Merlin@home with the MerlinOnDemand capability Transmitter REF EX1150 Merlin@home with the MerlinOnDemand" (MoD) capability is a transmitter configured for multiple patient use. It can receive information from an implantable heart device in unpaired mode. With MoD capability, the transmitter will retrieve full follow-up data or retrieve data for display on the Merlin.net" Patient Care Network (PCN). The MoD capability enables clinicians to do the following: . support remote care patients in emergency rooms and heart failure clinics . triage patients for in clinic follow-ups . support patients in a clinical setting with Merlin.net PCN
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number: EX1150 UDI-DI code: Pending Serial Numbers: Pending
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Medical
- Reason for Recall:
- Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Merlin@home with the MerlinOnDemand capability Transmitter REF EX1150 Merlin@home with the MerlinOnDemand" (MoD) capability is a transmitter configured for multiple patient use. It can receive information from an implantable heart device in unpaired mode. With MoD capability, the transmitter will retrieve full follow-up data or retrieve data for display on the Merlin.net" Patient Care Network (PCN). The MoD capability enables clinicians to do the following: . support remote care patients in emergency rooms and heart failure clinics . triage patients for in clinic follow-ups . support patients in a clinical setting with Merlin.net PCN
Product Codes/Lot Numbers:
Model Number: EX1150 UDI-DI code: Pending Serial Numbers: Pending
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0552-2025
Related Recalls
There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).
There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).
Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.