 12F056004 12F056024 12F056034 12F056054 12F056064 12F056094 12F056104 12F056124 12F056154 12F056164 12F056174 12F056184 12F056194 12F056204 12F056664 12F056674 12F056684 12F056694 12F056704 12F056714 12F056744 12F058884 12F058894 12F058904 12F058924 12F058934 12F058944 12F058964 12G054154 12G054174 12G054194 12G054214 12G054224 12G054234 12G054244 12G054264 12H011414 12H011434 12H011444 12H011454 12H011474 12H011494 12H050834 12H050844 12H050854 12I039444 12I039454 12I039464 12I039474 12I039494 12I039504 12I039524 12I039534 12I040844 12I040854 12I040864 12I040874 12I040894 12I040924 12I040934 12I047324. 12D030204 12D030214 12D030224 12F060834 12F060844 12F060854 12F060874 12F060894 12F060904 12F060914 12F060924 12F060934 12F060954 12F060964 12F060974 12F060984 12F060994 12F061024 12F061054 12F061064 12F061074 12F061084 12F061094 12F061104 12F061114 12F061134 12F061154 12F061164 12F061174 12F061184 12F061194 12F061234 12F061264 12F061274 12F061294 12F061304 12F061314 12F061324 12H001234 12H001244 12H001254 12H001264 12H001274 12H001284 12H001294 12H001304 12H001324 12H038814 12H038824 12H038834 12H038844 12H038854 12H038864 12H038874 12H038894 12H059034 12H059044 12H059054 12H059084 12H059094 12H059104 12H059124 12I004814 12I004834 12I004844 12I004854 12I004864 12I004874 12I004884 12I004894 12I004904 12I004914 12I004954 12I004994 12I005004 12I005014 12I005024 12I005044 12I005054 12I005064 12I005074 12I005084 12I005094 12I005104 12I014974 12I014984 12I014994 12I015004 12I015014 12I015024 12I015034 12I015044 12I015054 12I015064 12I016374 12I016384 12I016394 12I016404 12I016414 12I016424 12I016434 12I016444 12I016454 12I016464 12I016474 12I016484 12I016494 12I016504 12I016514 12I016524 12I016534 12I016544 12I016554 12I016564 12I018674 12I018684 12I018694 12I018704 12I018714 12I018724 12I018734 12I018744 12I018754 12I018764 12I018774 12I018784 12I018794 12I018804 12I018814 12I018824 12I018834 12I018844 12I018854 12I018864 12I018874 12I018884 12I018894 12I018904 12I018914 12I018924 12I018934 12I018944 12I019254 12I019264 12I019274 12I019284 12I019294 12I019304 12I019314 12I019324 12I019334 12I019344 12I033824 12I033874 12I033894 12I033904 12I033914 12I033924 12I033934 12I033944 12I033954 12I033964 12I033974 12I033984 12I033994 12I034004 12I034214 12I034224 12I034234 12I034244 12I034254 12I034264 12I034274 12I034284 12I034304 12I034814 12I034824 12I034834 12I034844 12I034854 12I034864 12I034874 12I034884 12I034894 12I034924 12I034934 12I034944 12I034954 12I034964 12I034974 12I034984 12I034994 12I035214 12I035224 12I035244 12I035254 12I035284 12I035294 12I035304 12J002614 12J002634 12J002664 12J002674 12J002694 12J002714 12J002724 12J002744 12J002804 12J002824 12J002854 12J002864 12J020084 12J020144 12J020154 12J020164 12J020594 12J020614 12J020634 12J020644 12J020654 12J020664 12J063964 12J063994 12J064004 12J064024 12J064034 12J064044 12J064064 12J064074 12J064084 12J064094 12J064104 12J064114 12J064124 12J064144 12J064154 12J064164 12J064174 12J064184 12J064194 12J064204 12J064214 12J064224 12J072934 12J072964 12J072994 12J073014 12J073034 12J073044 12J073054 12J073064 12J073074 12J073084 12K002644 12K002654.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Stryker Endoscopy
Reason for Recall:
In certain operating room environments, a colored screen image occurs with the 1488 Cameras that were manufactured during a specific time period
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

1488 HD 3-Chip Inline Camera; Model Number 1488-710-105 - HD 3-Chip Inline Camera, C-Mount; Product Usage: High definition camera to view endoscopic surgical sites on video monitors and capture still and video images of endoscopic surgical applications. Indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy and plastic surgery wherever a laparoscope/endoscope/arthroscope is indicated for use.

Product Codes/Lot Numbers:

Model 1488-710-105: 1488 HD 3-Chip Camera head, C-Mount; Serial numbers: 12F016274 12F016294 12F055824 12F055844 12F055874 12F055884 12F055894 12F055904 12F055914 12F055924 12F055934 12F055954 12F055964 12F055974 12F055994 12F056004 12F056024 12F056034 12F056054 12F056064 12F056094 12F056104 12F056124 12F056154 12F056164 12F056174 12F056184 12F056194 12F056204 12F056664 12F056674 12F056684 12F056694 12F056704 12F056714 12F056744 12F058884 12F058894 12F058904 12F058924 12F058934 12F058944 12F058964 12G054154 12G054174 12G054194 12G054214 12G054224 12G054234 12G054244 12G054264 12H011414 12H011434 12H011444 12H011454 12H011474 12H011494 12H050834 12H050844 12H050854 12I039444 12I039454 12I039464 12I039474 12I039494 12I039504 12I039524 12I039534 12I040844 12I040854 12I040864 12I040874 12I040894 12I040924 12I040934 12I047324. 12D030204 12D030214 12D030224 12F060834 12F060844 12F060854 12F060874 12F060894 12F060904 12F060914 12F060924 12F060934 12F060954 12F060964 12F060974 12F060984 12F060994 12F061024 12F061054 12F061064 12F061074 12F061084 12F061094 12F061104 12F061114 12F061134 12F061154 12F061164 12F061174 12F061184 12F061194 12F061234 12F061264 12F061274 12F061294 12F061304 12F061314 12F061324 12H001234 12H001244 12H001254 12H001264 12H001274 12H001284 12H001294 12H001304 12H001324 12H038814 12H038824 12H038834 12H038844 12H038854 12H038864 12H038874 12H038894 12H059034 12H059044 12H059054 12H059084 12H059094 12H059104 12H059124 12I004814 12I004834 12I004844 12I004854 12I004864 12I004874 12I004884 12I004894 12I004904 12I004914 12I004954 12I004994 12I005004 12I005014 12I005024 12I005044 12I005054 12I005064 12I005074 12I005084 12I005094 12I005104 12I014974 12I014984 12I014994 12I015004 12I015014 12I015024 12I015034 12I015044 12I015054 12I015064 12I016374 12I016384 12I016394 12I016404 12I016414 12I016424 12I016434 12I016444 12I016454 12I016464 12I016474 12I016484 12I016494 12I016504 12I016514 12I016524 12I016534 12I016544 12I016554 12I016564 12I018674 12I018684 12I018694 12I018704 12I018714 12I018724 12I018734 12I018744 12I018754 12I018764 12I018774 12I018784 12I018794 12I018804 12I018814 12I018824 12I018834 12I018844 12I018854 12I018864 12I018874 12I018884 12I018894 12I018904 12I018914 12I018924 12I018934 12I018944 12I019254 12I019264 12I019274 12I019284 12I019294 12I019304 12I019314 12I019324 12I019334 12I019344 12I033824 12I033874 12I033894 12I033904 12I033914 12I033924 12I033934 12I033944 12I033954 12I033964 12I033974 12I033984 12I033994 12I034004 12I034214 12I034224 12I034234 12I034244 12I034254 12I034264 12I034274 12I034284 12I034304 12I034814 12I034824 12I034834 12I034844 12I034854 12I034864 12I034874 12I034884 12I034894 12I034924 12I034934 12I034944 12I034954 12I034964 12I034974 12I034984 12I034994 12I035214 12I035224 12I035244 12I035254 12I035284 12I035294 12I035304 12J002614 12J002634 12J002664 12J002674 12J002694 12J002714 12J002724 12J002744 12J002804 12J002824 12J002854 12J002864 12J020084 12J020144 12J020154 12J020164 12J020594 12J020614 12J020634 12J020644 12J020654 12J020664 12J063964 12J063994 12J064004 12J064024 12J064034 12J064044 12J064064 12J064074 12J064084 12J064094 12J064104 12J064114 12J064124 12J064144 12J064154 12J064164 12J064174 12J064184 12J064194 12J064204 12J064214 12J064224 12J072934 12J072964 12J072994 12J073014 12J073034 12J073044 12J073054 12J073064 12J073074 12J073084 12K002644 12K002654.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0570-2015

Related Recalls

When operating the device in high flow mode (40L/min, 15 mmHg) and connecting to a house gas inlet source, within seconds of insufflating the flow rate would drop to 0L/min and the red Check Gas Supply symbol would appear on the display, immediately suspending insufflation.

Mar 28, 2016 Diagnostic Equipment Nationwide View Details →

Stryker Endoscopy is recalling all non expired SERFAS 90 degree Energy Probes due to reports of fragments of the probe breaking off into the patient.

Jun 3, 2015 Implants & Prosthetics Nationwide View Details →

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Nov 17, 2014 Surgical Instruments Nationwide View Details →