Philips Allura systems with monoplane fixed ceiling mounts: 722001 Allura Xper FD10C 722003 Allura Xper FD10 722006 Allura Xper FD20 722010 Allura Xper FD10 722012 Allura Xper FD20 722015 Allura Xper FD20 OR Table 722016 Integris H5000 C / Allura 9C 722018 Integris Allura 9 722022 Allura Xper FD10 OR Table 722023 Allura Xper FD20 OR Table 722026 Allura Xper FD10 722028 Allura Xper FD20 722033 Allura Xper FD10 OR Table 722035 Allura Xper FD20 OR Table 722043 Integris Allura 15 & 12 (monoplane) 72246 Poly G - OMCP - VISUB - CCD (H5000)
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model/Name/UDI-DI: 722001 Allura Xper FD10C; 722003 Allura Xper FD10; 722006 Allura Xper FD20 (01)00884838059115(21); 722010 Allura Xper FD10 (01)00884838059030(21); 722012 Allura Xper FD20 (01)00884838059054(21); 722015 Allura Xper FD20 OR Table; 722016 Integris H5000 C / Allura 9C ; 722018 Integris Allura 9; - 722022 Allura Xper FD10 OR Table; 722023 Allura Xper FD20 OR Table (01)00884838059085(21;) 722026 Allura Xper FD10 (01)00884838054189(21); 722028 Allura Xper FD20 (01)00884838054202(21); 722035 Allura Xper FD20 OR Table (01)00884838054240(21); 722043 Integris Allura 15 & 12 (monoplane).
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Reason for Recall:
- Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Philips Allura systems with monoplane fixed ceiling mounts: 722001 Allura Xper FD10C 722003 Allura Xper FD10 722006 Allura Xper FD20 722010 Allura Xper FD10 722012 Allura Xper FD20 722015 Allura Xper FD20 OR Table 722016 Integris H5000 C / Allura 9C 722018 Integris Allura 9 722022 Allura Xper FD10 OR Table 722023 Allura Xper FD20 OR Table 722026 Allura Xper FD10 722028 Allura Xper FD20 722033 Allura Xper FD10 OR Table 722035 Allura Xper FD20 OR Table 722043 Integris Allura 15 & 12 (monoplane) 72246 Poly G - OMCP - VISUB - CCD (H5000)
Product Codes/Lot Numbers:
Model/Name/UDI-DI: 722001 Allura Xper FD10C; 722003 Allura Xper FD10; 722006 Allura Xper FD20 (01)00884838059115(21); 722010 Allura Xper FD10 (01)00884838059030(21); 722012 Allura Xper FD20 (01)00884838059054(21); 722015 Allura Xper FD20 OR Table; 722016 Integris H5000 C / Allura 9C ; 722018 Integris Allura 9; - 722022 Allura Xper FD10 OR Table; 722023 Allura Xper FD20 OR Table (01)00884838059085(21;) 722026 Allura Xper FD10 (01)00884838054189(21); 722028 Allura Xper FD20 (01)00884838054202(21); 722035 Allura Xper FD20 OR Table (01)00884838054240(21); 722043 Integris Allura 15 & 12 (monoplane).
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0558-2024
Related Recalls
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.