VITROS XT 3400 Chemistry System Product Code 6844458 Running Software Version 3.8.0

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Product Code: 6844458, UDI-DI: 10758750031986; Serial Numbers: J34500140 J34500113 J34500180 J34500114 J34500257 J34500116 J34500289 J34500117 J34500319 J34500123 J34500386 J34500128 J34500422 J34500129 J34500447 J34500130 J34500473 J34500139 J34500474 J34500190 J34500510 J34500261 J34500539 J34500269 J34500547 J34500271 J34500579 J34500292 J34500588 J34500301 J34500622 J34500337 J34500656 J34500356 J34500777 J34500363 J34500818 J34500515 J34500887 J34500520 J34500893 J34500536 J34500896 J34500582 J34500911 J34500609 J34500931 J34500628 J34500658 J34500711 J34500793
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Ortho-Clinical Diagnostics, Inc.
Reason for Recall:
During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

VITROS XT 3400 Chemistry System Product Code 6844458 Running Software Version 3.8.0

Product Codes/Lot Numbers:

Product Code: 6844458, UDI-DI: 10758750031986; Serial Numbers: J34500140 J34500113 J34500180 J34500114 J34500257 J34500116 J34500289 J34500117 J34500319 J34500123 J34500386 J34500128 J34500422 J34500129 J34500447 J34500130 J34500473 J34500139 J34500474 J34500190 J34500510 J34500261 J34500539 J34500269 J34500547 J34500271 J34500579 J34500292 J34500588 J34500301 J34500622 J34500337 J34500656 J34500356 J34500777 J34500363 J34500818 J34500515 J34500887 J34500520 J34500893 J34500536 J34500896 J34500582 J34500911 J34500609 J34500931 J34500628 J34500658 J34500711 J34500793

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0546-2024

Related Recalls

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

Jul 29, 2025 Diagnostic Equipment Nationwide View Details →

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

Jul 29, 2025 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive "spikes" (imperfections) in slide images and generates the TH4-63J condition code when the threshold of 600 spikes is exceeded. Updated analysis revealed a 19-fold increase in failure rates from 0.0043% (2024 baseline) to 0.082% for Generation 67+ slides in 2025, with 100% of calibrator failures occurring at the lowest concentration level. This can result in delayed diagnosis and treatment of conditions such as hypercalcemic crisis, parathyroid disorders, and other calcium-related emergencies requiring urgent intervention.

Jul 9, 2025 Other Medical Devices Nationwide View Details →