V-Twin, REF 6002-800 or 6002-850, SN 18-3304, IVD, (01)03661540600180

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    (OUS) V-Twin¿ 6002-800 serial numbers from 3-3021 till 18-3345; (US) V-Twin¿6002-850 serial numbers from 3-3006 till 9-4037
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Vital Scientific N.V.
Reason for Recall:
During in-house testing it was determined that their is a possibility that during a sample run, the instrument may aspirate a sample from the wrong sample container resulting in a sample result and patient ID mismatch. If not detected, this may lead to an incorrect patient report.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

V-Twin, REF 6002-800 or 6002-850, SN 18-3304, IVD, (01)03661540600180

Product Codes/Lot Numbers:

(OUS) V-Twin¿ 6002-800 serial numbers from 3-3021 till 18-3345; (US) V-Twin¿6002-850 serial numbers from 3-3006 till 9-4037

Distribution:

Distributed in: NY

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0547-2020